NCT Number: NCT04995666
Quality and reliability study of hearing aid. Adults with mild to severe hearing loss will wear hearing aids daily for approximately 3 months.
Looking for future studies?
Notify Me18 year–90 year
All sexes
Interventional
Not applicable
Aurora, Illinois, 60504, United States
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Rechargeable hearing aid
Time frame: Day 1 of study
Test box electroacoustic measurements of hearing aid done at full on settings must meet manufacturer product specifications per manufacturer data sheet. This includes maximum and high frequency average output sound pressure level, maximum and high frequency average acoustic gain, frequency response range, total harmonic distortion and equivalent input noise level. This is done using an Audioscan Verifit2 electroacoustic analyzer.
Electroacoustic measurements listed above will also be taken with hearing aid set to user settings, as a baseline measure for all future appointments.
Time frame: Day 45 of study
Test box electroacoustic measurements (using the Audioscan Verifit 2 Electroacoustic Analyzer) of hearing aid done at user gain settings after daily use for extended period of time. Results will be compared to baseline electroacoustic measures taken on day 1 of study. A "pass" means that devices were within tolerance levels (per ANSI S3.22-2014) of baseline measures. A "fail" means that devices were not within tolerance levels of one or more baseline measures (i.e. output, gain, frequency range, total harmonic distortion, or equivalent input noise level).
Time frame: Day 90 of study
Test box electroacoustic measurements of hearing aid done at user gain settings after daily use for extended period of time. Test box electroacoustic measurements of hearing aid done at user gain settings after daily use for extended period of time. Results will be compared to baseline electroacoustic measures taken on day 1 of study. A "pass" means that devices were within tolerance levels (per ANSI S3.22-2014) of baseline measures. A "fail" means that devices were not within tolerance levels of one or more baseline measures (i.e. output, gain, frequency range, total harmonic distortion, or equivalent input noise level).
Sonova AG
Industry
Hearing Aid Quality Check and Reliability Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05219474
Ear Diseases, Hearing Disorders
New York, United States
View Trial DetailsNCT01805167
Ear Diseases, Hearing Disorders
Nice, France
View Trial DetailsNCT05070429
Aging, Ear Diseases
Hagerstown, Maryland, United States
View Trial DetailsNCT07501312
Ear Diseases, Hearing Disorders
Aurora, Illinois, United States
View Trial Details