Mitral valve replacement SJM™ Masters HP™ 15mm Rotatable Mechanical Heart Valve
DeviceImplantation of the SJM™ Masters HP™ 15mm Rotatable Mechanical Heart Valve in the mitral and aortic positions
NCT Number: NCT02097420
The purpose of the study is to provide evidence of safety and effectiveness of the study valve. The rationale is to offer a replacement mitral valve for patients with anatomy that is too small for the currently commercially available valves ranging in size from 16mm to 37mm.
20 subjects were enrolled as part of the IDE cohort and an additional 3 were enrolled under continued access.
Looking for future studies?
Notify MeUp to 5 year
All sexes
Interventional
Not applicable
University of Alabama at Birmingham, Birmingham, Alabama, United States
The study is a single arm, prospective, non-randomized, multi-center clinical investigation.
The objective of this study is to evaluate the safety and effectiveness of the 15mm MHV in subjects five years or less of age with a diseased, damaged, or malfunctioning mitral heart valve. The objective will be evaluated by assessing valve-related adverse events, subject survival, subject growth, and echocardiogram assessment of hemodynamic function through the five year follow-up visit as long as the valve remains implanted.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
IDE Cohort -
Inclusion criteria
Exclusion criteria
Continued Access Cohort-
Prospective Inclusion Criteria:
Prospective Exclusion Criteria:
Retrospective Eligibility Criteria:
OR
Implantation of the SJM™ Masters HP™ 15mm Rotatable Mechanical Heart Valve in the mitral and aortic positions
Time frame: 12 months
The Kaplan-Meier method is a statistical method that summarizes the rate at which subjects survive or are free from a specified event. The method uses whether or not a subject had the specified event and their time to the specified event or the end of their follow-up to calculate the rate.
Kaplan-Meier event rates with 95% confidence intervals would be calculated for overall survival. All follow-up time through 365 days post implant would be considered. If a subject discontinues prior to one year post implant, all available data would be included.
Time frame: 12 months
Peak gradient as assessed by echocardiography at 12 months post implant or when the valve is removed/replaced due to anatomical growth of the subject, whichever occurs first.
Time frame: 12 months
Mean gradient as assessed by echocardiography at 12 months post implant or when the valve is removed/replaced due to anatomical growth of the subject, whichever occurs first.
Time frame: 12 months
Valvular regurgitation as assessed by echocardiography at 12 months post implant or when the valve is removed/replaced due to anatomical growth of the subject, whichever occurs first.
Time frame: 12 months
Kaplan-Meier event rates with 95% confidence intervals would be calculated for valve-related adverse events. All follow-up time through 365 days post implant would be considered. If a subject discontinues prior to one year post implant, all available data would be included.
Time frame: 12 months
Percentiles would be summarized by visit interval from baseline to 12 months for all available data. Summary statistics including number of subjects, mean and standard deviation are provided at the 12-month visit.
Time frame: 12 months
Percentiles would be summarized by visit interval from baseline to 12 months for all available data.Summary statistics including number of subjects, mean, standard deviation, and difference between visits were used for this endpoint.
Abbott Medical Devices
Industry
HALO Clinical Study: A Single Arm, Prospective, Non-randomized, Multi Center Investigation of the SJM™ Masters HP™ 15mm Rotatable Mechanical Heart Valve
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07069673
Cardiovascular Diseases, Deglutition Disorders
Birmingham, Alabama, United States
View Trial DetailsNCT04045340
Cardiac Output, Low, Cardiovascular Diseases
Milan, Italy
View Trial DetailsNCT04787757
Cardiovascular Diseases, Heart Diseases
Milan, MI, Italy
View Trial DetailsNCT07027670
Aortic Valve Disease, Cardiac Surgical Procedures
Oxford, Oxfordshire, United Kingdom
View Trial Details