In this prospective study different methods of haemoglobin measurement in term and preterm neonates are compared with the gold standard. Non-invasive haemoglobin measurement with the Radical-7® (SpHb, Masimo®), point-of-care haemoglobin-measurement (HcHb, HemoCue@, Radiometer), blood-gas-analysis (BGAHb,ABL800®, Radiometer) are compared with haemoglobin measurement by an automated hematology analyzer (labHb, Siemens Advia®).
Looking for future studies?
Notify Me3 day–4 week
All sexes
Interventional
Not applicable
Mainz, 55131, Germany
Neonates admitted to neonatal intensive care unit or to a normal neonatal ward are enrolled in the study, if parents have given their written informed consent. Non-invasive SpHb measurements are recorded prior to routine venous sampling. Routine venous sampling includes labHb and BGAHb. Additionally HcHb is taken from the routine venous sampling (HcHbart/ven) and from an additionally capillary puncture of earlobe or heel (HcHbcap). Every neonate can participate in two measurements.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
gender both
age limits minimum age 3 days of life maximum age 44 postmenstrual weeks
accepts no healthy volunteers
Inclusion criteria
written informed consent of the parents
-
Exclusion criteria
Time frame: through study completion, an average of 3 months
accuracy is assessed by Bland-Altman-method of comparison, by Root mean square error and by out of range
Time frame: through study completion, an average of 3 months
measurement of how often the Radical-7 has to be replaced and of the time until a stable Signal appears
Time frame: through study completion, an average of 3 months
correlation of PI with the accuracy of SpHb
Time frame: through study completion, an average of 3 months
correlation of PI with the accuracy of SpHb
Time frame: through study completion, an average of 3 months
correlation of concentration of HbF with the accuracy of SpHb
Time frame: through study completion, an average of 3 months
correlation of concentration of Hb with the accuracy of SpHb
Time frame: through study completion, an average of 3 months
accuracy is assessed by Bland-Altman-method of comparison
Johannes Gutenberg University Mainz
Other
Comparison of Alternative Non-invasive and Invasive Methods of Haemoglobin Measuring With the Gold Standard in Preterm and Term Neonates
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04869319
Diagnostic Tests, Hemoglobins
New York, United States
View Trial DetailsNCT05506748
Albumins, Hemoglobins
Bharatpur, Bagmati, Nepal
View Trial DetailsNCT03816514
Blood Transfusion, Hemoglobins
Seoul, South Korea
View Trial DetailsNCT02655107
Hemoglobins
Kiel, Schleswig-Holstein, Germany
View Trial Details