Skip to main content
OpenTrials
Completed

NCT Number: NCT03501693

Fujifilm DBT Plus S-View Versus FFDM Alone in the Detection of Breast Cancer - A Pivotal Study

The purpose of the pivotal reader study is to assess the comparative accuracy of Fujifilm DBT plus S-View versus FFDM in the detection of breast cancer.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

Female

Study type

Observational

Primary location

International HealthCare, LLC

Norwalk, Connecticut, 06851, United States

About this study

This clinical research is a retrospective, pivotal, multi-reader, multi-case (MRMC) study with an enriched sample of 300 breast screening or diagnostic cases which were selected from the library of mammograms collected under Fujifilm protocol FMSU2013-004A (all subjects previously provided written informed consent agreeing their image data and supporting documentation could be used for future research and investigations). Approximately 18 qualified radiologists will independently perform two reads on all (approximately 300) cases. Each reader will read each case both as a FFDM read, and a DBT plus S-View read on the ASPIRE Bellus II workstation.

As its primary endpoint, this study is designed to evaluate whether the area under the receiver operating characteristic (ROC) curve (AUC) based on probability of malignancy (POM) scores and requiring correct lesion localization is statistically non-inferior for DBT plus S-View versus FFDM.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • •Female subjects participating in FMSU2013-004A protocol with known clinical status

Exclusion criteria

  • •Female subjects that did not have known clinical status in FMSU2013-004A

Treatment and study plan

DBT plus S-View

Device

DBT plus S-View images

FFDM alone

Device

FFDM alone images

Primary outcomes

  1. Compare Per Subject AUC: DBT Plus S-View Versus FFDM

    Time frame: 4 weeks

    Average of each radiologist reader's ROC (receiving operating characteristic curve) based on per-subject POM (probability of malignancy) scores requiring correct lesion localization. A ROC curve is a plot of Sensitivity versus 1-Specificity and is a summary of diagnostic performance of a device or clinician. The reader will first record whether there are mammographic findings and an initial BI-RADS assessment of 0, 1, or 2. If the answer to this question is "no" the reader will be asked to confirm for this case: a BI-RADS assessment category of 1 or 2, a Probability of Malignancy (POM) score following the POM guidance provided in the protocol, and recall decision of "no". If the reader answers "yes" to whether there are mammographic findings, the reader will be asked to confirm an initial BI-RADS assessment category of 0, and will then provide detailed information on each suspicious finding (up to three findings) and the decision whether or not to recall the subject.

Secondary outcomes

  1. Compare Per-subject Average Recall Rate for All Non-cancer Cases for DBT Plus S-View Versus FFDM

    Time frame: 4 weeks

    Based on recall rate for all non-cancer cases for DBT plus S-View versus FFDM, using non-inferiority margin delta = 0.05. The reader will first record whether there are mammographic findings and an initial BI-RADS assessment of 0, 1, or 2. If the answer to this question is "no" the reader will be asked to confirm for this case: a BI-RADS assessment category of 1 or 2, a Probability of Malignancy (POM) score following the POM guidance provided in the protocol, and recall decision of "no". If the reader answers "yes" to whether there are mammographic findings, the reader will be asked to confirm an initial BI-RADS assessment category of 0, and will then provide detailed information on each suspicious finding (up to three findings) and the decision whether or not to recall the subject, as well as overall POM score and BI-RADS assessment category.

  2. Compare Per-subject Average Recall Rate for DBT Plus S-View Versus FFDM for All Cancer Cases

    Time frame: 4 weeks

    Based on recall rate for all cancer cases for DBT plus S-View versus FFDM, using non-inferiority margin delta = 0.10. The reader will first record whether there are mammographic findings and an initial BI-RADS assessment of 0, 1, or 2. If the answer to this question is "no" the reader will be asked to confirm for this case: a BI-RADS assessment category of 1 or 2, a Probability of Malignancy (POM) score following the POM guidance provided in the protocol, and recall decision of "no". If the reader answers "yes" to whether there are mammographic findings, the reader will be asked to confirm an initial BI-RADS assessment category of 0, and will then provide detailed information on each suspicious finding (up to three findings) and the decision whether or not to recall the subject, as well as overall POM score and BI-RADS assessment category.

  3. Compare Per-subject Average Sensitivity for DBT Plus S-View Versus FFDM

    Time frame: Four weeks

    Based on forced BI-RADS scores requiring correct lesion localization, using non-inferiority margin delta = 0.10. The reader will first record whether there are mammographic findings and an initial BI-RADS assessment of 0, 1, or 2. If the answer to this question is "no" the reader will be asked to confirm for this case: a BI-RADS assessment category of 1 or 2, a Probability of Malignancy (POM) score following the POM guidance provided in the protocol, and recall decision of "no". If the reader answers "yes" to whether there are mammographic findings, the reader will be asked to confirm an initial BI-RADS assessment category of 0, and will then provide detailed information on each suspicious finding (up to three findings) and the decision whether or not to recall the subject, as well as overall POM score and BI-RADS assessment category.

Sponsors and collaborators

Lead sponsor

Fujifilm Medical Systems USA, Inc.

Industry

Registry information

Official study title

A Multi-Reader Multi-Case Controlled Clinical Trial to Evaluate the Comparative Accuracy of the Fujifilm DBT Plus S-View Versus FFDM Alone in the Detection of Breast Cancer - A Pivotal Study

Important dates

Study start
2018
Primary completion
2018
Study completion
2018
First posted
Apr 18, 2018
Registry last updated
Sep 21, 2023

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.