A Follow-up Study Validating a Blended Technique for Identifying the Proximal Humerus Intraosseous Vascular Access Insertion Site
Looking for future studies?
Notify Me21 year and older
All sexes
Interventional
Not applicable
Spring Branch, Texas, 78070, United States
In 2012, the investigators from Vidacare evaluated 4 techniques commonly used to identify the proximal humerus IO insertion site to determine if one technique resulted in more consistent successful placement of the IO needle set with a higher level of confidence among device operators performing the techniques. Results suggest that a combination of the best features of the 4 techniques may lead to optimal IO needle set placement more consistently. This study is needed to validate the "Blended" proximal humerus IO insertion technique to confirm that use of the technique results in proper site identification among device operators and results in increased confidence.To most closely evaluate the Blended proximal humerus insertion technique and how it is received by clinicians who perform these procedures in actual patients, device operators will include licensed/certified clinicians including, emergency medicine technicians/paramedics and nurses.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Device Operators:
Exclusion criteria
for Device Operators: None
Inclusion criteria
Healthy Subjects:
Exclusion criteria
Health Subjects:
Intraossesous Vascular Access in the proximal humerus
Other names: EZ-IO, Proximal Humerus IO Access, IO
Time frame: Day 1 Within 10 minutes following procedure
X-ray images of the inserted IO needle will be used to evaluate placement success and scored. Each insertion is scored by a musculoskeletal radiologist for two measures: Insertion Site Accuracy (scale range 0 - 3) and Needle Tip Location (scale range -7 to 7) as described below.
The Total score is the sum of Insertion Site Accuracy Score and Needle Tip Location Score. Scale range -7 to 10; higher score indicates greater accuracy.
Insertion Site Accuracy Score Scale is 0 to 3 where higher scores indicate greater accuracy: 3=center of greater tubercle, 2=perimeter of greater tubercle, 1=junction of greater tubercle and shaft or head, and 0=outside the greater tubercle.
Needle tip Location Score Scale ranges from -7 to 7 where higher scores indicate more favorable needle tip location: 7=within center of proximal humerus (PH), 5=within PH not centered, 3= Near miss of PH, -3=In humeral head, -3=on epiphyseal line, -5=lateral miss (of the humerus), -7=medial miss (of the humerus).
Time frame: 1 day, within 10 minutes of procedure
The amount of time it takes for the device operator to identify the proximal humerus intraosseous (IO) insertion site using their assigned proximal humerus IO site identification method and insert the IO needle. Timing starts from the time the participant begins to palpate the proximal humerus and stops when the IO needle is inserted.
Time frame: Day 1 within 10 minutes of completing training
Device operators scored their perceived Level of Confidence with their assigned Intraosseous (IO) Insertion method. The score was reported following the completion of training but before performing the unilateral IO insertion. The Level of Confidence scale scores range from 1 to 5, where. 1=not at all confident, 2=slightly confident, 3=moderately confident, 4=very confident, and 5=extremely confident. Higher scores indicate greater level of confidence.
Time frame: Day 1 within 10 minutes of completing the unilateral intraosseous insertion
Device operators scored their perceived Level of Confidence with their assigned Intraosseous (IO) Insertion method for their one unilateral proximal humerus IO site procedure. The score was reported after performing the unilateral IO insertion. The Level of Confidence scale scores range from 1 to 5, where. 1=not at all confident, 2=slightly confident, 3=moderately confident, 4=very confident, and 5=extremely confident. Higher scores indicate greater level of confidence.
Time frame: Day 1 within 10 minutes of completing the unilateral intraosseous insertion
Device operators will score their perceived Ease of Use for their assigned Intraosseous Insertion method for their one unilateral proximal humerus IO site procedure. The score was reported following the completion of training but before performing the unilateral IO insertion. The Ease of Use scale scores range from 1 to 5, where 1=very difficult, 2=difficult, 3=neutral, 4=easy, and 5=very easy. Higher scores indicate greater ease of use.
Time frame: Day 1 within 10 minutes of completing the unilateral insertion
Device operators will score their perceived Ease of Use for their assigned Intraosseous Insertion method for their one unilateral proximal humerus IO site procedure. The score was reported after completing the unilateral intraosseous insertion. The Ease of Use scale scores range from 1 to 5, where 1=very difficult, 2=difficult, 3=neutral, 4=easy, and 5=very easy. Higher scores indicate greater ease of use.
Time frame: Day 1 after establishing IO vascular access
Intraosseous infusion flow rates attained at tested infusion pressures, which may include Gravity; 100 mmHg, 200 mmHg and 300 mmHg.
Time frame: Day 1 after IO insertion
Time in seconds for fluid delivery from the proximal humerus to the heart, using visualization of contrast injection under fluoroscopy was performed and measured once for the participants in the Healthy Subjects: Blended Technique arm.
Time frame: Day 1 within 30 minutes after establishing proximal humerus IO vascular access and peripheral venous access
Number of Participants with Laboratory Results Successfully Obtained from both intraosseous (IO) and peripheral venous blood specimens. Success was determined if the point-of-care laboratory analysis system provided laboratory results values for both IO and Peripheral Venous specimens. It was deemed unsuccessful if the point-of-care system provided no laboratory results values and/or noted an error for at least one specimen. An intraosseous blood specimen and a peripheral venous blood specimen were collected from each subject
Vidacare Corporation
Industry
A Follow-up Study Validating a Blended Technique for Identifying the Proximal Humerus Intraosseous Vascular Access Insertion Site
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT01866514
Intraosseous Vascular Access
Spring Branch, Texas, United States
View Trial DetailsNCT01853215
Intraosseous Vascular Access
Spring Branch, Texas, United States
View Trial DetailsNCT01866475
Intraosseous Vascular Access
San Antonio, Texas, United States
View Trial DetailsNCT01742780
Vascular Access
Spring Branch, Texas, United States
View Trial Details