Sansum Diabetes Research Institute
Santa Barbara, California, 93105, United States
NCT Number: NCT01472406
This clinical trial is a feasibility study to assess the performance of an Artificial Pancreas (AP) device using the Artificial Pancreas System (APS©) platform for subjects with type 1 diabetes. The device is a closed-loop between a DexCom™ SEVEN® PLUS (DexCom™ Corp, San Diego, CA) continuous glucose monitor (CGM) and a OneTouch® Ping® Glucose Monitoring System (Animas Corp, Westchester, PA) subcutaneous insulin delivery pump (CSII). The AP device is controlled by a zone-Model Predictive Control (zone-MPC) algorithm augmented by a safety algorithm named the Health Monitoring System (HMS). The clinical study will include 12 to 20 adults subjects aged 21 to 65 years old.
Looking for future studies?
Notify Me18 year–65 year
All sexes
Interventional
Early Phase 1
Santa Barbara, California, 93105, United States
The study consists of an evaluation of the Artificial Pancreas device during a 24-hour closed-loop in a clinic environment (Sansum Diabetes Research Institute, Santa Barbara, CA). The 24-hour period includes:
The proposed study will evaluate the performance of the AP device in predicting the fall and rise of glucose values and in regulating insulin delivery to mitigate extreme blood glucose variations during the following challenges:
The goal is to demonstrate that the AP device is able to maintain the subject blood glucose within a safe range at all times. One of the objectives of the study will be to measure the percent of time spent in the following zones:
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The Sansum Closed-loop Artificial Pancreas Device is the infrastructure that allows communication between the insulin pump, the CGM, the zone-MPC, and the HMS.
Other names: Sansum AP
The insulin pump, which has received 510k clearance (K080639), delivers insulin subcutaneously through an infusion set.
The CGM, which is an approved device (P050012), measures interstitial glucose.
Other names: CGM
The zone-MPC predicts future glucose values, based on CGM data, to regulate blood glucose level by increasing, decreasing, or stopping insulin infusion via an insulin infusion pump.
Other names: zone-MPC
The HMS predicts future glucose values, based on CGM data, to warn the Sansum Closed-loop Artificial Pancreas Device of values below 70 mg/dL.
Other names: HMS
Time frame: 24 hours
Measures ability of AP device to maintain glucose level within a safe zone and to return to save zone following unannounced meals and exercises.
Time frame: 24 hours
Measures ability of AP device to maintain glucose level with a safe zone during the night and prevent hypoglycemia
Time frame: 24 hours
Measures ability of AP device to react to small meal and to prevent glucose level from dangerous excursions.
Time frame: 24 hours
Measures ability of AP device to maintain safe glucose level during 30 minutes of non-strenuous exercise
Time frame: 24 hours
Measures ability of AP device to address exercise and to prevent glucose level from dangerous excursions
Sansum Diabetes Research Institute
Other
Artificial Pancreas Device Feasibility Study for Type 1 Diabetes Patients Using Model-predictive Control and Health Monitoring System Algorithms With an Approved Subcutaneous Insulin Delivery Pump and Subcutaneous Continuous Glucose Monitor
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07696039
Autoimmune Diseases, Diabetes Mellitus
Bergamo, Italy
View Trial DetailsNCT03263494
Autoimmune Diseases, Diabetes Mellitus
Los Angeles, California, United States
View Trial DetailsNCT07090824
Autoimmune Diseases, Diabetes Mellitus
Palo Alto, California, United States
View Trial DetailsNCT05211479
Adolescent, Anxiety Disorders
Yalova, Turkey (Türkiye)
View Trial Details