Hvidovre Hospital
Hvidovre, 2650, Denmark
NCT Number: NCT01770561
The purpose of this study is to collect performance data on the integrated sensor and infusion set during 3 days of use and demonstrate performance adequacy of the Integrated sensor and infusion set with glucose sensing and insulin delivery functionalities. The hypothesis of this study is that having the infusion of insulin in close proximity to the sensor does not interfere with the sensor accuracy.
Looking for future studies?
Notify Me21 year–70 year
All sexes
Interventional
Not applicable
Hvidovre, 2650, Denmark
Sensor clinical accuracy will be evaluated based on the following parameters: Mean ARD (absolute relative difference), median ARD, Clarke Grid Analysis. Meter BG values will be used as reference. Descriptive statistics will be used to assess non-inferiority of the Integrated sensor and infusion sets sensors compare to control Enlite sensors.
Insulin delivery functionality will be evaluated based on BG profiles obtained during peri-prandial periods during which BG measurements will be performed at 15, 20, 30 min intervals according to a SMBG schedule. Post-prandial pharmacodynamic profiles will be analyzed to demonstrate the effect of insulin delivered using the Integrated sensor and infusion set. Specifically, rise in glucose levels followed by their fall will be used as an indication that insulin was delivered and absorbed. Post-prandial BG areas under the curve (AUC), Tmax (time to maximum rise of BG), and Cmax (maximum post-prandial BG) values will be calculated for each meal (breakfast, lunch, and dinner) consumed during both inpatient days and compared. Intra - and inter-subject mean values and standard deviations will be calculated. Paired and unpaired T-tests will be used to evaluate statistical significance.
Longevity of device performance will be assessed by comparing the listed parameters obtained during two inpatient days.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Change patients from current devices to the Integrated sensor and infusion set.
Time frame: 4 months
Sensor clinical accuracy will be evaluated based on the following parameters: Mean ARD (absolute relative difference), median ARD, Clarke Grid Analysis. Meter BG values will be used as reference. Descriptive statistics will be used to assess non-inferiority of the Integrated sensor and infusion set sensors compare to control Enlite sensors.
Time frame: 4 months
Insulin delivery functionality will be evaluated based on BG profiles obtained during peri-prandial periods during which BG measurements will be performed at 15, 20, 30 min intervals according to the SMBG schedule. Post-prandial pharmacodynamic profiles will be analyzed to demonstrate the effect of insulin delivered using the Integrated sensor and infusion set. Specifically, rise in glucose levels followed by their fall will be used as an indication that insulin was delivered and absorbed. Post-prandial BG areas under the curve (AUC), Tmax (time to maximum rise of BG), and Cmax (maximum post-prandial BG) values will be calculated for each meal (breakfast, lunch, and dinner) consumed during both inpatient days (visits 3 and 4) and compared. Intra - and inter-subject mean values and standard deviations will be calculated. Paired and unpaired T-tests will be used to evaluate statistical significance.
Time frame: 4 months
Longevity of device performance will be assessed by comparing the listed parameters obtained during visits 3 and 4.
Time frame: 4 months
Skin condition after removal of the device will be evaluated and assessed by the clinical staff in a study survey
Time frame: 4 months
Investigational Center and subject's report on physical duration of the device assessed in study survey Subjects' experiences are gathered in study surveys.
Medtronic Diabetes R&D Denmark
Industry
Feasibility Study to Assess Performance of an Integrated Sensor and Infusion Set. TRIAL II
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03529188
Autoimmune Diseases, Diabetes Mellitus
Brest, France
View Trial DetailsNCT02546401
Autoimmune Diseases, Diabetes Mellitus
Brest, France
View Trial DetailsNCT06529705
Autoimmune Diseases, Diabetes Mellitus
Montpellier, France
View Trial DetailsNCT06365255
Autoimmune Diseases, Diabetes Mellitus
Angers, France
View Trial Details