NCT Number: NCT02993601

Feasibility Study of a New Peripheral Oedema Monitor for Heart Failure

The study aims at assessing the usefulness of the Heartfelt1 device in patients suffering from heart failure, who have peripheral oedema. The device is non-contact and does not requires patient compliance to work. It gives an output reading of the patient's lower leg volume.

During the study, this measurement will be compared to clinical standard and gold standard measurements as well as clinical grading of peripheral oedema.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–99 year

Sex eligibility

All sexes

Study type

Observational

Primary location

Royal Brompton Hospital

London, SW3 6NP, United Kingdom

About this study

The Heartfelt-1 device is a CE (Conformité Européenne = European Conformity) mark medical device used to measure biometric parameters of the foot and lower leg. The study aims at assessing the usefulness in patients suffering from heart failure, who have peripheral oedema. Models used for the measurements of foot and lower legs have not yet been validated for this group of patients, who are likely to have feet with abnormal surface contours and appearance secondary to oedema. Peripheral oedema can be a leading indicator of heart failure de-compensation which can lead to hospitalisation of patients, and therefore the device could be a useful clinical tool to monitor patients in their home. The Heartfelt-1 device has the ability to recognise the patient's feet (amongst an average size household setting), and trigger automatically the measurements when the feet are naked (for example as the patient goes in and out of bed). This makes the device an obvious choice for patients with compliance issues with current methods of monitoring (e.g. daily use of weighing scales) and treatment (e.g. diuretics) if it could be validated in this patient group.

The study has two general aims:

The first is to collect a time-series of data on inpatients with peripheral oedema caused by heart failure, as they receive intravenous or oral diuretic treatment to remove excess fluid and achieve optimal body weight and fluid balance. This data will be used both to determine the attainable resolution of foot biometric measurements on a patient as their fluid status changes, and also to further develop the algorithms used in the device and make them usable on a wide range of feet/lower legs which have been deformed by oedema. The data required for this study are images taken using the Heartfelt-1 device, as well as validation measurements such as the figure of eight using a medical grade tape measure, callipers, or the water displacement method using a volumetric gauge, at multiple time-points whilst the peripheral oedema reduces.

The second, on a second group of stable ambulatory (Heart Failure) HF patients with variable levels of oedema, is to evaluate the performance of the Heartfelt-1 device in determining foot biometric parameters one a wide range of patients at a single point in time. Patients with stable chronic heart failure will have their foot/lower legs measured by the Heartfelt-1 device, as well as standard medical measurement (figure of eight using a medical grade tape measure, callipers) and gold standard measurement (water displacement method using a gauge), as well as the patient's weight and a clinical grading of oedema. These will be compared and an estimate of the accuracy of the Heartfelt-1 device over this wider patient group can be extrapolated.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Inpatients with admission for de-compensated heart failure requiring increase or initiation of diuretic therapy for fluid congestion.
  • Patients attending cardiology outpatients at the Royal Brompton hospital with a diagnosis of heart failure and fluid retention.
  • Cardiology patient without fluid retention (cardiology controls) Non-Cardiology patient without fluid retention (controls)

Exclusion criteria

  • Patients unable to provide informed consent.
  • Unstable arrhythmias in previous 48 hours
  • Systolic BP<90mmHg or need for inotropic support
  • Ulceration on lower limbs (below shin) - excluded from water displacement but could have image recording and calipers
  • Inability to transfer from bed to chair
  • Patients with broken skin or cellulitis will not have their foot volume measured using the water displacement equipment to reduce the risk of infection for the patient and other participants.

Treatment and study plan

Heartfelt-1

Device

The patients will stand in front of the Heartfelt-1 in order for images to be captured and the volume of their lower legs measured.

Other names: HF-1

Weighing scales

Device

The patient will be asked to stand on standard medical weighing scales, to measure their weight.

Medical grading of oedema

Other

A member of the research team will assess the severity of the peripheral oedema on a scale of 1-5.

Baseline assessements

Other

At the point study participants are recruited, their age, sex, diagnosis, aetiology of heart failure, LVEF (Left ventricular ejection fraction), medical and device therapy, time since diagnosis, BNP (B-type Natriuretic Peptide), CRP (C-Reactive Protein), renal function (EGFR), Haemoglobin, albumin and urinalysis will be assessed unless recent (less that 6 months old) results are available. The number of Heart failure hospital admissions in the last 2 years, LACE score, co-morbidities will also be noted.

Water displacement method

Device

This will involve filling a gauge with lukewarm water (27-35oC +-2oC) up to the appropriate level. The appropriate level is determined by the water stabilising after some has flown out of the gauge into the waste bucket. Once the water level is stable the patient, who is sitting on a chair will be gently plunging the foot (naked) in the gauge until their foot is flat on the bottom on the gauge. The water flowing out of the gauge into the measuring bucket will be collected until the flow of water stops. Four small marks with non-toxic /non-permanent pen will be made on the lower leg of the patient to indicate where the water level was at. The water which has been collected in the measuring bucket will be weighted with precision scales to determine the volume of water displaced by the insertion of the foot. The patient's foot will then be dried.

Callipers measurements

Device

This will involve making 4 marks on one of the foot/lower leg (small cross) using non-toxic / non- permanent pen. The measurement will consist of making 6 measurements between those marks using callipers, and reporting the distance on a calibrated tape.

Measuring tape

Device

The measurements taken using a medical tape measure will involve the circumference of the ankle to be measured as well as the circumference at the level of the arch of the foot. The medical tape used will be a CE marked disposable medical measuring tape to minimise the risk of infection. The measurements will be done on both feet.

Primary outcomes

  1. Volume as measured by the Heartfelt-1 device

    Time frame: Within 3 months after patient recruitment or before the end of the study (whichever is soonest).

Secondary outcomes

  1. Weight

    Time frame: Within 3 months after patient recruitment or before the end of the study (whichever is soonest).

    Patient weight on the day

  2. Distance measurements

    Time frame: Within 3 months after patient recruitment or before the end of the study (whichever is soonest).

    Distance measurements on the foot / lower leg of the patient

  3. Circumference and "figure of eight"

    Time frame: Within 3 months after patient recruitment or before the end of the study (whichever is soonest).

    Lower leg / foot circumference and "figure of eight" measurement

  4. Volume of the lower leg/foot

    Time frame: Within 3 months after patient recruitment or before the end of the study (whichever is soonest).

    As measured using the water displacement method.

  5. Medical grading of oedema

    Time frame: Within 3 months after patient recruitment or before the end of the study (whichever is soonest).

    Medical grading of peripheral oedema on a scale 1-5

Sponsors and collaborators

Lead sponsor

Heartfelt Technologies

Industry

Collaborators

  • Royal Brompton & Harefield NHS Foundation Trust

Registry information

Official study title

Evaluation of the Safety and Feasibility of a New Peripheral Oedema Monitor for Home and Clinical Use for Heart Failure Patients, Using the Heartfelt-1 Device

Important dates

Study start
2017
Primary completion
2017
Study completion
2017
First posted
Dec 15, 2016
Registry last updated
Oct 13, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.