Feasibility Study for GORE® TAG® Thoracic Branch Endoprosthesis to Treat Proximal Descending Thoracic Aortic Aneurysms

The purpose of this study is to assess the feasibility of the use of the GORE® TAG® Thoracic Branch Endoprosthesis to treat aneurysms involving the proximal Descending Thoracic Aorta (DTA)

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Leland Stanford Junior University, Stanford, California, United States

Loading trial locations.

About this study

The GORE® TAG® Thoracic Endoprosthesis (TAG® Device) received premarket approval (PMA) for use in endovascular aneurysm repair of the descending thoracic aorta (DTA) on 23-Mar-2005 under P040043, and design changes that resulted in the conformable GORE® TAG® Thoracic Endoprosthesis (CTAG Device) received premarket approval for the treatment of aneurysms of the DTA on 23-Aug-2011 (P040043/S039). The TAG® Device and CTAG Device are intended to exclude an aneurysm from the blood circulation in patients diagnosed with DTA aneurysms. However, endovascular-only treatment options for patients with aneurysms approaching the aortic arch are limited as current stent graft technology would require coverage of the left subclavian artery (LSA) which may result in the need for a surgical revascularization procedure. This prompted the creation of the GORE® TAG® Thoracic Branch Endoprosthesis (previously known as Branched TAG® Device), which was evaluated in this study under IDE G130120 for Zone 2 aneurysms of the DTA.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Presence of DTA aneurysm deemed to warrant surgical repair which requires proximal graft placement in Zone 2:
  • Fusiform (≥ 55 mm), or
  • Fusiform (>2 times native aortic diameter), or
  • Saccular (no diameter criteria)
  • Age ≥18 years at time of informed consent signature
  • Subject is capable of complying with protocol requirements, including follow-up
  • Informed Consent Form (ICF) is signed by Subject or legal representative
  • Must have appropriate proximal aortic landing zone, defined as:
  • Must require coverage of the left subclavian artery (LSA) origin for exclusion of the lesion
  • Aortic inner diameters between 16-48 mm
  • Landing zone, which must include the LSA ostium, cannot be aneurysmal, heavily calcified, or heavily thrombosed,
  • Acceptable proximal landing zone outer curvature length for the required device
  • Landing zone must be native aorta
  • Must have appropriate distal aortic landing zone, defined as:
  • Outer curvature length must be ≥2cm proximal to the celiac artery
  • Aortic inner diameters between 16-48mm (diameter should be between 16-42mm if using distal TAG® Device extension)
  • Landing zone cannot be aneurysmal, heavily calcified, or heavily thrombosed
  • Native aorta or previously implanted GORE® TAG® Device
  • Must have appropriate LSA landing zone, defined as:
  • LSA length of ≥3 cm proximal to first major branch vessel if using Aortic Component with 8 mm portal diameter, or LSA length of ≥2.5 cm proximal to first major branch vessel if using Aortic Component with 12 mm portal diameter
  • LSA inner diameters of 6-15 mm if using Aortic Component with 8 mm portal diameter, or inner diameters of 11-18 mm if using Aortic Component with 12 mm portal diameter
  • Target branch vessel landing zone cannot be aneurysmal, heavily calcified, or heavily thrombosed

Exclusion criteria

  • Concomitant aneurysm/disease of the ascending aorta, aortic arch, or abdominal aorta requiring repair
  • Previous endovascular repair of the ascending aorta
  • Previous endovascular repair of the DTA with a non-Gore device
  • Surgery within 30 days of treatment
  • Infected aorta
  • Dissection of the DTA
  • Intramural hematoma of the DTA without DTA aneurysm
  • Life expectancy <2 years
  • Myocardial infarction or stroke within 6 weeks prior to treatment
  • Patient has a systemic infection and may be at increased risk of endovascular graft infection
  • Pregnant female at time of informed consent signature
  • Degenerative connective tissue disease, e.g. Marfan's or Ehler-Danlos Syndrome
  • Participation in another drug or medical device study within one year of study enrollment
  • Known history of drug abuse within one year of treatment
  • Significant thrombus or atheroma in the aortic arch
  • Tortuous or stenotic iliac and/or femoral arteries preventing introducer sheath insertion and the inability to use a conduit for vascular access
  • Planned coverage of left carotid or celiac arteries
  • Patient has known sensitivities or allergies to the device materials
  • Patient has known hypersensitivity or contraindication to anticoagulants or contrast media, which is not amenable to pre-treatment
  • Previous instance of Heparin Induced Thrombocytopenia type 2 (HIT-2) or known hypersensitivity to heparin
  • Diameter taper outside of the device sizing range between proximal and distal landing zones of aorta and the inability to use additional devices of different diameters to compensate for the taper
  • Mycotic aneurysm
  • Persistent refractory shock (systolic blood pressure <90 mm Hg)
  • Patient has body habitus or other medical condition which prevents adequate visualization of the aorta

Additionally, the following exclusion criteria apply for Subjects enrolled in Stage 1:

  • Aberrant right subclavian or left vertebral arteries
  • Occluded/stenosed/hypoplastic right vertebral artery
  • Presence of a patent left internal mammary artery (LIMA) graft
  • Bilateral carotid artery disease
  • Known incomplete Circle of Willis
  • Known left vertebral artery ending in posterior inferior cerebellar artery (PICA)

Treatment and study plan

GORE® TAG® Thoracic Branch Endoprosthesis

Device

Primary outcomes

  1. Number of Participants With Successful Study Device Access

    Time frame: During treatment procedure (day 0)

    Access to the aneurysm and target landing zone location is obtained via conventional vascular access and endovascular techniques.

  2. Number of Participants With Successful Study Device Deployment

    Time frame: During treatment procedure (day 0)

    Absence of deployment failure will be considered a successful deployment. Deployment failure will be considered the failure of any Branched TAG® Device component (Aortic Component, Aortic Extender, or SB Component) to be released from the delivery catheter resulting in a serious adverse event (SAE) due to mechanical failure or use error.

  3. Number of Participants With Primary Procedural Side Branch Patency as Assessed by Angiography

    Time frame: At conclusion of the treatment procedure (day 0)

    The presence of forward flow through the implanted Side Branch Component into the target branch vessel.

Secondary outcomes

  1. Number of Participants With 1 Month Side Branch Primary Patency Assessed by an Independent Core Lab

    Time frame: 1 Month

  2. Number of Participants Without 1 Month Device Related Endoleaks Assessed by an Independent Core Lab

    Time frame: 1 Month

    Device-related endoelaks are defined as the presence of contrast within the aneurysm sac originating from the junction between any Branched TAG® Device component and the adjacent tissue (endoleak type IA or IB) OR the junction between the Aortic Component and either the SB Component or the Aortic Extender (type III endoleak).

Sponsors and collaborators

Lead sponsor

W.L.Gore & Associates

Industry

Registry information

Official study title

Evaluation of the GORE® TAG® Thoracic Branch Endoprosthesis in the Treatment of Proximal Descending Thoracic Aortic Aneurysms

Important dates

Study start
2014
Primary completion
2016
Study completion
2022
First posted
Dec 27, 2013
Registry last updated
Sep 9, 2022

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.