Skip to main content
OpenTrials
Completed

NCT Number: NCT05399914

Feasibility and Reliability of Synchronized Non-invasive Intermittent Positive Pressure Ventilation in Preterm Infants After Birth

The aim of this prospective observational study (pilot study) is to test if synchronized noninvasive positive pressure ventilation (S-NIPPV) is feasible and works appropriately during transition from intrauterine to extrauterine life in very preterm infants in the delivery room.

Completed

Looking for future studies?

Notify Me

Key information

About this study

The study period will be the first 10 minutes after birth.

All patients will be non-invasively ventilated using an EVE NEO ventilator (CE 0482, Fritz Stephan GmbH, Gackenbach, Germany). A facemask will be used throughout the study period. Synchronization will be achieved using graseby capsule (Stephan, Vio Healthcare), which will be secured at the abdominal wall on the midaxillary line below the ribs. For the detection of spontaneous breaths respiratory induction plethysmography (RIP) is used. For this purpose, an extra small plethysmography belt (Sleepsense, Elgin, USA) is placed around the abdomen. The data of the plethysmography belts are processed and stored with the NewLifebox-XL (Advanced Life Diagnostics UG, Weener, Germany) and merged with the data from the ventilator for later evaluation.

Infants according to the clinician's assessment only needing CPAP will receive S-NIPPV with very low PIP levels. This enables to have enough ventilator breaths applied for analysis to obtain sufficient data about the synchronization rate of S-NIPPV in the delivery room and meanwhile consider the infant's safety by not applying high pressures where it might not be needed. The ventilator support settings will be PIP 10, PEEP 6 in the noninvasive pressure controlled assist control ventilation (nPC-ACV+) mode. In this mode every spontaneous breath is supported by a ventilator breath (if the triggering level is exceeded). The triggering level will be set at 0,5 to avoid autotriggering and unsupported breaths. The inspiratory time will be set at max. 0,3 sec. The backup ventilation rate is set at 60 per minute.The EVE NEO ventilator has in the nPC-ACV+ mode an option where the ventilator breaths are interrupted/stopped as soon as the graseby capsule detects a start of spontaneous expiration.

If the infant according to the clinician's assessment needs S-NIPPV (e.g. the infant is apnoeic, and/or heart rate is below 100 bpm), the pressure of S-NIPPV will be augmented to target PIP 20, PEEP 6 (unless clinical concerns require an adjustment). The respiratory support can be deescalated to PIP 10 once the clinician deems the higher pressure no longer necessary.

The following demographical characteristics will be collected from the patient charts (pseudonymized): gender, gestational age, birth weight, antenatal steroids, mode of delivery, reason for prematurity (prolong premature rupture of membrane, infection, pathologic CTG, pathological doppler, gestosis), time of cord clamping, umbilical cord pH, FiO2 requirement, APGAR.

The following interventions within the first 10 minutes after birth including the time will be noted: tactile stimulation, placement of an intravenous line, non-invasive surfactant administration, intubation, chest compression, administration of volume, catecholamine, blood or other medication.

Who can participate

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Preterm infants with a gestational age (GA) between 26 ≥ and < 32 weeks

Exclusion criteria

  • Major congenital malformations
  • Refused parental consent (prenatal parental consent is necessary)

Treatment and study plan

S-NIPPV

Other

Synchronized noninvasive positive pressure ventilation during the first 10 minutes after birth.

Primary outcomes

  1. Synchronisation rate

    Time frame: First 10 minutes after birth

    Synchronisation rate (sensitivity/true positive rate) of S-NIPPV in the delivery room meaning the proportion of graseby triggered synchronized ventilator breaths with corresponding RIP signal related to all spontaneous breaths detected by RIP.

Secondary outcomes

  1. False negative rate

    Time frame: First 10 minutes after birth

    Proportion of non-synchronized ventilator breaths with corresponding RIP signal. This will be related to all spontaneous breaths detected by RIP.

  2. False positive rate

    Time frame: First 10 minutes after birth

    Proportion of graseby triggered ventilator breaths without corresponding RIP signal. This will be related to all ventilator breaths without a corresponding RIP signal (i.e. movement artefacts).

  3. Proportion of artifacts

    Time frame: First 10 minutes after birth

    Proportion of graseby triggered ventilator breaths without corresponding RIP signal. This will be related to all ventilator breaths.

  4. Specificity (true negative rate)

    Time frame: First 10 minutes after birth

    Proportion of ventilator breaths without corresponding RIP signal and with no graseby signal (i.e. during an apnoea). This will be related to all ventilator breaths without a corresponding RIP signal.

  5. Proportion of back up breaths

    Time frame: First 10 minutes after birth

    Proportion of ventilator breaths without corresponding RIP signal and with no graseby signal. This will be related to all ventilator breaths.

  6. Proportion of time of synchronised breaths, non-synchronized breaths, periods of apnoea and periods of movement artefacts occure within the first ten minutes of respiratory support.

    Time frame: First 10 minutes after birth

  7. Time until the graseby capsule is attached to the infant's abdomen and the proportion of infants in whom the graseby capsule is attached within the first two minutes after birth.

    Time frame: First 10 minutes after birth

  8. Proportion of time non invasive ventilator breaths will be applied.

    Time frame: First 10 minutes after birth

  9. Peak inspiratory pressure (PIP)

    Time frame: First 10 minutes after birth

  10. Positive end expiatory pressure (PEEP)

    Time frame: First 10 minutes after birth

  11. Mean airway pressure (MAP)

    Time frame: First 10 minutes after birth

  12. Td (time delay) graseby

    Time frame: First 10 minutes after birth

    Time between the initial increase of the abdominal RIP curve and the initial increase of the graseby curve.

  13. Td (time delay) ventilator

    Time frame: First 10 minutes after birth

    Time between the initial increase of the graseby curve and the beginning of the pressure curve.

  14. Ti ventilator

    Time frame: First 10 minutes after birth

    Inspiratory time of a ventilator breath

  15. Ti RIP

    Time frame: First 10 minutes after birth

    Inspiratory time of a spontaneous breath

  16. Ti excess

    Time frame: First 10 minutes after birth

    Calculated: ((Ti ventilator - Ti RIP)/Ti RIP)) x 100

  17. Swing RIP

    Time frame: First 10 minutes after birth

    Amplitude of the RIP curve calculated by subtracting the baseline RIP from the peak RIP curve, which is a surrogate parameter for the patient's respiratory effort.

  18. Number of tactile stimulations and other interventions on the patient

    Time frame: First 10 minutes after birth

    Number of tactile stimulations and other interventions on the patient (e.g. placement of intravenous line, non-invasive administration of surfactant, intubation, chest compression) are documented by video analysis.

  19. Heart rate

    Time frame: First 10 minutes after birth

  20. Oxygen saturation

    Time frame: First 10 minutes after birth

  21. Fraction of inspired oxygen

    Time frame: First 10 minutes after birth

Sponsors and collaborators

Lead sponsor

University Hospital Tuebingen

Other

Registry information

Acronym: SenSyNoPP

Important dates

Study start
2022
Primary completion
2023
Study completion
2023
First posted
Jun 1, 2022
Registry last updated
Sep 23, 2024

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.