NCT Number: NCT02321514

Expanded Clinical Study of the Tendyne Mitral Valve System

The purpose of this study is to evaluate the safety and performance of the Tendyne Mitral Valve System in the treatment of severe mitral regurgitation in patents with functional disability greater than or equal to NYHA Class II, who are not suitable candidates for surgical replacement with otherwise available devices. Follow-up evaluations will be conducted through 5 years post implantation.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Flinders Medical Center, Bedford Park, Australia

Loading trial locations.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Severe mitral regurgitation of primary or secondary etiology according to MVARC (Mitral Valve Academic Research Consortium) 2015 defined as:
  • For Degenerative MR: EROA ≥ 40 mm^2 or regurgitant volume ≥ 60ml
  • For Secondary MR: EROA ≥ 20 mm^2 or regurgitant volume ≥ 30ml
  • New York Heart Association (NYHA) functional Class ≥ II while on guideline directed medical therapy (GMDT), including device therapy (CRT) if indicated.
  • Heart team determines patient is not a suitable candidate for traditional surgical treatment according to valid guidelines.
  • Age 18 years or older.

Exclusion criteria

  • Severe mitral annular calcification, severe mitral stenosis, valvular vegetation or mass.
  • Left Ventricle (LV) or Left Atrium (LA) thrombus.
  • Patient has a chest condition that prevents transapical access.
  • Left ventricular ejection fraction (LVEF) less than 30% by echocardiogram.
  • Left Ventricular End Diastolic Diameter (LVEDD) > 7.0 cm.
  • Prior surgical or interventional treatment of mitral or aortic valves (e.g. valve repair or replacement, MitraClip, edge to edge repair, aortic balloon valvuloplasty, etc.).
  • Any planned surgery or interventional procedure within the period of 30 days prior to 30 days following the implant procedure. This includes any planned concomitant cardiovascular procedure such as PCI, pulmonary vein ablation, left atrial appendage occlusion, septal defect repair, etc.
  • Cardiac resynchronization therapy device or implantable pulse generator implanted within three months of planned implant procedure.
  • Myocardial Infarction (MI) within 30 days of the planned implant procedure.
  • Symptomatic, unresolved multi-vessel coronary artery disease (CAD) or unprotected left main coronary artery disease requiring stenting or Coronary Artery Bypass Grafting (CABG).
  • Cerebrovascular accident (CVA) within six months of planned implant procedure.
  • Unresolved severe symptomatic carotid stenosis (> 70% by ultrasound).
  • Cardiogenic shock or hemodynamic instability requiring inotropes or mechanical support devices at the time of planned implant procedure.
  • Severe tricuspid regurgitation, tricuspid valve disease requiring surgery or severe right ventricular dysfunction.
  • Hypertrophic or restrictive cardiomyopathy, constrictive pericarditis or any other structural heart disease causing heart failure other than dilated cardiomyopathy of either ischemic or non-ischemic etiology.
  • Any of the following: leukopenia, acute anemia, thrombocytopenia, history of bleeding diathesis, or coagulopathy if cannot be adequately treated.
  • History of endocarditis within six months of planned implant procedure.
  • Active systemic infection requiring antibiotic therapy.
  • Known hypersensitivity or contraindication to procedural or post-procedural medications (e.g., contrast solution, anti-coagulation therapy) which cannot be adequately managed medically or hypersensitivity to nickel or titanium.
  • Patient is undergoing hemodialysis due to chronic renal failure.
  • Patient has pulmonary arterial hypertension (fixed PAS >70mmHg).
  • Patient has COPD and is on home oxygen.
  • Patient refuses blood transfusions.
  • Pregnant, lactating, or planning pregnancy within next 12 months.
  • Participating or planning participation in an investigational drug or another device study.
  • Patient or legal guardian unable or unwilling to give informed consent.
  • Patient unable or unwilling to comply with study required testing and follow-up visits.
  • Patients with non-cardiac co-morbidities that are likely to result in a life expectancy of less than one year.

Treatment and study plan

Tendyne Mitral Valve System

Device

Patients will undergo transcatheter mitral valve replacement using Tendyne Mitral Valve system

Primary outcomes

  1. Safety Endpoint: Number of Participants With Composite of Device Success and Freedom From Device or Procedure Related Serious Adverse Events (SAEs)

    Time frame: 30 days post-index procedure

    Device success and freedom from the following device- or procedure-related serious adverse events (SAEs) at 30 days post the index procedure, will be classified by the Clinical Events Committee (CEC):

    • Cardiovascular death
    • Reintervention caused by valve-related dysfunction
    • Disabling stroke
    • Myocardial infarction (MI)
    • Life-threatening bleeding (BARC Type 2, 3, and 5)
    • Major Vascular Complications
    • Renal failure requiring dialysis
    • Other device-related SAEs
    • Other procedure-related SAEs
  2. Performance Endpoint: Number of Participants With MR Grade ≤ 2

    Time frame: 30 days post-index procedure

    Proportion of subjects with mitral regurgitation (MR) grade ≤ 2, per echocardiography core laboratory will be assessed.

    MR severity is graded on a scale of 0+ to 4+ where 0+ means absence of mitral regurgitation, 1+ is mild, 2+ is moderate, 3+ is moderate-to-severe and 4+ is severe.

Sponsors and collaborators

Lead sponsor

Abbott Medical Devices

Industry

Registry information

Important dates

Study start
2014
Primary completion
2020
Study completion
2025
First posted
Dec 22, 2014
Registry last updated
Jul 17, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.