NCT Number: NCT00399646

Evaluation of the Metricath Gemini System, a Device Intended to Make Arterial Measurements and Perform Angioplasty Dilations.

The purpose of the Metricath Gemini System is to pre-dilate the target treatment area, take arterial lumen measurements in a native section of artery or within the deployed stent and perform further dilation of the deployed stent if required. In this manner, the Metricath Gemini System can assist the Interventionalist by providing arterial measurements prior to stenting, performing pre-stenting angioplasty, taking within stent measurements to help determine if the stent is fully deployed, and by performing post-stenting dilations to further dilate the stent if required.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Sex eligibility

All sexes

Study type

Interventional

Phase

Phase 2

Primary location

St. Paul's Hospital, Vancouver, British Columbia, Canada

Loading trial locations.

About this study

Several recent clinical studies have suggested that proper vascular stent deployment directly affects clinical outcome, and the rate of re-stenosis. These studies suggest that angiography alone is not sufficient to ensure proper vascular stent deployment, and that re-stenosis rates will decline if proper stent apposition has occurred. The importance of proper stent sizing and apposition is further emphasised with the increasing use of drug eluting stents, the drug effects of which are only realized upon contact with the arterial wall.

Angiometrx Inc. developed the Metricath Arterial Measurement System in response to the need for increased stent sizing and deployment. The Metricath System measures the diameter and cross sectional area of arteries using an intravascular balloon catheter attached to a computerized console. Considering the ease and rapidity of obtaining Metricath results, this technique may form an alternative to evaluate vessel area and stent expansion.

The Metricath System was 510(k) cleared in the U.S. in June, 2003, and received European and Canadian market approvals in 2004.

Angiometrx Inc. has now developed the next generation of Metricath System, which incorporates a second balloon on the catheter, intended for angioplasty purposes. The Metricath Gemini System is comprised of a dual balloon catheter which is attached to the same Metricath computerized console as the Metricath System. As with the Metricath System, the Metricath Console operates the inflation and deflation of the Gemini measurement balloon and provides arterial cross-sectional area and diameter. The Metricath Gemini angioplasty balloon is controlled separately in the traditional manner using a hand-held inflation device, attached to a port in the hub of the catheter.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patients in whom coronary artery angiography is indicated and angioplasty and stenting is being considered.
  • Single vessel disease of an artery with a lesion assessable to standard coronary guidewires and are scheduled or indicated for percutaneous coronary intervention.
  • A maximum of one lesion requiring treatment (angioplasty +/- stenting) to an intended arterial diameter of 2.5 mm - 4.0 mm.
  • Target stent length of 15 mm - 32 mm, where a single stent can adequately cover the lesion.
  • Stenotic de novo lesions in native coronary arteries.

Exclusion criteria

  • Women who are pregnant.
  • Participation in a study involving investigational drugs or devices during the last six months.
  • Patients with previous stenting in the target vessel area.
  • Patients who experienced a myocardial infarction, Braunwald Class C, and very unstable Braunwald Class B3 within 6 weeks prior to the procedure or have unstable angina pectoris.
  • Patients who are experiencing TIA's or minor or major strokes within six months prior to procedure.
  • Contraindicated for antiplatelet and/or anticoagulation medications.
  • Symptoms of cardiogenic shock.
  • Patients who underwent a staged procedure 30 days prior to enrolment and/or were scheduled to have a staged procedure 14 days after treatment.
  • Significant liver or kidney disease or malignancy.
  • Patients scheduled for any other surgery or other procedure within 30 days.
  • Severe peripheral vascular disease preventing femoral access.
  • Left ventricle ejection fraction of < 35%.
  • Total occlusion of target vessel.
  • A blood pressure in excess of 180 mmHg at the time of the angiography.
  • Visible thrombus, filling defect, or ulceration in the target artery.
  • Severely calcified lesions which suggest that balloon pre-dilatation will not achieve adequate luminal diameter to allow successful stent delivery and deployment.
  • The target lesion is beyond a left main artery stenosis > 50%.
  • Contraindicated for CABG.
  • Coronary spasm in the absence of a significant stenosis.
  • Requiring treatment of more than two lesions.
  • Requiring treatment with atherectomy.
  • Lesions in surgical conduits - saphenous vein grafts, internal mammary arteries, or radial arteries.
  • Unprotected left main coronary artery.
  • Diffuse disease.
  • Excessive tortuosity of proximal segment.
  • Extremely angulated segments > 90°.

Treatment and study plan

Metricath Gemini System

Device

Primary outcomes

  1. To observe the safety of the device, as determined by freedom from MACE (major adverse cardiac events) within 30 days post-procedure.

    Time frame: 30 days

  2. To observe the success of the device as determined by residual diameter stenosis

    Time frame: 30 days

  3. To evaluate the inflation of dilation balloon to desired pressure

    Time frame: immediate

Secondary outcomes

  1. To evaluate the successful delivery of catheter to target area

    Time frame: immediate

  2. To evaluate the performance of the device to measure arterial size

    Time frame: immediate

Sponsors and collaborators

Lead sponsor

Shockwave Medical, Inc.

Industry

Registry information

Official study title

Gemini Angioplasty and Arterial Measurement Evaluation (GAAME) An Evaluation of the Metricath Gemini System, a Percutaneous Interventional Device Intended to Make Arterial Lumen Measurements and Perform Percutaneous Transluminal Angioplasty Dilations.

Acronym: GAAME

Important dates

Study start
2005
Primary completion
2007
Study completion
2009
First posted
Nov 15, 2006
Registry last updated
May 13, 2009

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.