Diagnostic TestBlood tube will be collected from standard venepuncture
NCT Number: NCT04375982
A Multicenter Study Conducted to Evaluate the Agreement between Fingerstick Whole Blood, Venous Whole Blood and Plasma Determined on the LumiraDx Point of Care D-Dimer and Point of Care CRP Tests to Results on the Reference Analyzer
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Rancho Paseo Medical Group, Banning, California, United States
The purpose of the study is to gather data to determine agreement between sample types, which together with performance evaluation data will be used to support CE marking under Directive 98/79/EC of the European Parliament and of the Council of 27 October 1998 on in vitro diagnostic medical devices 3, using ISO 13612:2002 - Performance Evaluation of In Vitro Diagnostic Medical Devices as the standard. The investigative devices are the LumiraDx Platform D Dimer Test and the LumiraDx Platform CRP Test. The LumiraDx Platform employs a portable diagnostic instrument used with single use D Dimer or CRP Assay Test Strips.
Agreement of the different sample types to the reference method will be demonstrated across a patient population representative of the indication for use of the products. Operators will be trained clinical site staff who are representative of the intended users of the product, i.e. health care professionals at the point of care such as nurses, technicians, doctors, etc. A panel of whole blood and plasma samples will be obtained from each patient in order to assess the accuracy of the investigative device across all sample types.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
D-dimer:
D-dimer Exclusion Criteria:
CRP Inclusion criteria:
CRP Exclusion Criteria:
Blood tube will be collected from standard venepuncture
Capillary blood will be collected by fingerstick
Time frame: 2 months
Verification that all sample types give an equivalent result when tested in the LumiraDx assays. Results generated from capillary blood and venous whole blood will be compared to those generated from plasma to ensure that all 3 sample types give equivalent values. Results will be compared by standard statisitical techniques which may include Regression analysis (e.g. Passing Bablok) or Bias analysis (e.g. Altman Bland)
LumiraDx UK Limited
Industry
A Multicenter Study Conducted to Evaluate the Agreement Between Fingerstick Whole Blood, Venous Whole Blood and Plasma Determined on the LumiraDx Point of Care D-Dimer and Point of Care CRP Tests to Results on the Reference Analyzer
Acronym: NOVEL-3
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05568966
Cardiovascular, Cardiovascular Diseases
Airdrie, Scotland, United Kingdom
View Trial DetailsNCT04657120
Cancer, Cardiovascular Diseases
Antwerp, Belgium
View Trial DetailsNCT04439383
Cardiovascular Diseases, Deep Vein Thrombosis
Bern, Switzerland
View Trial DetailsNCT05128591
Cardiovascular Diseases, Embolism and Thrombosis
Seoul, South Korea
View Trial Details