NCT Number: NCT04732351

Evaluation of the Active Sentry System During Cataract Surgery With the Centurion Phacoemulsifier

Each new generation of phacoemulsification system has aimed to optimize the control of fluidics and the delivery of energy during phacoemulsification. Nicoli CM et al., (J Cataract Refract Surg 2016;42:157-162) suggested a better anterior chamber stability when using an active fluidics system. Malik PK et al., (Eye and Vision 2017;4:22) showed that using an IOP based phacoemulsification system in association with the use of an Intrepid balanced tip provides a reduction in CDE (Cumulated Dissipated Energy and EFU (Estimated Fluid Usage). It has also been shown that a lower CDE provides obvious benefits: less endothelium cells damages and lower complication rates (Mady MA et al., Clin.Ophthalmol. 2012; 6:503-10).

The latest technologies, i.e. the Active Sentry Handpiece and Hybrid tips, have been introduced to further improve the safety of phacoemulsification. The Active Sentry system has moved pressure sensors closer to the operated eye, hence further reducing the risk of surge during procedures. (Thorne A et al. Phacoemulsifier occlusion break surge volume reduction, J Cataract Refract Surg. 2018;44:1491-1496). The Hybrid tips have been designed to prevent posterior capsule rupture in the event of contact with the capsule.

This new technology allows phacoemulsification procedures under reduced infusion pressures and a "high-vacuum - low energy" strategy which may lead to a safer procedure.

Data from this study will try to demonstrate that cataract surgeries with Active Sentry will need less CDE and therefore may be safer, due to the surge effects decrease. This CDE parameter will be analysed according to differents subgroups: Presence of hybrid tip or balanced tip, cataract score based on the WHO nuclear grading recommendation (Thylefors B et al., (2002) The WHO Cataract Grading Group, Ophthalmic Epidemiology, 9:2, 83-95) The recent experiences conducted in real world condition in different sites have suggested that Centurion with AS was able to improved settings: maximum vacuum and UltraSound (US) time reduction and as a consequence a 20 to 30% CDE reduction.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Ophtalmopôle, Hôpital Cochin

Paris, Île-de-France Region, 75014, France

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • All consecutive patients undergoing unilateral or bilateral cataract surgery with the Centurion phacoemulsifier, with the ability to give informed consent.
  • Patients will undergo the cataract surgery with or without the Active Sentry module and with or without hybrid tip according to centers practice.
  • Patients will undergo a cataract scoring (nuclear grade 1 to 3) prior surgery as described in the WHO recommendation:

Exclusion criteria

  • Patients with :
  • pupil dilation <5mm
  • zonular defects
  • corneal opacities

Treatment and study plan

Primary outcomes

  1. Mean CDE (Cumulated Dissipated Energy)

    Time frame: The day of the cataract procedure

    Mean CDE measured by the phacoemulsifier in different groups according to baseline patients and/or eyes characteristics (nuclear cataract grading etc...)

Secondary outcomes

  1. Mean total procedure duration

    Time frame: The day of the cataract procedure

    Mean total duration of each procedure in different groups

  2. Mean total Ultrasound time (US time)

    Time frame: The day of the procedure

    Mean total Ultrasound time (US time) in different groups

  3. Mean procedure settings

    Time frame: The day of the procedure

    Mean procedure settings : infusion pressure, aspiration flow and vacuum levels

  4. Mean BSS volume

    Time frame: The day of the procedure

  5. Frequency of adverse events

    Time frame: The day of the procedure

    Mean frequency of adverse events during procedures, including posterior capsule ruptures

Sponsors and collaborators

Lead sponsor

Assistance Publique - Hôpitaux de Paris

Other

Collaborators

  • URC-CIC Paris Descartes Necker Cochin

Registry information

Official study title

Observational Multicenter Registry to Assess the Enefits of Active Sentry Handpiece During Cataract Surgery With the Centurion Phacoemulsifier (SASCA)

Acronym: SASCA

Important dates

Study start
2021
Primary completion
2022
Study completion
2022
First posted
Feb 1, 2021
Registry last updated
Jan 15, 2026

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.