NCT Number: NCT02255214

Evaluation of the Accuracy of the Hemochron Signature Elite Point of Care Device Compared With Laboratory Assay

Central Neuraxial Blockade (CNB) is commonly used in anaesthetics. It has several advantages over a general anaesthetic however contraindications include coagulopathy. Guidelines suggest a CNB should not be undertaken with an International Normalised Ratio (INR) of 1.5 or greater. A common cause of an elevated INR is warfarin therapy and regular monitoring of its effects is required.

The Hemochron Signature Elite is a portable micro coagulation system that can be used as a point of care (POC) anticoagulation monitoring device. Marketing is focussed on its accuracy, convenience and efficiency in delivering a coagulation result.

This study will assess the accuracy of this device in calculating the International Normalised Ratio (INR) result. This will be compared to another INR result obtained on the same sample via the current gold standard testing equipment in the Royal Infirmary Edinburgh (RIE) laboratories.

The study will include patients presenting for surgery who have had their warfarin therapy stopped within the previous 7 days. Patients on warfarin therapy requiring surgery are currently identified in preoperative assessment clinics, on average 12 weeks prior to their date of surgery.

Anecdotally there appears to be a difference between the 'expected' and actual INR when a POC device is used. Concern exists that patients may be inappropriately denied the potential benefits of central neural blockade (CNB) if a POC device provides a falsely high result, or conversely the investigators may be proceeding with CNB in patients with an unacceptably high INR if POC provides a falsely low result. There is also the possibility of improving hospital efficiency using this device as it provides a result much faster than the laboratory.

Completed

Looking for future studies?

Notify Me

Key information

Sex eligibility

All sexes

Study type

Observational

Primary location

Royal Infirmary Edinburgh

Edinburgh, Lothian, EH16 4SA, United Kingdom

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Adult patients presenting for elective surgery who have stopped their warfarin therapy within the last 7 days.

Exclusion criteria

  • Adults with incapacity

Treatment and study plan

Primary outcomes

  1. The International Normalised Ratio (INR) Will be Measured by the Laboratory Assay and the Hemochron Signature Elite Device.

    Time frame: 18 months

    The correlation between the international normalised ratio measured by the laboratory assay and the Hemochron Signature Elite device was assessed.Number of participants with a higher INR measurement by the Hemochron Signature Elite device than laboratory assay

Sponsors and collaborators

Lead sponsor

NHS Lothian

Other Gov

Registry information

Official study title

Evaluation of the Accuracy of the Hemochron Signature Elite Point of Care Micro Coagulation Instrument Compared With Laboratory Assay

Important dates

Study start
2013
Primary completion
2015
Study completion
2015
First posted
Oct 2, 2014
Registry last updated
Jan 30, 2026

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.