Galilee Medical Center
Nahariya, 22100, Israel
NCT Number: NCT03964675
The SenseGuard™ is a new respiratory rate measuring and logging non-invasive device, and measures the respiratory rate by responding to changes in air moisture content in exhaled air. A prospective single-center, single arm study, designed to assess the bias, accuracy , precision and safety of the SenseGuard™ device in measuring of respiratory rate in adult patients. The study shall include a single invitational visit with a respiratory rate measurement duration of 30 minutes for each of 25 adult subjects hospitalized in internal wards.Subjects will be invited to participate should they meet all of the eligibility criteria. Following informed consent, subjects will be measured with both Microstream capnograph and SenseGuard™ measuring simultaneously so their results can be compared.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Nahariya, 22100, Israel
Study description:
Respiratory rate (RR) is a vital sign used to monitor clinical condition of a patient. It is defined as the number of breaths per minute. Various devices using different techniques are available to measure RR. Currently this is measured at the bedside in clinical practice by counting the breaths, however the gold standard for measuring this vital sign is the capnograph. For the patient, this involves limiting the movement and talking so that the measurements can be taken.
A new respiratory rate measuring and logging device, SenseGuard™, is non-invasive and measures the respiratory rate by responding to changes in air moisture content (small amounts of water condensation) in exhaled air.
The study is planned as a prospective single-center, single arm study, designed to assess the bias, accuracy , precision and safety of the SenseGuard™ non-invasive device in measuring of respiratory rate in adult patients.
Subjects will be invited to participate should they meet all of the eligibility criteria. Following informed consent, subjects will be measured with both Microstream capnograph and SenseGuard™ measuring simultaneously so their results can be compared.
Study population:
25 hospitalized patients from Internal/General wards aged 18 and older.
Objective:
Evaluate the performance and safety of the SenseGuard™ non-invasive device in measuring respiratory rate in hospitalized adult patients.
Inclusion criteria
Exclusion criteria
Participant Duration:
A single invitational visit with a respiratory rate measurement duration of 30 minutes
Study Duration:
1 year.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
A single invitational visit with a respiratory rate measurement duration of 30 minutes.
Time frame: 4 months
Time frame: 4 months
Time frame: 4 months
The easiness and comfort shall be evaluated using a questionnaire that includes 5 questions with possible scores ranging from 1(worst)-5(best)
NanoVation
Industry
Evaluation of SenseGuard™ - a Noninvasive Respiratory Rate Measuring Device for Adult Patients
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06226779
Behavior, Condition
Kaohsiung City, Taiwan
View Trial DetailsNCT05972187
Body Temperature Changes, Condition
Maastricht, Netherlands
View Trial DetailsNCT03393598
Condition, Disease
Basel, Switzerland
View Trial DetailsNCT05253664
Anxiety Disorders, Anxiety State
Kütahya, Turkey (Türkiye)
View Trial Details