PillCam™ (Capsule Endoscopy) Colon 2 capsule
DeviceMedical Device
NCT Number: NCT01087528
To evaluate the performance of PillCam COLON 2 in regards to detection of patients with polyps where colonoscopy is considered as the gold standard reference.
Looking for future studies?
Notify Me18 year–70 year
All sexes
Interventional
Not applicable
Indiana University Medical Center, Indianapolis, Indiana, United States
Standard evaluation of subjects with suspected colon diseases includes endoscopic imaging by colonoscopy and radiology testing such as: air-contrast barium enema, abdominal/pelvic CT, or virtual colonoscopy.
The Given® Diagnostic System offers an alternative approach for endoscopic visualization of the colon. Advantages of the Given® Diagnostic System include the elimination of the need for conscious sedation, the minimally invasive, painless nature of the exam, and the ability to pursue normal daily activities immediately following the procedure . Furthermore, compared to standard colonoscopy, the Given® Diagnostic System may be more readily accepted by the subjects, thereby improving subjects' willingness to undergo a diagnostic evaluation of the colon and comply with colorectal cancer screening recommendations.
The PillCam™ SB capsule (formerly M2A® Capsule) that was cleared by the FDA- in August 2001 for small bowel evaluation has been ingested to date by more than 1,000,000 people worldwide and is well accepted by patients and physicians as well as the professional societies. However, adequate visualization of the colon cannot be achieved with the standard PillCam™ SB capsule because of the anatomical and physiological properties of the colon which are significantly different than the small bowel. Moreover, other issues that limit the evaluation of the colonic mucosa by the standard PillCam™ SB procedure include an unsatisfactory level of colon cleanliness and slow progression of the PillCam™ SB capsule through the colon during the desired examination time. Therefore, the development and introduction of a specially designed, customized colon capsule combined with a dedicated capsule colonoscopy procedure protocol will allow for more efficient evaluation of the colonic mucosa. This is expected to improve the capability of the Given® Diagnostic System to detect colonic pathologies and to serve as a diagnostic and screening tool for colonic disease. To date, several clinical studies have been conducted with Given® Diagnostic System and the PillCam™ Colon 1 capsule A pilot multicenter study is currently being conducted with a new Given® Diagnostic System and an advanced generation of the the PillCam™ Colon capsule - PillCam™ Colon 2. Capsule. The new development of Given® Diagnostic System and PillCam™ Colon 2 capsule is primarily aimed to increase sensitivity and specificity for polyp detection.
Further details of the Given® Diagnostic System and PillCam™ Colon 2 capsule can be found in the device description section.
This study is designed to evaluate the performance of the new version of Given® Diagnostic System and PillCam™ Colon 2 capsule in detecting patients with polyps and other colonic lesions as compared to conventional colonoscopy.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Medical Device
Time frame: within 7 days
The accuracy parameters (i.e. sensitivity) were evaluated for detection of patients with polyps equal or larger than 6 mm and for detection of patients with polyps equal or larger than 10 mm by constructing 2X2 matrix (i.e. PCCE-2 performance versus the standard colonoscopy), where sensitivity and specificity, are presented with 95% Cl based on the binomial distribution. Standard colonoscopy is considered the gold standard.
Time frame: within 7 days
The accuracy parameters (i.e. specificity) were evaluated for detection of patients with polyps equal or larger than 6 mm and for detection of patients with polyps equal or larger than 10 mm by constructing 2X2 matrix (i.e. PCCE-2 performance versus the standard colonoscopy), where sensitivity and specificity, are presented with 95% Cl based on the binomial distribution. Standard colonoscopy is considered the gold standard.
Time frame: within 7 days
The diagnostic yield (%) of PCCE-2 and colonoscopy in visualizing a variety of colonic lesion excluding polyps (e.g. inflammation, diverticulosis and bleeding lesions) is provided.
Time frame: within 7 days
Overall colon cleanliness was judged for capsule endoscopy and colonoscopy on a four-point grading index scale as follows:
Time frame: within 7 days
Capsule excretion time is defined as the duration of ingestion to capsule excretion time. The distribution of excretion times were categorized as follows:
Medtronic - MITG
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03037385
Adenocarcinoma, Adenocarcinoma, Papillary
Phoenix, Arizona, United States
View Trial DetailsNCT02647866
Abdominal Pain, Autoimmune Disease
Greenville, South Carolina, United States
View Trial DetailsNCT03559543
Colonic Diseases, Colorectal Neoplasm
Havana, Cuba
View Trial DetailsNCT04263818
Colonic Diseases, Digestive System Diseases
Strasbourg, France
View Trial Details