The goal of this prospective controlled study is to measure the reliability of two intraocular pressure (IOP)-measuring instruments in comparison to the gold standard, Goldmann applanation tonometry (GAT) following vitreo-retinal surgery. The main question the current study aims to answer is: are Icare and Accupen as accurate as GAT in measuring IOP one day post eye surgery?
Participants in this study will have their IOP measured by the three different instruments one day post-surgery.
Interested in participating?
Request Info18 year and older
All sexes
Interventional
Not applicable
Québec, Quebec, G1S4L8, Canada
Location status: Recruiting
Imad E Hachem, MD
CONTACT
Mathieu Caissie, MD
CONTACT
Goldmann applanation tonometry (GAT), used since 1948, is the reference technique for measuring intraocular pressure (IOP) in both healthy patients and those suffering from glaucoma or other eye disorders. However, the use of GAT requires the topical application of fluorescein, local anesthesia, and direct contact with the cornea. In addition to its lack of portability, these characteristics of GAT quickly led clinicians to try to find simpler, less cumbersome, but just as reliable techniques to measure IOP. Among the techniques developed to potentially replace GAT, two portable techniques are particularly easy to use: rebound tonometry (Icare) and portable applanation tonometry (Accupen). Before these techniques are universally and interchangeably used with GAT, their reliability and accuracy must be measured. The majority of studies have evaluated these techniques in glaucomatous or healthy eyes, presenting mixed but promising results. However, only one recent study has evaluated the validity of the two techniques (Icare and TonoPen) in a post-operative context of vitreo-retinal surgery. More data is therefore necessary to come to a definitive conclusion regarding the reliability of Icare and Accupen.
The objective of this study is therefore to determine whether the IOP measured by Accupen and/or Icare is as accurate as that taken by GAT in patients who have undergone vitreo-retinal surgery.
To answer this question, this project will be a prospective controlled study of 67 patients who have undergone vitreo-retinal surgery and who will have their intraocular pressure taken 1 day post-operatively by the 3 instruments mentioned above, in a randomized order. At their 1-day post-operative visit, an ocular medical history will be taken, including the preoperative diagnosis and the type of surgery performed. A resident in ophthalmology at Laval University, will take a series of 2 intraocular pressures with these 3 devices: GAT, Icare, and Accupen and will average for each group.
The project research hypothesis would be that the taking of intraocular pressure by Icare and/or Accupen is as valid as that taken by Goldman's applanation tonometry.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The intraocular pressure of the subjects using either the Goldmann applanation tonometer, Icare or Accupen. The order of the measure is randomized.
Time frame: at the post-operative visit (1 day after surgery)
No difference in the intraocular pressure measured by Accupen or Icare compared to Goldmann tonometer.
Time frame: at the post-operative visit (1 day after surgery)
The devices have the same intraocular pressure measured by Goldmann tonometer at High level of pressure (>24mmHg), moderate level (10-24mmHg) and low level (<10mmHg).
Contact information is provided by the study sponsor or research team.
Julie Mauger, BSc
CONTACT
418-682-7511 ext. 84821
Sébastien Méthot, PhD
CONTACT
418-682-7511 ext. 84821
CHU de Quebec-Universite Laval
Other
Reliability Evaluation of Intraocular Pressure At Day 1 Following Vitreoretinal Surgery, Comparing Goldmann Applanation Tonometry (Gold Standard) to Icare and Accupen
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06275308
Intraocular Pressure
Pittsburgh, Pennsylvania, United States
View Trial DetailsNCT05716945
Congenital, Hereditary, and Neonatal Diseases and Abnormalities, Corneal Diseases
Apeldoorn, Gelderland, Netherlands
View Trial DetailsNCT04628663
Intraocular Pressure
Athens, Attica, Greece
View Trial DetailsNCT06016712
Delivery, Eye Diseases
Prague, Czechia
View Trial Details