CSF-3 is a wrist-worn medical device similar to a watch. It is designed to monitor a user's heart rate based on ECG (electrocardiogram) and PPG (Photoplethysmograph) readings, using off-the-shelf ECG and unique PPG chipsets with proprietary algorithms. ECG and PPG are the two primary technologies for measuring heart rate. ECG measures the bio-potential generated by electrical signals that control the expansion and contraction of heart chambers, while PPG uses light-based technology to sense the rate of blood flow as controlled by the heart's pumping action.
Looking for future studies?
Notify MeThe proposed clinical study is a prospective, open, single-center study to evaluate the performance and accuracy of the ECG and heart rate calculation of the CSF-3 investigational Device.
Subjects will be screened by the Principal Investigator. Subjects who meet the study's enrolment criteria will be recruited and fitted with an FDA cleared ECG Holter, and with the investigational Device (CSF-3).
At least 40 eligible subjects will be enrolled in a single center. Data collected from the subjects during the study will be included in the overall performance analysis of the ECG output. Subjects will perform an ECG and CSF-3 Device simultaneously Subjects will remain connected to both the ECG Holter and the CSF-3 Device at the study site for at least 7 minutes duration. The ECG and Holter will be stopped simultaneously. Both recordings data (ECG and CSF-3) will be uploaded and then synchronized offline using a dedicated and validated tool.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The subjects will be connected to an ECG Holter, and an investigational Device will be placed simultaneously on the subject's wrist. All subjects in this phase will be asked to participate in a single continuous data recording measurement for seven (7) minutes
Other names: FDA cleared ECG Holter
Time frame: 6 months
Performance endpoint is defined as the ability of CSF-3 to detect beat-by beat Heart Rate in ECG signal with a sensitivity of no less than 96%, considering a false detection rate of not higher than 2%, and PPV of 80%.
Time frame: 6 months
Performance endpoint is defined as the calculation of average Heart Rate with ARMS error<3BPM
CardiacSense Ltd.
Industry
A Prospective Study for Evaluating the Safety and Performance of CSF-3 ECG Lead-I, HR and PR When Compared to ECG Holter
Acronym: ECG
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06068257
Breast Cancer, Breast Diseases
Orlando, Florida, United States
View Trial DetailsNCT06702423
Healthy
London, United Kingdom
View Trial DetailsNCT07223164
Healthy
Irvine, California, United States
View Trial DetailsNCT05595902
Healthy
Kadıköy, Istanbul, Turkey (Türkiye)
View Trial Details