NCT Number: NCT06428799

Evaluation of Carebot AI MMG Medical Device for Breast Lesion Detection and Density Assessment

Comparison of accuracy of clinician and DLAD image evaluation (Carebot AI MMG v2.2)

1. Comparison of the Accuracy of Density Assessment by Clinician and DLAD (DENS) 2. Comparison of Accuracy of Lesion Assessment by Clinician and DLAD (MASS, CLASS)

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

Female

Study type

Observational

Primary location

EUC Mamocentrum Brno, Brno, Czechia

Loading trial locations.

About this study

The mammography studies were acquired from three independent sites: Site 1 (EUC Mamocentrum Brno) and Site 2 (Hospital Šumperk, a.s.) specialise in routine screening mammography, and Site 3 (Masaryk Memorial Cancer Institute) is a comprehensive oncology facility primarily dedicated to diagnostic mammography, i.e. performing additional examinations in case of a suspicious finding (recall).

The ground truth was obtained by consensus of two board-certified radiologists with expertise in radiology and diagnostic methods, and 13 and 27 years of experience with mammography image interpretation, respectively.

For comparative analysis, a team of five independent radiologists with clinical experience in interpreting mammography images was established. Three of the clinicians were junior (2, 2, and 4 years of experience, respectively) without board-certification; two physicians were senior (7 and 8 years of experience, respectively), board-certified.

Who can participate

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • The medical device is intended for use in women over 18 years of age who are indicated for screening mammography using digital mammography.

Exclusion criteria

  • The medical device cannot be used in patients with breast implants.
  • The medical device cannot be used in male breast examination.
  • The medical device cannot be used in patients under 18 years of age.

Treatment and study plan

Carebot AI MMG

Device

Carebot AI MMG is a software solution that utilizes artificial intelligence methods, specifically deep learning and computer vision algorithms, to evaluate and localize suspicious regions of potential lesions during the interpretation of digital breast x-rays as part of standard mammography screening procedures. The Carebot AI MMG medical device is not intended for use in diagnostic mammography. The Carebot AI MMG is intended for use in women over the age of 18.

The predictive outputs of the Carebot AI MMG medical device are intended to aid decision-making in screening clinical practice, always in conjunction with other relevant patient information and based on the professional judgment of the examining clinician. The Carebot AI MMG is specifically designed to provide a supporting layer of analysis that helps in evaluating or prioritizing mammography images with additional patient information and the professional judgment of the examining physician.

Primary outcomes

  1. Performance Test

    Time frame: 2024

    A multicenter, multi-reader, retrospective study was designed to validate the clinical efficacy of the proposed Carebot AI MMG (also referred to as "DLAD"). Using a non-certified medical device, a test set of retrospectively collected mammography studies in standard projections (CC and MLO). The performance of the DLAD was evaluated against the ground truth for individual indications (breast density evaluation, breast lesion detection) using Accuracy.

Secondary outcomes

  1. Comparison of Accuracy of Clinician and DLAD Image Evaluation (Carebot AI MMG v2.2) in Breast Density Assessment

    Time frame: 2024

    Carebot AI MMG v2.2 classified mammography studies according to the ACR BI-RADS 5th edition, i.e. breast tissue density assessment into A/B/C/D classes. The performance of the Carebot AI MMG device was assessed relative to the ground truth and then compared with the performance of five independent radiologists with varying levels of experience (RAD 1-RAD 5). A rigorous statistical analysis was used in the BI-RADS breast density classification to evaluate the performance of each method - the proposed Carebot AI MMG v2.2 medical device and the compared radiologists in the multi-reader study. The analysis focused on key metrics including Accuracy, F1 Score (Macro-Averaged), Precision (Macro-Averaged), Recall (Macro-Averaged) and Cohen's Kappa (κ) to assess the strength of agreement. Given that all scans were evaluated by all radiologists in the comparison, a bootstrapping method that involves resampling the test data 1000 times.

  2. Comparison of Accuracy of Clinician and DLAD Image Evaluation (Carebot AI MMG v2.2) in Breast Lesion Detection

    Time frame: 2024

    The medical device (DLAD, Carebot AI MMG v2.2) analyzed mammography studies in standard projections (CC and MLO) and classified the presence of lesions ("present" x "absent") at the mammography study level. DLAD performance was assessed relative to the ground truth and then compared to the performance of five independent radiologists with varying levels of experience (RAD 1-RAD 5). The investigators quantified the performance of diagnostic tests based on Sensitivity, Specificity and Balanced Accuracy. The investigators further assessed the statistical significance of differences between DLAD and individual radiologists using appropriate statistical tests. To determine the reliability of the metrics examined, the investigators calculated 95% confidence intervals using Wilson scores. To evaluate the statistical significance of differences in Sensitivity and Specificity between DLAD and individual radiologists, the investigators applied McNemar's test with a continuity correction.

Sponsors and collaborators

Lead sponsor

Carebot s.r.o.

Industry

Registry information

Official study title

Evaluation of Carebot AI MMG Medical Device for Detection and Radiographic Assessment of Breast Lesions and Quantitative Analysis of Breast Density: A Multicentric, Multi-Reader Study

Acronym: EMBLEDDA-MMG

Important dates

Study start
2022
Primary completion
2024
Study completion
2024
First posted
May 24, 2024
Registry last updated
Mar 18, 2026

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.