NCT Number: NCT01756274
The purpose of this study was to evaluate the performance of three Bayer Blood Glucose Monitoring Systems (BGMS) with neonatal blood. Blood samples used in this study were 'left-over samples'. The blood samples were from heel sticks of neonates, collected (into a tube) and sent to the laboratory. All meter Blood Glucose (BG) results were compared with capillary plasma results obtained with a reference laboratory glucose method (Cobas® 6000 instrument).
Looking for future studies?
Notify MeUp to 30 day
All sexes
Interventional
Not applicable
St Louis, Missouri, 63110, United States
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Lab professionals tested the BG concentration using Contour® NEXT BGMS (Blood Glucose Monitoring System). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method, Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).
Lab professionals tested the BG concentration using Contour® PLUS BGMS (Blood Glucose Monitoring System). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method, Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).
Lab professionals tested the BG concentration using Contour® Next EZ BGMS (Blood Glucose Monitoring System). All meter BG results were compared with plasma results obtained with a reference laboratory glucose method, Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).
Time frame: 30 minutes
Laboratory professionals tested the BG concentration of 'left-over samples' using plasma referenced Blood Glucose Monitoring Systems. BGMS results were compared with capillary plasma BG results obtained with a Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).
Time frame: 30 minutes
Laboratory professionals tested the BG concentration of 'left-over samples' using plasma referenced Blood Glucose Monitoring Systems. BGMS results were compared with capillary plasma BG results obtained with a Cobas® 6000 instrument (Roche Diagnostics Corp., Indianapolis, IN).
Time frame: 30 minutes
To evaluate the effect of neonatal age on the meter systems' performances, the number of blood samples that produced unusual meter results out of the total number of blood samples that came from babies less than 24 hours old was reported. Studentized residuals were used to measure the degree to which meter BG results departed from what would be expected using a linear model. (This analysis is not related to BGM accuracy status, which has already been reported.)
Time frame: 30 minutes
To evaluate how the meter systems perform with blood samples drawn in the Neonatal Intensive Care Unit, the number of blood samples that produced unusual meter results out of the total number of blood samples that came from babies in Neonatal Intensive Care was reported. Studentized residuals were used to measure the degree to which meter BG results departed from what would be expected using a linear model. (This analysis is not related to BGM accuracy status, which has already been reported.)
Ascensia Diabetes Care
Industry
Performance of Three Bayer Blood Glucose Monitoring Systems With Blood Samples From Neonates
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00837122
Cardiovascular Diseases, Diabetes
Accra, Ghana
View Trial DetailsNCT03899883
Diabetes, Diabetes Complications
Aurora, Colorado, United States
View Trial DetailsNCT00065676
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial DetailsNCT00862433
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial Details