Bayer HealthCare LLC, Diabetes Care
Mishawaka, Indiana, 46544, United States
NCT Number: NCT01699763
The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and two additional Blood Glucose Monitoring Systems (BGMS) from other companies. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Mishawaka, Indiana, 46544, United States
The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and two additional Blood Glucose Monitoring Systems (BGMS) from other companies. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the three systems was evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing was performed by study staff and some samples were tested from subject fingertips. Additionally, some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with and without diabetes (up to 10% of subjects without diabetes were included). Performance of Contour® NEXT LINK BGMS was evaluated across the glucose range of the BGMSs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with and without diabetes (up to 10% of subjects without diabetes were included). Performance of OneTouch® UltraLink® BGMS was evaluated across the glucose range of the BGMSs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with and without diabetes (up to 10% of subjects without diabetes were included). Performance of Nova Max Link® BGMS was evaluated across the glucose range of the BGMSs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Time frame: 8 hours
Using the overall Blood Glucose (BG) range (34 to 561 mg/dL according to YSI subject plasma results), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI plasma results (BG reference) were compared. MARD is calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value. Higher MARD value indicates larger difference between meter value and the reference value.
Time frame: 8 hours
Using fresh and glycolyzed samples with YSI plasma Blood Glucose (BG) ≤80 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI subject plasma results (BG reference) were compared. MARD is calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value. Higher MARD value indicates larger difference between meter value and the reference value.
Time frame: 8 hours
Using samples with YSI plasma Blood Glucose (BG) >180 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI subject plasma results (BG reference) were compared. MARD is calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value. Higher MARD value indicates larger difference between meter value and the reference value.
Ascensia Diabetes Care
Industry
Evaluation of Blood Glucose Meter Systems - Contour® NEXT LINK Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00837122
Cardiovascular Diseases, Diabetes
Accra, Ghana
View Trial DetailsNCT03899883
Diabetes, Diabetes Complications
Aurora, Colorado, United States
View Trial DetailsNCT00065676
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial DetailsNCT00862433
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial Details