CRT-D
DeviceFor all subjects with the AUTOGEN CRT-D, the autothreshold algorithms will be evaluated
NCT Number: NCT02097290
This study will evaluate the PaceSafe Right Ventricular Autothreshold (RVAT) and Left Ventricular Autothreshold (LVAT) features for AUTOGEN Cardiac Resynchronization Therapy Defibrillator (CRT-D) devices.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Glendale Adventist Medical Center, Glendale, California, United States
The objective of the CAPTIVATE Clinical Study is to gather data to establish the safety and effectiveness of the PaceSafe Right Ventricular Autothreshold (RVAT) and Left Ventricular Autothreshold (LVAT) features to support the regulatory approval of the AUTOGEN CRT-D family of devices.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
For all subjects with the AUTOGEN CRT-D, the autothreshold algorithms will be evaluated
Time frame: 3 months
Safety of the AUTOGEN was evaluated by the system-related complication-free rate (CFR) at 3-months post-implant. The system consists of the implanted AUTOGEN CRT-D pulse generator, RA lead (if implanted), RV lead, and LV lead.
Time frame: 1-month and 3-month follow-up visits
Accuracy of the algorithm will be measured for all patients, by comparing algorithm determined threshold to a core lab determined threshold at both the 1-month and 3-month follow-up visits. The RVAT determined threshold for all patients at 1-month and 3-month follow-up visits are pooled for final analysis. An accurate commanded threshold is defined by: |commanded threshold - core lab determined threshold| ≤ 0.2 V; if the commanded threshold is ≤ 3.5V or |commanded threshold - core lab determined threshold| ≤ 0.5 V; if the commanded threshold is > 3.5V. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets from the 1-month and 3-month visits were pooled for the purpose of this endpoint analysis.
Time frame: 1-month and 3-month follow up visits
Accuracy of the algorithm will be measured for all patients, by comparing algorithm determined threshold to a core lab determined threshold at both the 1-month and 3-month follow-up visits. The LVAT determined threshold for all patients at 1-month and 3-month follow-up visits are pooled for final analysis. An accurate commanded threshold is defined by: commanded threshold - core lab determined threshold| ≤ 0.2 V; if the commanded threshold is ≤ 3.5V or |commanded threshold - core lab determined threshold| ≤ 0.5 V; if the commanded threshold is > 3.5V. Paired datasets from the 1-month and 3-month visits were pooled for the purpose of this endpoint analysis. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets from the 1-month and 3-month visits were pooled for the purpose of this endpoint analysis.
Time frame: 1-month and 3-month follow up visits
Accuracy of the algorithm will be measured for all patients, by comparing algorithm determined threshold to a core lab determined threshold at both the 1-month and 3-month follow-up visits.The RVAT determined threshold for all patients at 1-month and 3-month follow-up visits are pooled for final analysis. An accurate Ambulatory threshold is defined by:|Ambulatory threshold - ECG threshold| ≤ 0.6 V; if the ECG threshold is ≤ 3.5V or |Ambulatory threshold - ECG threshold| ≤ 1.0 V; if the ECG threshold is > 3.5V. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets form the 1-month and 3-month visits were pooled for purposes of endpoint analysis.
Time frame: 1-month and 3-month follow up visits
Accuracy of the algorithm will be measured for all patients, by comparing algorithm determined threshold to a core lab determined threshold at both the 1-month and 3-month follow-up visits. The LVAT determined threshold for all patients at 1-month and 3-month follow-up visits are pooled for final analysis. An accurate Ambulatory threshold is defined by: |Ambulatory threshold - ECG threshold| ≤ 1.0 V. Paired datasets from the 1-month and 3-month visits were pooled for the purpose of this endpoint analysis. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets form the 1-month and 3-month visits were pooled for purposes of endpoint analysis.
Time frame: 3-month follow up visit
This endpoint will evaluate the percent of RVAT commanded tests that result in an appropriate outcome. There are three possible outcomes to a commanded test:
Time frame: 3-month follow up visit
This endpoint will evaluate the percent of LVAT commanded tests that result in an appropriate outcome. There are three possible outcomes to a commanded test:
Boston Scientific Corporation
Industry
CAPTure Information Via Automatic Threshold Evaluation
Acronym: CAPTIVATE
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03446651
Cardiovascular Diseases, Heart Diseases
New Haven, Connecticut, United States
View Trial DetailsNCT05806606
Behavior, Cardiovascular Diseases
Hong Kong
View Trial DetailsNCT06847438
Cardiovascular Diseases, Heart Diseases
Atlanta, Georgia, United States
View Trial DetailsNCT07721909
Cardiovascular Diseases, Heart Diseases
Muscat, Oman
View Trial Details