NCT Number: NCT02041377

Evaluation of an Karajishi TS Investigational Blood Glucose Monitoring System

The purpose of this study was to determine if subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) with no training and obtain valid glucose results.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

AMCR Institute, Escondido, California, United States

Loading trial locations.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Males and females, 18 years of age and older
  • People with type 1 or type 2 diabetes
  • Able to speak, read, and understand English
  • Willing to complete all study procedures

Exclusion criteria

  • Hemophilia or any other bleeding disorder
  • Pregnancy
  • Physical, visual, or neurological impairments that would make the person unable to perform testing with the BGM
  • Previously participated in a BG monitor study using the Karajishi TS BGMS (or used a Bayer Contour TS meter)
  • Working for a medical laboratory, hospital, or other clinical setting that involves training on and clinical use of blood glucose monitors
  • Working for a competitive medical device company, or having an immediate family member who works for such a company
  • A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk

Treatment and study plan

Karajishi TS Investigational Blood Glucose Monitoring System

Device

Subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Karajishi TS Investigational Blood Glucose Monitoring System with no training. All BG results were compared to reference method results obtained from subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.

Primary outcomes

  1. Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

    Time frame: 1 hour

    Subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma).

Secondary outcomes

  1. Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

    Time frame: 1 hour

    Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI venous plasma) and +/-15% (>=100 mg/dL YSI venous plasma).

  2. Number of Blood Glucose (BG) Results From Alternative Site Testing (AST) Palm Blood Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

    Time frame: 1 hour

    Subjects with diabetes self-tested Alternative Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS AST palm results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma).

  3. Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff

    Time frame: 1 hour

    Study staff obtained and tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma).

  4. Number of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire Statements

    Time frame: 1 hour

    Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond Strongly Agree; Agree; Neutral; Disagree; or Strongly Disagree.

Sponsors and collaborators

Lead sponsor

Ascensia Diabetes Care

Industry

Registry information

Official study title

User Performance of the Karajishi Contour TS Blood Glucose Monitoring System

Important dates

Study start
2014
Primary completion
2014
Study completion
2014
First posted
Jan 22, 2014
Registry last updated
Feb 29, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)