Evaluation of Accuracy of Continuous Glucose Monitoring (CGM) in Patients With End Stage Renal Disease (ESRD) on Intermittent Hemodialysis (iHD).

Recent advances in continuous glucose monitors (CGMs) and availability of commercial CGM products to patients with type 1 and type 2 diabetes has made the use of CGM more widespread. CGMs work by placing a probe underneath the skin of a patient, into the interstitial space. Patients with end stage renal disease (ESRD) who are on intermittent hemodialysis (iHD) or peritoneal dialysis (PD) undergo fluid shifts between the interstitial fluid and intravascular space during dialysis treatments.These fluid shifts, uremia, acidosis, and volume overload (increase in interstitial fluid volume due to ESRD) have the potential to impact the performance of the most advanced and commercially available CGMs; however, use of CGM in these patients has not yet been studied.Use of CGM, and potentially hybrid closed loop insulin delivery systems that are dependent on accurate continuous glucose monitoring, has the potential to improve glucose control and quality of life in these patients (7). This study team feels that this study will be valuable in collecting preliminary data needed with the goal of validating the use of CGM in this patient population. The specific aim is to conduct a pilot study to evaluate the accuracy of continuous glucose monitors (CGM) in End Stage Renal Disease (ESRD) patients on intermittent hemodialysis (iHD).

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Meaghan Stumpf, MD

Charlottesville, Virginia, 22903, United States

About this study

Recent advances in continuous glucose monitors (CGMs) and availability of commercial CGM products to patients with type 1 and type 2 diabetes has made the use of CGM more widespread (1). CGMs work by placing a probe underneath the skin of a patient, into the interstitial space. The probe is an electroenzymatic sensor which uses glucose oxidase to break down glucose to create hydrogen peroxidase and other elements. Hydrogen peroxidase then interacts with a base metal layer of the sensor and is oxidized, which results in release of electrons which creates a current. The current is proportional to the glucose concentration. The current is measured by the probe and transmits a calculated glucose concentration to a receiving device (2). Substances that are widely distributed in body water, and thereby present in the interstitial space, potentially affect this technology. Acetaminophen and aspirin are substances that are have been known to affect the accuracy of these devices (3); however, more recently developed CGMs such as the Dexcom G6, were able to demonstrate no interference by acetaminophen (4). Patients with end stage renal disease (ESRD) who are on intermittent hemodialysis (iHD) or peritoneal dialysis (PD) undergo fluid shifts between the interstitial fluid and intravascular space during dialysis treatments. They are also often uremic and have metabolic acidosis (5). These fluid shifts, uremia, acidosis, and volume overload (increase in interstitial fluid volume due to ESRD) have the potential to impact the performance of the most advanced and commercially available CGMs; however, use of CGM in these patients has not yet been studied (3). Moderate to severe CKD is associated with both increase in insulin resistance and decrease in insulin clearance, which results in often unpredictable and labile glucose concentrations and increased risk of hypoglycemia in these patients (6). Use of CGM, and potentially hybrid closed loop insulin delivery systems that are dependent on accurate continuous glucose monitoring, has the potential to improve glucose control and quality of life in these patients (7). This study team feels that this study will be valuable in collecting preliminary data needed with the goal of validating the use of CGM in this patient population.OBJECTIVE: The specific aim is to conduct a pilot study to evaluate the accuracy of continuous glucose monitors (CGM) in End Stage Renal Disease (ESRD) patients on intermittent hemodialysis (iHD). Accuracy will be assessed by calculating the mean absolute relative difference (MARD) between CGM values and concurrent finger stick or capillary blood glucose (CBG) in these patients during hemodialysis, and on non-dialysis days.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Ages 18+
  • Type 1 diabetes mellitus on intermittent HD thrice weekly OR Type 2 diabetes mellitus on intermittent HD thrice weekly
  • Willingness and ability to comply with scheduled visits and study procedures

Exclusion criteria

  • Inability to comply with finger stick blood glucoses at least four times daily
  • Noncompliant with HD therapies
  • Pregnant women

Treatment and study plan

Continuous Glucose Monitor

Device

Use of a continuous glucose monitor during study period.

Primary outcomes

  1. Mean Absolute Relative Difference Between CGM Value and Capillary Blood Glucose (Self-monitoring Blood Glucose [SMBG])

    Time frame: From CGM placement to CGM removal (10 days)

    Mean Absolute Relative Difference (MARD) between CGM value and capillary blood glucose (SMBG) performed at home 4 to 7 times per day by the participant for 684 matched pairs.

  2. Mean Absolute Relative Difference (MARD) Between Continuous Glucose Monitor (CGM) Value and Venous Blood Glucose (vBGM)

    Time frame: From CGM placement to CGM removal (10 days)

    Venous blood glucose samples were collected approximately 12 blood samples from the existing hemodialysis (HD) intravenous (IV) line during each (three) HD session the CGM sensor was worn. These blood samples were immediately processed using the i-STAT System. While the goal was to have the subject participate in three hemodialysis sessions, two sessions were acceptable. 624 matched data pairs were analyzed.

Sponsors and collaborators

Lead sponsor

University of Virginia

Other

Collaborators

  • DexCom, Inc.

Registry information

Important dates

Study start
2020
Primary completion
2021
Study completion
2021
First posted
Sep 18, 2019
Registry last updated