Evaluation of Accuracy of CFD-based RuiXin-FFR by Comparing With Pressure-wire-based FFR

Coronary Artery Disease (CAD) is the top killer nowadays. Pressure-wire-based Fractional Flow Reserve (FFR) is the gold standard for measuring ischemia in coronary arteries. CFD-based RuiXin-FFR, which is noninvasive, is developed recently. But its accuracy is not verified. This is a multi-center and prospective study to evaluate the sensitivity, specificity, and accuracy of CFD-based RuiXin-FFR compared with wire-based FFR.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Shaoping Nie

Beijing, China

About this study

The participants are entered with suspended CAD and angiography ischemia between 30%-90% from 6 hospitals in China. Pressure-wire-based FFR is conducted in the standard protocol by each hospital. FFR>0.8 is chosen as the threshold to indicate non-ischemia. CFD-based RuiXin-FFR is performed by Raysight Medical (Shenzhen, China). CT images are acquired in a standard protocol by each hospital. Images with poor quality are excluded in this trial. Based on these images, RuiXin-FFR reconstructs the 3D model of coronary trees and obtains RuiXin-FFR values by conducting CFD simulations in a blinded fashion. RuiXin-FFR>0.8 is chosen as the threshold to indicate non-ischemia as well. By comparing the clinical diagnosis outcome of pressure-wire-based FFR and CFD-based RuiXin-FFR, the sensitivity and specificity of the RuiXin-FFR are obtained, which is the primary end point of this trial. The secondary end point includes the measurement of accuracy, NPV, PPV and ROC of the RuiXin-FFR.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • able to understand the purpose of the study and sign the informed consent
  • with diagnosed or confirmed coronary artery disease according to the comprehensive clinical assessment
  • with CTA image indicating that the diameter of the reference vessel in the stenosis segment was ≥2mm
  • with CTA image indicating that the stenosis degree of coronary artery lumen diameter ≥30% and ≤90%

Exclusion criteria

  • prior coronary artery bypass bypass (CABG) surgery, coronary interventional therapy (PCI), artificial heart valve implantation, cardiac pacemaker or implantable defibrillator implantation
  • persistent or active symptoms of clinical instability, including acute chest pain (sudden onset), cardiogenic shock, unstable blood pressure status (systolic pressure less than 90 mmHg), severe congestive heart failure (NYHA grade III or IV), or acute pulmonary edema
  • acute myocardial infarction occurred within 7 days before inclusion
  • complex congenital heart disease, sick sinus syndrome, long QT syndrome, severe arrhythmia, tachycardia, severe asthma, severe or very severe chronic obstructive pulmonary disease (COPD) and chronic renal damage (serum creatinine value > 1.5 mg/dl or creatinine clearance < 45 ml/Kg*1.73 m2)
  • there are contraindications for the use of adenosine disodium triphosphate
  • allergic to iodized contrast media
  • pregnancy or pregnancy status unknown
  • life expectancy less than 2 months
  • there are any factors that other researchers consider not suitable for inclusion or completion of this study
  • obvious mismatch of coronary artery CTA images
  • CTA image showing calcification occupies the cross-sectional area of the lumen > 80%
  • CT value standard deviation of aortic root image >30HU
  • coronary artery occlusion.

Treatment and study plan

Pressure wire fractional flow reserve

Diagnostic Test

Pressure wire is inserted into coronary arteries. The ratio of the average distal pressure and average aorta pressure is FFR.

CFD-based RuiXin-FFR

Diagnostic Test

CT-FFR was evaluated in a blinded fashion with a "Coarse-to-Fine Subpixel" algorithm for lumen contour

Primary outcomes

  1. Sensitivity and Specificity

    Time frame: 9 days

    By comparing with pressure-wire-based FFR, the sensitivity and specificity of CFD-based RuiXin-FFR are measured

Sponsors and collaborators

Lead sponsor

Beijing Anzhen Hospital

Other

Registry information

Official study title

Accuracy of RuiXin-FFR by Comparing With Pressure-wire-based FFR in Detecting Hemodynamically Significant Stenosis: A Prospective Multicenter Study

Important dates

Study start
2019
Primary completion
2020
Study completion
2020
First posted
Jan 29, 2021
Registry last updated
Sep 5, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.