Electronic Repositioning With Acuity and Easytrak Leads

Evaluate appropriate performance of Electronic Repositioning™ (ER) in the clinical routine, cumulative incidence of phrenic nerve stimulation (PNS) & success rates of Electronic Repositioning (ER) to remove it, various incl. left ventricular pacing (LVP) thresholds and PNS

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

University Hospital

Düsseldorf, 40225, Germany

About this study

Bipolar leads for left ventricular pacing (LVP) such as the newly available ACUITY™ or the standard EASYTRAK™ II and EASYTRAK™ III leads from GUIDANT in combination with suitable devices for cardiac resynchronization therapy (CRT) allow for the non-invasive programming of different Left Ventricular Pacing (LVP) configurations. Hence, this Electronic Repositioning (ER) may help physicians to overcome problems in cardiac resynchronization therapy (CRT) such as ineffective biventricular pacing, high Left Ventricular Pacing (LVP) thresholds and phrenic nerve stimulation (PNS) without the need for an additional patient's operation to reposition LV leads. The ERACE study it a multi-center, prospective, non-randomized clinical study to document and evaluate appropriate performance of GUIDANT's Electronic Repositioning™ in the clinical routine with respect to the following major aspects:

avoidance of PNS and low LVP thresholds (for low battery consumption and extended device longevity) in standard cardiac resynchronization therapy (CRT)

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patients receiving their FIRST implantation of a Left Ventricular (LV) lead in combination with a CRT device with implantable cardioverter defibrillator (ICD) backup and the possibility for Electronic Repositioning™
  • Patients willing (= signed written consent) and capable to participate in all procedures of the study"

Exclusion criteria

  • Patients who will not be available for routine Follow up
  • 18 years old
  • Known pregnancy.

Treatment and study plan

Electronic Repositioning

Device

Change of pacing vectors for CRT pacing

Primary outcomes

  1. Proportion of Patients Without the Occurrence of Cardiac Resynchronization Therapy (CRT)-Induced Phrenic Nerve Stimulation (PNS) Between Implant and First Regular Follow-up (FU)

    Time frame: From implant until first follow up (occurred between 3 or 6 month after implant procedure)

    The primary endpoint was the percentage of patients in whom PNS induced by Left Ventricular (LV) pacing could be avoided by electronic repositioning from implantation to first regular follow up (3-6 months postimplant). In this evaluation, prevention of PNS was defined as PNS appearing with the standard pacing configuration-LV tip as cathode versus Right Ventricle (RV) coil as anode-and being avoided by at least one of the three alternative reprogrammable configurations.

Sponsors and collaborators

Lead sponsor

Boston Scientific Corporation

Industry

Registry information

Acronym: ERACE

Important dates

Study start
2006
Primary completion
2008
Study completion
2008
First posted
Feb 7, 2008
Registry last updated
Apr 21, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.