NCT Number: NCT03938220
The aim of this study is to evaluate the diagnostic accuracy of electrical cardiometry (EC) for the noninvasive determination of fluid responsiveness in sepsis and agreement of (EC) compared to transthoracic echocardiography (TTE).
Looking for future studies?
Notify Me19 year–65 year
All sexes
Observational
Tanta, ElGharbiaa, 31511, Egypt
Usual investigations [complete blood count (CBC), C reactive protein (CRP), serum procalcitonin, serum lactate, arterial blood gases (ABG), blood culture, electrocardiogram (ECG) and liver and renal function tests] will be done.
Once the diagnosis of sepsis is definite and the patient develops hypotension, all patients will undergo simultaneous measurement by (EC) using the ICON_ device and (TTE).
Electrical cardiometry (EC) measurements: by the ICON_ hemodynamic monitor (ICON Cardiotronics, Inc., La Jolla, CA 92307; Osyka Medical GmbH, Berlin, and Germany, model C3, Serial no: 1725303). Four sensors of (EC) will be applied (1st: 5 cm above the base of the neck, 2nd: on the base of the neck, 3rd: lower thorax at the level of the xiphoid and 4th: 5 cm below the 3rd electrode at the level of the anterior axillary line). The ICON continuously displays heart rate, stroke volume, and cardiac output.
Transthoracic echocardiography (TTE) measurements: will be performed by using Philips (CX50 - Extreme edition) equipped with echo probe. SV of the left ventricle will be calculated using LVOT diameter (D) just below the aortic valve from parasternal long-axis view and velocity time integral (VTI) measured in left ventricle outflow tract (LVOT) from apical 5 chamber view (by pulsed wave Doppler), respectively. The machine's built-in software uses the formula ''(πD2/4) × VTI × HR'' to calculate CO.
Management of sepsis will be done according to surviving sepsis campaign guidelines in 2016 and its update 2018.
Fluid resuscitation will be 30 mL/kg of intravenous infusion of lactated ringer to be given within the first 3 h (≈5 mL/kg/30 min) guided by fluid responsiveness (fluid responder if SV increases by > 10% after the fluid challenge). If the patient becomes fluid non-responder, vasopressor infusion (norepinephrine 0.05-0.3µg/kg/min) will start. The end of the study is when mean arterial blood pressure > 65 mmHg (either by fluid or both fluid and vasopressor).
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Electrical cardiometry (EC) measurements: by the ICON_ hemodynamic monitor (ICON Cardiotronics, Inc., La Jolla, CA 92307; Osyka Medical GmbH, Berlin, and Germany, model C3, Serial no: 1725303). Four sensors of EC will be applied (1st: 5 cm above the base of the neck, 2nd: on the base of neck, 3rd: lower thorax at the level of the xiphoid and 4th: 5 cm below the 3rd electrode at the level of anterior axillary line). The ICON continuously displays HR, SV and CO.
Time frame: 24 hours
Fluid resuscitation will be 30 mL/kg of intravenous infusion of lactated ringer to be given within the first 3 h (≈5 mL/kg/30 min) guided by fluid responsiveness (fluid responder if stroke volume increases by > 10% after the fluid challenge). If the patient becomes fluid non-responder, vasopressor infusion (norepinephrine 0.05-0.3µg/kg/min) will start. The end of the study is when mean arterial blood pressure> 65 mmHg (either by fluid or both fluid and vasopressor).
Time frame: 24 hours
Fluid resuscitation will be 30 mL/kg of intravenous infusion of lactated ringer to be given within the first 3 h (≈5 mL/kg/30 min) guided by fluid responsiveness (fluid responder if stroke volume increases by > 10% after the fluid challenge). If the patient becomes fluid non-responder, vasopressor infusion (norepinephrine 0.05-0.3µg/kg/min) will start. The end of the study is when mean arterial blood pressure> 65 mmHg (either by fluid or both fluid and vasopressor).
Tanta University
Other
Validation of Electrical Cardiometry Measurements Compared to Transthoracic Echocardiography in Fluid Responsiveness in Sepsis
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06075407
Electrical Cardiometry, Fluid Resuscitation
Cairo, Egypt
View Trial DetailsNCT05272982
Electrical Cardiometry, Intensive Care Unit
Tanta, El-Gharbia Governorate, Egypt, Egypt
View Trial DetailsNCT03131102
Biosignals, Electrical Cardiometry
Berlin, Germany
View Trial DetailsNCT04618419
Electrical Cardiometry, Ultrasound of IJV
Alexandria, Egypt
View Trial Details