NCT Number: NCT04799964

Efficacy and Safety of the Flow Diverter (Tonbridge) for Endovascular Treatment of Intracranial Aneurysms

The purpose of this study is to assess the efficacy and safety of the Flow Diverter (Tonbridge) for endovascular treatment of intracranial aneurysms.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–80 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

The First Affiliated Hospital of USTC, Hefei, Anhui, China

Loading trial locations.

About this study

This is a prospective, multi-center, randomized, open, positive-controlled, non-inferiority trial. This clinical trial is conducted at more than 5 centers in China. Eligible patients are randomized into experimental group using Flow Diverter (Tonbridge) or control group using Tubridge (MicroPort) in a 1:1 ratio. The primary objective of this study is to verify the complete occlusion rate of aneurysms at 12 months of Flow Diverter (Tonbridge) is non-inferior to Tubridge, for treatment of intracranial aneurysms, and evaluate the effectiveness and safety of the Flow Diverter (Tonbridge).

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Age 18 to 80 years, any gender;
  • Unruptured intracranial aneurysms locate in the internal carotid artery (ICA) or vertebral artery (VA) with a neck ≥ 4 mm and a maximal diameter ≥ 10 mm, which diagnosed by DSA;
  • The target aneurysm with a parent vessel diameter of 2.0-6.5 mm;
  • Subject or guardian is able to understand the purpose of study, shows sufficient compliance with the study protocol and provides a signed informed consent form.

Exclusion criteria

  • Diagnosed as multiple aneurysms which cannot be covered by a single stent or with cerebrovascular malformations;
  • Subarachnoid hemorrhage of the target aneurysm within 30 days pre-procedure;
  • Modified Rankin Scale (mRS) score > 2 in pre-procedure;
  • Platelet (PLT) < 60×10^9/L or known platelet dysfunction or International Normalized Ratio (INR) > 1.5;
  • Heart, lung, liver and renal failure or other severe diseases (such as brain tumor, systemic infection, disseminated intravascular coagulation, myocardial infarction within 12 months before enrollment, history of severe psychosis);
  • The lesion which is not appropriate for flow diverter delivery and deployment judged by investigators (severe stenosis of parent artery, severe tortuosity of parent artery, stent failing to reach the lesion, etc.);
  • Major surgery within 30 days before enrollment, or intending to receive surgery within 180 days after enrollment;
  • Allergic history of anesthetics and contrast agents, or contraindication for dual antiplatelet or/and anticoagulant therapy;
  • Allergic History of metals such as nickel-titanium alloy;
  • Life expectancy < 12 months;
  • Pregnant or breastfeeding women;
  • Subject has participated in other drug or medical device clinical trials within 1 month before signing informed consent;
  • Other conditions judged by the investigators as unsuitable for enrollment.

Treatment and study plan

Flow Diverter (Tonbridge)

Device

The device is a self-expanding vascular stent system, which consists of a stent and a delivery system. The stent is made of platinum core nickel-titanium alloy material. The delivery system is composed of a delivery guide wire and an introducer sheath. The stent and the delivery guide wire are pre-installed in the introducer sheath.

Tubridge (MicroPort)

Device

Tubridge consists of a stent system and a microcatheter system. The stent system consists of a stent and a conveyor, the stent is self-expanding as well.

Primary outcomes

  1. Complete occlusion rate of aneurysms at 12 months

    Time frame: 360±45 days post-procedure

    Complete occlusion is defined as no contrast agent in the aneurysm (Raymond-Roy class I) which is diagnosed by cerebrovascular DSA at 12 months post-procedure.

Secondary outcomes

  1. Immediate procedural success rate

    Time frame: intra-procedure

    Intraoperative cerebrovascular DSA is performed to determine whether the investigational device positions exactly and covers the aneurysm neck effectively.

  2. Complete occlusion rate of aneurysms at 6 months

    Time frame: 180±30 days post-procedure

    Complete occlusion is defined as no contrast agent in the aneurysm (Raymond-Roy class I) which is diagnosed by MRA/CTA at 6 months post-procedure.

  3. Subtotal occlusion rate

    Time frame: 360±45 days post-procedure

    Subtotal occlusion is defined as contrast agents in the aneurysm sac less than 10% which is diagnosed by DSA at 12 months post-procedure.

  4. Incidence of technical complications

    Time frame: intra-procedure

    Technical complications include but not limited to: stent incomplete opening, stent migration, stent herniation into the aneurysm sac, in-stent thrombosis, operation injury, etc.

  5. Incidence of stroke

    Time frame: within 360±45 days

    Stroke includes hemorrhagic stroke and symptomatic ischemic stroke.

  6. Incidence of parent artery stenosis ( > 50%) in target area at 360 days and incidence of parent artery occlusion in target area at 360 days

    Time frame: 360±45 days post-procedure

    Parent artery stenosis in target area is defined as stenosis degree > 50%. Using postoperative cerebrovascular DSA at 360 days is performed to determine.

  7. Incidence of adverse events and incidence of serious adverse events

    Time frame: within 360±45 days

    "Incidence of adverse events" is the proportion of subjects with adverse events using the investigational device or the comparator;

    "Incidence of serious adverse events" is the proportion of subjects with serious adverse events using the investigational device or the comparator.

  8. Mortality rate

    Time frame: within 360±45 days

    Deaths due to any cause are calculated.

  9. Operation satisfaction rate

    Time frame: intra-procedure

    The 5-point Likert scale is used to evaluate the operation satisfaction from three aspects: push performance, retracting performance and release performance. "Operation satisfaction rate" is defined as the proportion of investigational devices or comparators with Likert score ≥ 12 points.

  10. Incidence of device deficiency

    Time frame: within 360±45 days

    Device deficiency is the unreasonable risk that may endanger human health and life safety in the normal use of medical devices during clinical trials, such as labeling errors, quality problems, malfunctions, etc.

Sponsors and collaborators

Lead sponsor

Ton-Bridge Medical Tech. Co., Ltd

Industry

Registry information

Official study title

Efficacy and Safety of the Flow Diverter (Tonbridge) for Endovascular Treatment of Intracranial Aneurysms: a Prospective, Multi-center, Randomized, Open, Positive-controlled, Non-inferiority Trial

Important dates

Study start
2021
Primary completion
2023
Study completion
2023
First posted
Mar 16, 2021
Registry last updated
Sep 19, 2024

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.