Department of Cardiac Surgery, University of Verona
Verona, 37126, Italy
NCT Number: NCT03937570
After the implantation of a left ventricular assist device (LVAD), many patients continue to experience exercise intolerance. LVAD echo guided optimization (EO) determines a more favorable hemodynamic profile and could provide an improvement on functional capacity (FC). VAFRACT is a prospective randomized trial to evaluate the additional benefit of an EO approach on FC, measured by cardiopulmonary exercise test (CPET), in LVAD optimization free population.
Looking for future studies?
Notify Me18 year–80 year
All sexes
Interventional
Not applicable
Verona, 37126, Italy
The use of Left Ventricular Assist Devices (LVAD) is a viable therapeutic option to improve survival and quality of life of patients with advanced and refractory heart failure (HF). However, after the implantation of an LVAD, patients' functional capacity (FC) is still reduced with maximum oxygen uptake (VO2 peak) values calculated at the cardiopulmonary exercise test (CPET) ranging from 11 to 20 ml/kg/min. LVAD echo guided optimization (EO) determines a more favorable hemodynamic profile, in particular for the right ventricle, and could provide an improvement on FC. VAFRACT is a prospective randomized trial to evaluate the additional benefit of an EO approach on FC, measured by CPET in LVAD optimization free population.
The optimal device speed is defined as the one that allows an intermittent aortic valve opening and a neutral position of the interventricular septum without increasing aortic or tricuspid regurgitation associated or not to a dilatation of the right ventricle (RV). In this condition it is reasonable to expect a more favorable hemodynamic profile for the RV and a better response in terms of functional capacity.
Subjects studied are HF patients supported with a continuous-flow LVAD: HeartMate II (Thoratec Inc., Pleasanton, CA) and HeartMate 3 (HM 3, Abbott, North Chicago, IL). Patients are recruited from our Day Hospital of the Cardiac Surgery Department at the University Hospital of Verona.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
LVAD echo-optimization consists of routine comprehensive transthoracic echocardiography at the baseline speed setting, followed by stepwise incremental adjustments to the LVAD speed (revolutions per minute: rpm), with collection of prespecified echocardiographic parameters at each new speed (eg, left ventricle end-diastolic diameter, interventricular septal position, aortic valve opening frequency/duration, tricuspid and/or mitral regurgitation severity). The optimal velocity is defined as the one that allows an intermittent aortic valve opening and a neutral position of the interventricular septum without increasing aortic or tricuspid regurgitation, associated or not to a dilatation of the right ventricle. The recommended speed range varies according to the indications given in the data sheet for each specific device.
Time frame: three months
peak oxygen uptake measured by cardiopulmonary exercise test
Time frame: Three months
Rate of hospitalizations for: device thrombosis, hemorrhagic events, infections, right heart failure, arrhythmias.
Time frame: Three months
Changes in quality of life perceived by EuroQoL (quality of life) Scale. It is an instrument which evaluates the quality of life measuring with one question: mobility, self-care, usual activities, pain/discomfort, and anxiety/depression. The answers given to these question (1 = no problems; 2 = some problems; 3 = extreme problems) are converted into EQ-5D index, an utility scores anchored at 0 for death and 1 for perfect health. The EQ-5D questionnaire also includes a Visual Analog Scale (VAS), by which respondents can report their perceived health status with a grade ranging from 0 (the worst possible health status) to 100 (the best possible health status).
Time frame: Three months
Nt-proBNP levels
Time frame: Three months
Fractional area change of right ventricle assessed by echocardiography
Time frame: Three months
Cardiopulmonary exercise time
Universita di Verona
Other
The Effects of Echo-optimization of Left Ventricular Assist Devices on Functional Capacity, a RAndomized Controlled Trial (VAFRACT)
Acronym: VAFRACT
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03446651
Cardiovascular Diseases, Heart Diseases
New Haven, Connecticut, United States
View Trial DetailsNCT05806606
Behavior, Cardiovascular Diseases
Hong Kong
View Trial DetailsNCT06847438
Cardiovascular Diseases, Heart Diseases
Atlanta, Georgia, United States
View Trial DetailsNCT07721909
Cardiovascular Diseases, Heart Diseases
Muscat, Oman
View Trial Details