mCRM Therapy System
DeviceCommunication testing between S-ICD and LCP in 4 body postures as well as required electrical testing.
Other names: Communication of S-ICD to Leadless Cardiac Pacemaker (LCP)
NCT Number: NCT04798768
The MODULAR ATP Clinical Study is designed to demonstrate safety, performance, and effectiveness of the Modular Cardiac Rhythm Management (mCRM) Therapy System.
This study is active but is not currently recruiting participants.
18 year and older
All sexes
Interventional
Not applicable
Kepler Universitaetsklinikum, Linz, Austria
The MODULAR ATP Clinical Study will enroll subjects with a standard Implantable Cardioverter Defibrillator (ICD) indication applying international practice guidelines, as well as those who already have an implanted S-ICD System and satisfy the inclusion criteria for this study, while not meeting any exclusion criteria. Subjects will be followed for at least 6 months following mCRM Therapy System implantation.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
[i]Al-Khatib, et al. 2017 AHA/ACC/HRS Guideline for Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death. A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. (2018); 138:e272-e391.
[ii] 2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: The Task Force for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death of the European Society of Cardiology (ESC). European Heart Journal (2015) 36, 2793-2867.
[iii] TV-ICD system is expected to be fully explanted during or prior to full Coordinated System implant
[iv] Potential subjects with a Model 1010 S-ICD Pulse Generator are only eligible for MODULAR ATP if they are getting upgraded to Model A209, A219 or future BSC S-ICD Pulse Generator; Patients with an existing S-ICD PG subject to the electrical overstress field action are only eligible for MODULAR ATP if they are getting a new BSC Model A209 or A219, or future BSC S-ICD Pulse Generator
*Significant cardiac scar is defined as a scar involving at least one ventricular myocardial segment (i.e., basal infero-septum) as identified in the official findings of a cMRI, or nuclear viability study, or echo report by the interpreting radiologist/ cardiologist who is not affiliated with the study
**meaningful survival means that a patient has a reasonable quality of life and functional status
Communication testing between S-ICD and LCP in 4 body postures as well as required electrical testing.
Other names: Communication of S-ICD to Leadless Cardiac Pacemaker (LCP)
Time frame: Implant through 6 Months Post-Implant
Major EMPOWER MPS System- and Procedure-related Complication-Free Rate Subjects will be assessed for safety issues related to the procedure or system through 6 months post implant
Time frame: Implant through 12 Months Post-Implant
Major EMPOWER MPS System- and Procedure-related Complication-free Rate
Subjects will be assessed for safety issues related to the procedure or system through 12 months post implant
Time frame: At the 6 Month Follow-up
Communication Success between the S-ICD and EMPOWER PG
Data from subjects will be assessed for effectiveness of communication between S-ICD and the EMPOWER PG by evaluating if a paced beat is present during communication testing in four postures: upright, supine, and right and left side.
Time frame: At the 6 Month Follow-up
Proportion of Subjects with Adequate Pacing Capture Threshold
Effectiveness will be confirmed by evaluating the percentage of subjects considered to be a PCT Responder, defined as a subject with a PCT measurement of ≤ 2.0 V @ 0.4 ms pulse width
Time frame: At the 3 Month Visit
Metabolic-Chronotropic Relation Slope (MCR Slope) from the Kay-Wilkoff Model
Using the Kay-Wilkoff model, data will be assessed to evaluate the proportionality of the EMPOWER PG's sensor-indicated rate to the subject's workload during the treadmill test
Time frame: Implant through 2 years post-implant
All-Cause Survival; data not reported as it is still being collected
Boston Scientific Corporation
Industry
Acronym: MODULAR ATP
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT01940081
Arrhythmias, Cardiac, Cardiac Conduction System Disease
Leiden, Netherlands
View Trial DetailsNCT00000480
Arrhythmia, Arrhythmias, Cardiac
View Trial DetailsNCT00000531
Arrhythmia, Arrhythmias, Cardiac
View Trial DetailsNCT00004559
Arrhythmia, Arrhythmias, Cardiac
View Trial Details