Abbott Sensor Based Glucose Monitoring System
DeviceSubjects will wear the Abbott Sensor Based Glucose Monitoring Systems and will receive no treatment except for safety purposes.
NCT Number: NCT02283411
The purpose of the study is to evaluate the safety and effectiveness of the Abbott Sensor Based Glucose Monitoring System Personal (System-P) and Abbott Sensor Based Glucose Monitoring System Professional (System-Pro) when used as an adjunct to blood glucose testing over a 14-day wear period in adult subjects. The Abbott Sensor Based Glucose Monitoring System - Professional Sensors will be worn by adult subjects over a 14-day wear period.
The primary objective is to characterize the Systems performance with respect to Yellow Spring Instrument (YSI) reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the Abbott Sensor Based Glucose Monitoring Systems in reference to YSI.
Safety of the Abbott Sensor Based Glucose Monitoring Systems will be characterized by Adverse Device Effects and Serious Adverse Device Effects experienced by study participants.
240 subjects will be enrolled and additional healthy subjects may also be enrolled in the study as training subjects.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
William Sansum Diabetes Center, Santa Barbara, California, United States
Sensors are inserted into the back of the subject's upper arm for up to fourteen (14) days. The subjects are expected to perform capillary BG tests during the 14-day sensor wear. Three in clinic visits are scheduled during the 14-day sensor wear period. During each in-clinic visit, study staff performed IV blood draws to obtain venous blood for YSI reference glucose measurements.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Additional Criteria
Subjects will wear the Abbott Sensor Based Glucose Monitoring Systems and will receive no treatment except for safety purposes.
Time frame: 14 days
Point accuracy of the system was evaluated as the proportion of System readings that are within ±20% of the YSI reference value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for YSI glucose levels <80 mg/dL. All 125 subjects wore both System-P and System-Pro Sensors and were included in the outcome measure. System P, the Personal System which is intended for single patient use and System Pro is intended for use by healthcare professionals (HCP). Each subject had up to 136 YSI samples collected during the study which were paired with sensor glucose readings measured at the same time. Point accuracy of the system was evaluated as the percentage of sensor glucose readings that are within ±20% of the YSI reference value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for YSI glucose levels <80 mg/dL.
Time frame: Safety was evaluated throughout the subject's study participation.
Safety of the Abbott Sensor Based Glucose Monitoring Systems, including adverse device effects and serious adverse device effects were assessed for all participants enrolled in the study.
Time frame: 14 days
Trend accuracy of the device performance was assessed by comparing of the glucose rate of change results between the sensor glucose readings and YSI reference results for System P only.
Time frame: 14 days
The Continuous Glucose Error Grid Analysis (CG-EGA) in the Hypoglycemic range (YSI ≤ 70 mg/dL), in the Euglycemic range (70 < YSI ≤ 180 mg/dL) and in the Hyperglycemic range (YSI > 180 mg/dL) were evaluated.
Time frame: 14 days
Point accuracy of the system was evaluated as the proportion of System readings that are within ±20% of the capillary blood glucose (BG) value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for capillary blood glucose (BG) levels <80 mg/dL. All 125 subjects wore both System-P and System-Pro Sensors and were included in the outcome measure. System P, the Personal System which is intended for single patient use and System Pro is intended for use by healthcare professionals (HCP). Each subject had approximately112 capillary blood glucose (BG) samples collected during the study which were paired with sensor glucose readings measured at the same time. Point accuracy of the system was evaluated as the percentage of sensor glucose readings that are within ±20% of the BG value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for BG levels <80 mg/dL.
Abbott Diabetes Care
Industry
Effectiveness and Safety Study of the Abbott Sensor Based Glucose Monitoring System - Personal and Abbott Sensor Based Glucose Monitoring System - Professional
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06890208
Anxiety Disorders, Arrhythmias, Cardiac
Hamburg, Germany
View Trial DetailsNCT07052292
Autoimmune Diseases, Diabetes Mellitus
Wroclaw, Kuyavian-Pomeranian Voivodeship, Poland
View Trial DetailsNCT03987802
Autoimmune Diseases, Diabetes Mellitus
Hyderabad, Andhra Pradesh, India
View Trial DetailsNCT06330194
Autoimmune Diseases, Chronic Disease
Copenhagen, Denmark
View Trial Details