The primary objective of the study is to evaluate the safety and effectiveness of the G4 System (System) when used as an adjuvant to blood glucose testing over a 7-day period in pediatric subjects with diabetes mellitus.
System effectiveness will be primarily evaluated with comparison to laboratory reference Yellow Springs Instrument (YSI) venous sample measurements. The system performance will be evaluated in terms of point and rate accuracy of the system in reference to YSI. The point accuracy is measured as the proportion of G4 System values that are within ±20% of YSI reference value for glucose levels >80 mg/dL and ±20 mg/dL at YSI glucose levels <80 mg/dL. The trend accuracy of the device performance, i.e., continuous glucose error grid analysis, will be evaluated as well as the temporal accuracy at different glucose rates of changes and different glucose ranges (hypoglycemic, euglycemic, and hyperglycemic ranges). The primary matched paired (Sensor-YSI) measurements will be collected during one in-clinic session. System effectiveness will also be evaluated with SMBG measurements similarly.
Safety data of the system will also be collected and safety will be characterized by the incidence of Adverse Device Effects (ADE) experienced by study participants.
Other objectives include the evaluation of device performance at an alternate site (the upper buttocks).
Looking for future studies?
Notify Me2 year–17 year
All sexes
Observational
Advanced Metabolic Care and Research, Escondido, California, United States
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: 7 days
The primary objective is to characterize System effectiveness in comparison with the laboratory reference YSI venous sample measurements.
Time frame: 7 days
An additional objective is to evaluate System performance at an alternate wear site. The System performance will be assessed, using the same statistical methodology as the primary wear site.
DexCom, Inc.
Industry
Effectiveness and Safety of the Dexcom™ G4 Continuous Glucose Monitoring System in Pediatric Subjects With Diabetes Mellitus.
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06406439
Diabetes (Insulin-requiring, Type 1 or Type 2), Diabetes Mellitus
Gwangmyeong, Deokan-ro, South Korea
View Trial DetailsNCT00038727
Body Weight, CVD
Rockville, Maryland, United States
View Trial DetailsNCT06029517
Arterial Occlusive Diseases, Arteriosclerosis
Buffalo, New York, United States
View Trial DetailsNCT01934660
Cardiovascular Disease, Cardiovascular Diseases
Bethesda, Maryland, United States
View Trial Details