Servicio de Endocrinología y Nutrición, Hospital Clínico Universitario de Valencia
Valencia, 46010, Spain
NCT Number: NCT01714895
Achieving near-normoglycaemia has been established as the main objective for most patients with diabetes. However, it is well known that intensification of treatment is associated with an increase in the frequency of hypoglycemia, especially in the context of insulin therapy. The burden of hypoglycemia in terms of psychological implications, morbidity and even mortality, explains why it has been defined as the main limiting factor to achievement of good metabolic control.
Continuous subcutaneous glucose monitoring (CGM) devices have been claimed to be useful in hypoglycemia detection/prevention, allowing theoretically for safer intensification of therapy in diabetic patients. However, accuracy of CGM devices, especially in the hypoglycemic range, raises some concerns. In fact, commercially available CGM devices estimate plasma glucose from measurements in the interstitial fluid and not in plasma. However, the relationship between plasma and interstitial glucose is not fully understood, especially under dynamic conditions, and this may explain the poor CGM performance during rapid changes in blood glucose and hypoglycemia.
In this project, the relationship between plasma and interstitial glucose will be evaluated under conditions of normal glucose concentrations and hypoglycemia. Experiments will be performed to assess the role, if any, of different plasma insulin concentrations on the accuracy of CGM.
All the information obtained may be relevant to the improvement of the ability of CGM devices to detect hypoglycemia and hypoglycemic risk.
Looking for future studies?
Notify Me18 year–60 year
All sexes
Interventional
Not applicable
Valencia, 46010, Spain
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.
Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.
Time frame: 330 minutes
CGM devices are electrochemical sensors. Currently available sensors are placed in the subcutaneous tissue where they react with the glucose of the interstitial fluid giving a current signal (in nano Amperes), which is proportional to glucose concentrations. The current signal is then filtered and transformed into plasma glucose estimations through a calibration algorithm built into the CGM device.
In this study we will analyze the current signal from CGM obtained under two different experimental conditions:
This will allow for establishing the role, if any, of different plasma insulin concentrations on the accuracy of CGM to identify plasma glucose falls resulting in hypoglycaemia.
Since CGM sensors can exhibit different sensitivities to glucose concentrations, the current signal will be normalized before statistical analysis
Fundación para la Investigación del Hospital Clínico de Valencia
Other
New Strategies for Automated Glycaemic Control: the Issue of Continuous Glucose Monitoring Accuracy Under Hypoglycaemic Conditions
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04614168
Autoimmune Diseases, Diabetes Mellitus
Edinburgh, United Kingdom
View Trial DetailsNCT05616273
Autoimmune Diseases, Diabetes Mellitus
Exeter, Devon, United Kingdom
View Trial DetailsNCT05379686
Autoimmune Diseases, Behavior
Herlev, Denmark
View Trial DetailsNCT05378672
Autoimmune Diseases, Diabetes Mellitus
Atlanta, Georgia, United States
View Trial Details