Faculty of Biomedical Engineering, Czech Technical University in Prague
Kladno, 27201, Czechia
NCT Number: NCT06369090
Twenty-two healthy female volunteers were monitored using electrical impedance tomogprahy (EIT) in two different situations. First, with electrode belt placed over the breast tissue and second, with the electrode belt under the breast tissue. The EIT measurement was performed with adjunct spirometric measurement. The changes of regional ventilation related to the breast tissue and the size of breast tissue were evaluated.
Looking for future studies?
Notify Me18 year–50 year
Female
Interventional
Not applicable
Kladno, 27201, Czechia
The EIT measurement (PulmoVista 500, Dräger Medical, Germany) was performed in supine lying position. On every subject, an electrode belt of proper size based on the measured thorax circumference was placed. It was crucial to find a right position where it was possible to place the electrode belt over the breast tissue and then underneath the breast tissue, without changing the level and the position of the electrode belt on the thorax. Also, the position of the belt had to be between 4th and 6th intercostal space, as recommended by the manufacturer. The change of the position of the belt relative to the breast tissue was performed in a lying position by a slight pull of the buckled belt away from the subject´s body, so the subject could relocate the breast tissue over the belt using her own hands into the position as desired by the researcher.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Non-invasive, radiation-free bedside lung monitoring using small alternating currents with no known side effects.
Time frame: 30 minutes
The dependence of the regional ventilation evaluated by EIT on the breast size on the regional ventilation was not confirmed.
Time frame: 30 minutes
Recommendation to include women in studies and clinical applications that use electrical impedance tomography for lung monitoring
Czech Technical University in Prague
Other
Effect of Women Breast Size on Lung Monitoring Using Electrical Impedance Tomography
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.