Skip to main content
OpenTrials
Completed

NCT Number: NCT03699397

EEG Controlled Triage in the Ambulance for Acute Ischemic Stroke

Endovascular thrombectomy (EVT) is the standard treatment for patients with a large vessel occlusion (LVO) stroke. Direct presentation of patients with an LVO to a comprehensive stroke center (CSC) reduces onset-to-treatment time by approximately an hour and thereby improves clinical outcome. However, a reliable tool for prehospital LVO-detection is currently not available. Previous electroencephalography (EEG) studies have shown that hemispheric hypoxia quickly results in slowing of the EEG-signal. Dry electrode EEG caps allow reliable EEG measurement in less than five minutes. We hypothesize that dry electrode EEG is an accurate and feasible diagnostic test for LVO in the prehospital setting.

ELECTRA-STROKE is a diagnostic pilot study that consists of four phases. In phases 1, 2 and 3, technical and logistical feasibility of performing dry electrode EEGs are tested in different in-hospital settings: the outpatient clinic (sample size: max. 20 patients), Neurology ward (sample size: max. 20 patients) and emergency room (sample size: max. 300 patients), respectively. In the final phase, ambulance paramedics will perform dry electrode EEGs in 386 patients with a suspected stroke. The aim of the ELECTRA-STROKE study is to determine the diagnostic accuracy of dry-electrode EEG for diagnosis of LVO-a stroke when performed by ambulance personnel in patients with a suspected AIS. Sample size calculation is based on an expected specificity of 70% and an incidence of LVO stroke of 5%.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Amsterdam University Medical Centers, location AMC

Amsterdam, North Holland, 1105AZ, Netherlands

About this study

RATIONALE

Endovascular thrombectomy (EVT) is standard treatment for acute ischemic stroke (AIS) if there is a large vessel occlusion in the anterior circulation (LVO-a). Because of its complexity, EVT is performed in selected hospitals only. Currently, approximately half of EVT eligible patients are initially admitted to hospitals that do not provide this therapy. This delays initiation of treatment by approximately an hour, which decreases the chance of a good clinical outcome. Direct presentation of all patients with a suspected AIS in EVT capable hospitals is not feasible, since only approximately 7% of these patients are eligible for EVT. Therefore, an advanced triage method that reliably identifies patients with an LVO-a in the ambulance is necessary. Electroencephalography (EEG) may be suitable for this purpose, as preliminary studies suggest that slow EEG activity in the delta frequency range correlates with lesion location on cerebral imaging. Use of dry electrode EEG caps will enable relatively unexperienced paramedics to perform a reliable measurement without the EEG preparation time associated with 'wet' EEGs. Combined with algorithms for automated signal analysis, we expect the time of EEG recording and analysis to eventually be below five minutes, which would make stroke triage in the ambulance by EEG logistically feasible.

HYPOTHESIS

We hypothesize that EEG accurately identifies the presence of an LVO-a stroke in patients with a suspected AIS when applied in the ambulance.

OBJECTIVE

To determine the diagnostic accuracy of dry-electrode EEG for diagnosis of LVO-a stroke when performed by ambulance personnel in patients with a suspected AIS.

STUDY DESIGN

This diagnostic study consists of four phases:

Phase 1: Optimization of measurement time and software settings of the dry electrode cap EEG in a non-emergency setting in patients in whom a regular EEG is/will be performed for standard medical care. Sample size: maximum of 20 patients.

Phase 2: Optimization of measurement time and software settings of the dry electrode cap EEG in patients close to our target population in a non-emergency setting. Sample size: maximum of 20 patients.

Phase 3: Validation of several existing algorithms and development of one or more new algorithms for LVO-a detection, as well as optimization of logistics and software settings of the dry electrode EEG cap in patients close to our target population in an in-hospital emergency setting. Sample size: maximum of 300 patients.

Phase 4: Validation of several existing algorithms and algorithms developed in phase 3 for LVO-a detection in patients with a suspected AIS in the ambulance, as well as assessment of technical and logistical feasibility of performing EEG with dry electrode caps in patients with a suspected AIS in the ambulance. Sample size: maximum of 386 patients.

STUDY POPULATION

Phase 1: Patients in the outpatient clinic of the Clinical Neurophysiology department of the AMC, in whom a regular EEG has been/will be performed for standard medical care.

Phase 2: Patients with an AIS admitted to the Neurology ward of the coordinating hospital with an LVO-a (after reperfusion therapy).

Phase 3: Patients with a suspected AIS in the emergency room (ER) of the coordinating hospital (before endovascular treatment).

Phase 4: Patients with a suspected AIS in the ambulance.

INTERVENTION

Performing a dry electrode cap EEG (in phase 1 in the outpatient clinic, in phase 2 during hospital admission, in phase 3 in the ER and in phase 4 in the ambulance).

MAIN END POINTS

Primary end point: the diagnostic accuracy of dry electrode cap EEG to discriminate LVO-a stroke from all other strokes and stroke mimics in the prehospital setting (study phase 4) expressed as the area under the receiver operating characteristics (ROC) curve of the theta/alpha ratio.

Secondary end points:

  • Sensitivity, specificity, PPV and NPV of the theta/alpha ratio, and test characteristics of other existing EEG data based algorithms for LVO-a detection (e.g. Weighted Phase Lag Index, delta/alpha ratio);
  • Logistical and technical feasibility of paramedics performing dry electrode cap EEG in the ambulance in suspected AIS patients;
  • Developing one or more novel EEG data based algorithms with an optimal diagnostic accuracy for LVO-a detection in suspected AIS patients with ambulant dry electrode cap EEG.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

STUDY PHASE 1

Inclusion criteria

  • Age of 18 years or older;
  • Patient is in the outpatient clinic of the Clinical Neurophysiology department of the AMC, because a regular EEG has been/will be performed on him/her for standard medical care;
  • Written informed consent by patient.

Exclusion criteria

  • Injury or active infection of electrode cap placement area.

STUDY PHASE 2

Inclusion criteria

  • A diagnosis of acute ischemic stroke caused by a large vessel occlusion in the anterior circulation (intracranial carotid artery or proximal (M1/M2) middle cerebral artery confirmed by neuro-imaging (CTA or MRA);
  • Stroke onset <72 hours before expected time of performing EEG;
  • Age of 18 years or older;
  • Written informed consent by patient or legal representative.

Exclusion criteria

  • Injury or active infection of electrode cap placement area.

STUDY PHASE 3

Inclusion criteria

  • Suspected acute ischemic stroke, as judged by the paramedic presenting the patient to the ER or known AIS with an LVO-a;
  • Onset of symptoms or, if onset was not witnessed, last seen well <24 hours ago;
  • Age of 18 years or older;
  • Written informed consent by patient or legal representative (deferred).

Exclusion criteria

  • Injury or active infection of electrode cap placement area.

STUDY PHASE 4

Inclusion criteria

  • Suspected acute ischemic stroke as judged by the attending paramedic;
  • Onset of symptoms or, if onset not witnessed, last seen well <24 hours ago;
  • Age of 18 years or older;
  • Written informed consent by patient or legal representative (deferred).

Exclusion criteria

  • Injury or active infection of electrode cap placement area.

Treatment and study plan

Dry electrode EEG

Diagnostic Test

A single dry electrode electroencephalography (EEG) will be performed in each patient that is included in this study. To do this, the investigators will use the WaveguardTM dry electrode EEG cap and compatible eegoTM amplifier, developed by ANT Neuro B.V. Netherlands and both CE marked as medical devices in the European Union (see appendices 1 and 2). Both products will be used within the intended use as described in the user manuals. The dry electrode cap is put on the patients head and records the EEG signal for several minutes; the amplifier is used to amplify the EEG signal and reduce artefacts.

Primary outcomes

  1. The diagnostic accuracy of dry electrode cap EEG to discriminate LVO-a stroke in the prehospital setting expressed as the area under the receiver operating characteristics (ROC) curve of the theta/alpha ratio.

    Time frame: The presence or absence of an LVO-a will be assessed based on CT angiography data obtained at the emergency department (within 24 hours after inclusion in the study). EEG data will be collected at baseline.

    The diagnostic accuracy of dry electrode cap EEG to discriminate LVO-a stroke from all other strokes and stroke mimics in the prehospital setting (study phase 4) expressed as the area under the receiver operating characteristics (ROC) curve of the theta/alpha ratio.

Secondary outcomes

  1. Sensitivity of dry electrode EEG for diagnosis of LVO-a

    Time frame: The presence or absence of an LVO-a will be assessed based on CT angiography data obtained at the emergency department (within 24 hours after inclusion in the study). EEG data will be collected at baseline.

    Sensitivity of the theta/alpha ratio, and test characteristics of other existing EEG data based algorithms for LVO-a detection (e.g. Weighted Phase Lag Index, delta/alpha ratio).

  2. Specificity of dry electrode EEG for diagnosis of LVO-a

    Time frame: The presence or absence of an LVO-a will be assessed based on CT angiography data obtained at the emergency department (within 24 hours after inclusion in the study). EEG data will be collected at baseline.

    Specificity of the theta/alpha ratio, and test characteristics of other existing EEG data based algorithms for LVO-a detection (e.g. Weighted Phase Lag Index, delta/alpha ratio).

  3. Positive predictive value of dry electrode EEG for diagnosis of LVO-a

    Time frame: The presence or absence of an LVO-a will be assessed based on CT angiography data obtained at the emergency department (within 24 hours after inclusion in the study). EEG data will be collected at baseline.

    Positive predictive value of the theta/alpha ratio, and test characteristics of other existing EEG data based algorithms for LVO-a detection (e.g. Weighted Phase Lag Index, delta/alpha ratio).

  4. Negative predictive value of dry electrode EEG for diagnosis of LVO-a

    Time frame: The presence or absence of an LVO-a will be assessed based on CT angiography data obtained at the emergency department (within 24 hours after inclusion in the study). EEG data will be collected at baseline.

    Negative predictive value of the theta/alpha ratio, and test characteristics of other existing EEG data based algorithms for LVO-a detection (e.g. Weighted Phase Lag Index, delta/alpha ratio).

  5. Technical feasibility of performing dry electrode EEGs in the ambulance

    Time frame: Feedback on technical issues by the paramedic that performs the EEG and by the EEG-expert, will be collected directly at arrival in the emergency department (within 24 hours after the patient is included in the study).

    Technical feasibility of performing dry electrode cap EEGs on patients with a suspected acute ischemic stroke in the ambulance

  6. Logistical feasibility of performing dry electrode EEGs in the ambulance

    Time frame: Feedback on logistical issues by the paramedic that performs the EEG, will be collected directly at arrival in the emergency department (within 24 hours after the patient is included in the study).

    Logistical feasibility of performing dry electrode cap EEGs on patients with a suspected acute ischemic stroke in the ambulance

  7. Algorithms with an optimal diagnostic accuracy for LVO-a detection in suspected AIS patients with ambulant dry electrode cap EEG.

    Time frame: The presence or absence of an LVO-a will be assessed based on CT angiography data obtained at the emergency department (within 24 hours after inclusion in the study). EEG data will be collected at baseline.

    Developing one or more novel EEG data based algorithms with an optimal diagnostic accuracy for LVO-a detection in suspected AIS patients with ambulant dry electrode cap EEG.

Sponsors and collaborators

Lead sponsor

Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)

Other

Registry information

Acronym: ELECTRA-STROKE

Important dates

Study start
2018
Primary completion
2022
Study completion
2023
First posted
Oct 9, 2018
Registry last updated
Apr 11, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.