Fingerprint sweat drug screen
Diagnostic TestQualitative lateral-flow immunoassay analysis of fingerprint sweat specimen
NCT Number: NCT06618183
Prospective, open-label, single-dose, randomized, multi-site performance evaluation of the accuracy and usability Intelligent Fingerprinting Drug Screening System in the hands of intended users: "operators" and "donors."
Interested in participating?
Request Info18 year–65 year
All sexes
Observational
CenExel iResearch, Decatur, Georgia, United States
This study will compare positive and negative test results obtained from fingerprint sweat using the Intelligent Fingerprinting DSC-7 Drug Screening Cartridge versus opiates in sweat collected via the Fingerprint Collection Kit for Laboratory Analysis, within healthy males and non-pregnant females.
The primary objective is to compare the sensitivity, specificity, and accuracy of the Intelligent Fingerprinting Drug Screening System to the validated confirmatory LC-MS/MS technique.
The secondary objectives are to assess drug screening system usability and reproducibility.
"Operators" comprise nine (9) adults: male and female subjects who operate the drug screening system. These subjects are not administered any drug. "Operators" represent end users who utilize the screening system in a laboratory setting. "Operators" will be provided the Intelligent Fingerprinting Drug Screening Guide, Confirmatory Laboratory Analysis Guide, and Quick Reference Guide, and be trained in system use. After receiving and reviewing all materials, the "operator" will complete a Training Exam to determine if he or she may proceed with testing.
"Donors" comprise 135 healthy adults: male and non-pregnant female subjects who provide fingerprint sweat specimens. Subjects will be randomized into the blocks administered 0, 30, or 60mg of codeine sulfate in tablet form. "Donors" represent end users either applying for employment or working at an organization that conducts pre-employment and/or periodic workplace drug screening.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Qualitative lateral-flow immunoassay analysis of fingerprint sweat specimen
Laboratory analysis of fingerprint sweat specimen by liquid chromatography/tandem mass spectrometry
Time frame: 30 minutes to 3 hours
Comparison of the sensitivity, specificity, and accuracy of the Intelligent Fingerprinting Drug Screening System (qualitative lateral-flow immunoassay) to the validated confirmatory LC-MS/MS technique
Time frame: 72 hours
Usability of the Intelligent Fingerprinting Drug Screening System based on scores obtained by operators on written training exam (Drug Screening Process and Laboratory Process Exam). Training exam score must meet or exceed 85% for each operator.
Intelligent Bio Solutions Inc.
Industry
User Performance Evaluation of the Intelligent Fingerprinting Drug Screening System: Method Comparison and Usability
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07601425
Arteriovenous Malformations, Cardiovascular Abnormalities
Boston, Massachusetts, United States
View Trial DetailsNCT07728513
AI-Assisted Diagnosis, Disease
Lahore, Punjab Province, Pakistan
View Trial DetailsNCT07725510
Disease, Pathologic Processes
Guangzhou, Guangdong, China
View Trial DetailsNCT06203119
Cancer, Disease
Beijing, Beijing Municipality, China
View Trial Details