NCT Number: NCT07735403

Dozee Validation for Contactless Heart Rate and Respiration Monitoring in Clinical Settings

The Dozee VS system (K221555) is a contactless device that tracks heart rate and breathing rate using a sensor sheet placed under the mattress. It captures body vibrations and converts them into vital signs, which can be monitored remotely in real time. The system also supports add-ons like a blood pressure cuff, pulse oximeter, and thermometer, offering a complete monitoring solution. This study aims to validate the updated AI-powered heart rate and improved breathing rate algorithms across different sensor sheet types to confirm clinical accuracy

Recruiting

Interested in participating?

Request Info

Key information

Age range

22 year–100 year

Sex eligibility

All sexes

Study type

Observational

Primary location

Oak Village Healthcare

Lake Jackson, Texas, 77566, United States

Location status: Recruiting

Location contact

Lindsay Darhtard, LVN LNFA

CONTACT

[email protected]

979-297-0425

About this study

The Dozee VS system (K221555), developed by Turtle Shell Technologies, is a non-invasive, contactless device that continuously monitors heart rate (HR) and breathing rate (RR). A thin sensor sheet placed under the mattress detects tiny body vibrations from the heart and lungs, which are processed by smart algorithms and displayed on a secure online dashboard for real-time monitoring.Unlike traditional methods such as ECGs, nasal tubes, or thermistors-which are invasive, require skilled placement, and are uncomfortable for long-term use-Dozee is simple, comfortable, and doesn't rely on patient compliance like wearables do. The system can also connect to add-on devices, including a blood pressure cuff, pulse oximeter, and thermometer, giving a full picture of a patient's vital signs and helping spot early signs of deterioration. Hardware upgrades now include dual-sensor sheets that improve accuracy across different mattresses and patient types. The new HR algorithm uses AI/ML for better accuracy, while the improved RR algorithm performs more reliably even with movement or weak signals.Study Goal: This study aims to validate the updated algorithms (Gen 2) for HR and RR across the three sensor sheet types (DS, S1, and S1-I) to ensure they meet clinical accuracy standards

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Adults Male or female. (≥ 22 years of age)
  • Provide written informed consent.
  • Weight over 40kg/88.18lbs

Exclusion criteria

  • Are connected to a device that may interfere with the device monitoring in this study.
  • Are receiving any bedside care which may be incompatible with the study procedures.
  • A likely need to receive or undergo a procedure during the testing period.
  • Cannot accept a nasal cannula, or have an ECG leads placed on the chest.
  • Have a significant medical condition in the judgement of the investigator, which may compromise the study testing procedures.
  • Prisoners

Treatment and study plan

Telemetry monitoring

Device

All participants will be on standard telemetry monitoring for Heart Rate, Respiration Rate and movements using the gold standard (Electrocardiography and Capnography respectively)

Primary outcomes

  1. Mean Absolute Error (MAE) of Dozee's Heart Rate a from the Gold Standard ECG

    Time frame: 6 months

    Heart Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Heart Rate will be calculated

  2. Mean Absolute Error (MAE) of Dozee's Respiratory Rate from the Gold Standard Capnography

    Time frame: 6 months

    Respiratory Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Respiratory Rate will be calculated.

Study contacts

Contact information is provided by the study sponsor or research team.

Lindsay Darthard, LVN LNFA

CONTACT

[email protected]

979-297-0425

Sponsors and collaborators

Lead sponsor

Turtle Shell Technologies Pvt. Ltd.

Industry

Registry information

Official study title

Dozee System Validation for Contactless Heart Rate and Respiration Monitoring in Clinical Settings

Important dates

Study start
2026
Primary completion
2026
Study completion
2026
First posted
Jul 30, 2026
Registry last updated
Jul 30, 2026

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.