Development of Advanced Double Intravenous Vasopressor Automated (ADIVA) System

Maternal hypotension during spinal anaesthesia for Caesarean delivery may bring various adverse effects. It is therefore important to closely monitor hypotension during spinal anaesthesia, however current technology does not allow the blood pressure monitoring to respond in timely manner should there be any occurrence of hypotension. The investigators developed an advanced double-intravenous vasopressor automated system (ADIVA) so as to tackle this issue via novel algorithm to control blood pressure more rigorously with more stable haemodynamic profiles.

Active, Not Recruiting

This study is active but is not currently recruiting participants.

Notify Me

Key information

Age range

21 year–50 year

Sex eligibility

Female

Study type

Interventional

Phase

Not applicable

Primary location

KK Women's and Children's Hospital

Singapore, 229899

About this study

The incidence of maternal hypotension during spinal anaesthesia for Caesarean delivery is reported to be as high as 70-90%. The potential adverse effects resulting from maternal hypotension are nausea, vomiting, cardiac dysfunction, foetal acidosis and hypoxia.

Hypotension during spinal anaesthesia is currently detected using an intermittent blood pressure (BP) monitor cycling and reactive administration of vasopressors upon detection. The use of conventional non-invasive BP monitoring is limited by the time required to inflate and to deflate the cuffs commonly applied to the arm, which subsequently leads to a failure to react in a timely manner to BP changes when they occur.

The DIVA system (double-intravenous vasopressor automated system) previously developed helped tackle the limitations associated with the conventional management of hypotension in Caesarean sections under spinal anaesthesia. In a recent randomized controlled trial the DIVA system achieved less incidence of maternal hypotension compared to conventional management, however the side effects in mother and baby were still not fully eliminated.

In this proposed study a new algorithm for an advanced DIVA (ADIVA) system will be developed to control BP more rigorously by detecting BP with more stable haemodynamic profiles and thereby improve patient outcomes. The clinical trial will be conducted in 2 phases. The first phase included the use of ADIVA in 76 subjects who undergo spinal anaesthesia for Caesarean delivery to test its safety and efficacy. The second phase was a randomised controlled trial to investigate the use of ADIVA in comparison to the existing DIVA system in 97 patients undergoing spinal anaesthesia for Caesarean section.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Healthy adult (American Society of Anesthesiologists physical status 1 and 2) parturients;
  • With singleton full-term pregnancy;
  • The indication for an elective cesarean delivery;
  • The use of spinal anaesthesia for cesarean delivery;
  • Anthropometric profile within the following range: age 21-50 years old, weight 40-90 kg and height of 145-170 cm.

Exclusion criteria

  • Obstetric (e.g. pre-eclampsia, premature rupture of amniotic membranes for more than 48 hours, gestational diabetes on insulin, pregnancy-induced hypertension on medication) and uncontrolled medical (e.g. cardiac disease) complications;
  • Contraindication to spinal anaesthesia and/or allergy to opioids.

Treatment and study plan

Phenylephrine

Drug

Phenylephrine will be administered with conditions as below:

  • When systolic blood pressure is between 90-110% of baseline, 25mcg phenylephrine will be given either in infusion or bolus if heart rate ≥ 60 beats/min;
  • When systolic blood pressure is less than 90% of baseline, 25 to 75mcg phenylephrine will be given either in infusion or bolus if heart rate ≥ 60 beats/min.

Other names: Vasopressor

Ephedrine

Drug

Ephedrine will be administered with conditions as below:

  • When systolic blood pressure is between 90-110% of baseline, 2mg ephedrine will be given either in infusion or bolus if heart rate < 60 beats/min;
  • When systolic blood pressure is less than 90% of baseline, 2 to 6mg ephedrine will be given either in infusion or bolus if heart rate < 60 beats/min.

Other names: Vasopressor

ADIVA System

Device
  • When systolic blood pressure (SBP) is 100-110% of baseline, 25mcg phenylephrine (infusion) is given if heart rate ≥ 60 beats/min, and 2mg ephedrine (infusion) is given if heart rate < 60 beats/min;
  • When SBP is 90-100% of baseline, 25mcg phenylephrine (infusion or bolus) is given if heart rate ≥ 60 beats/min, and 2mg ephedrine (infusion or bolus) is given if heart rate < 60 beats/min;
  • When SBP is 80-90% of baseline, 50mcg phenylephrine (infusion or bolus) is given if heart rate ≥ 60 beats/min, and 4mg ephedrine (infusion or bolus) is given if heart rate < 60 beats/min;
  • When SBP < 80% of baseline, 75mcg phenylephrine (infusion or bolus) is given if heart rate ≥ 60 beats/min, and 6mg ephedrine (infusion or bolus) is given if heart rate < 60 beats/min.

Other names: Advanced Double Intravenous Vasopressor Automated System

DIVA System

Device

DIVA system works as below:

  • When systolic blood pressure is between 90 to 100% of baseline, 25mcg phenylephrine (bolus) will be given if heart rate ≥ 60 beats/min, whereas 2mg ephedrine (bolus) will be given if heart rate < 60 beats/min;
  • When systolic blood pressure is less than 90% of baseline, 50mcg phenylephrine (infusion or bolus) will be given if heart rate ≥ 60 beats/min, whereas 4mg ephedrine (infusion or bolus) will be given if heart rate < 60 beats/min.

Other names: Double Intravenous Vasopressor Automated System

Primary outcomes

  1. Number of events of Hypotension occurrence

    Time frame: 1 day (during cesarean delivery)

    Systolic blood pressure less than 80% of baseline systolic blood pressure

Secondary outcomes

  1. Number of events of Hypertension occurrence

    Time frame: 1 day (during cesarean delivery)

    Systolic blood pressure more than 120% of baseline systolic blood pressure

  2. Number of events of Nausea and vomiting occurrence

    Time frame: 1 day (during and after cesarean delivery)

    Nausea and vomiting during and after cesarean delivery

Sponsors and collaborators

Lead sponsor

KK Women's and Children's Hospital

Other Gov

Registry information

Official study title

Development of Advanced Double Intravenous Vasopressor Automated (ADIVA) System With Improved Hemodynamic Trend Control During Spinal Anaesthesia for Caesarean Section

Acronym: ADIVA

Important dates

Study start
2018
Primary completion
2025
Study completion
2026
First posted
Aug 8, 2018
Registry last updated
Oct 9, 2024

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.