Hypo Sense is a non- invasive method for detection of hypoglycemia. The Hypo Sense combines an array of non-invasive sensors which monitors the patient's physiological parameters (heart & respiration rate, perspiration, skin temperature and arm motion) designed as a wrist watch device.
The Hypo sense is intended for monitoring symptoms of hypoglycemia in diabetic patients in hospital environment among type 1 and type 2 diabetes adults as an adjunctive device to reference methods
The proposed study will be consisting of two main segments:
The primary aim of segment 1 of the study is data collection and calibration of the Hypo Sense sensor prototype compared to standard invasive reference glucometer.
The primary aim of segment 2 of the study is to validate the Hypo Sense prototype performance in detecting hypoglycemic events.
During the first segment of the study we intend to collect in parallel measurements of blood glucose using reference method (capillary glucometer) and continuous data generated by the non- invasive study device during approximately 4 hours, in which a hypoglycemic event will be induced.
The reference and study device data will be analyzed using multivariate regression model to formulate a calibration algorithm model. This model will translate the set of the physiological recorded parameters into detection of hypoglycemic events.
During the second segment of the study we intend to evaluate the validity of the Hypo Sense sensor ability to detect hypoglycemic events compared to standard invasive reference method (capillary glucometer).
Looking for future studies?
Notify Me18 year–65 year
All sexes
Interventional
Not applicable
Petah Tikva, 49202, Israel
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Parallel measurements of capillary blood glucose using reference methods (both capillary glucometer and continuous sensor) and data generated by the non- invasive study device during approximately 4 hours, in which a hypoglycemic event will be induced.
Time frame: At final visit ( week 4 for participants at segment 1 of the study and week 6 for participants at segment 2 of the study)
Sensitivity of the Hypo Sense-the proportion of hypoglycemia events detected by the Hypo Sense sensor from total number of hypoglycemic events detected by reference capillary glucometer
Time frame: At final visit ( week 4 for participants at segment 1 of the study and week 6 for participants at segment 2 of the study)
Positive Predictive Value of the hypo sense- the proportion of validated ("true") hypoglycemia events detected by the hypo sense sensor from total alerts detected by the Hypo-Sense sensor
Rabin Medical Center
Other
Development and Assessment of Feasibility of Non-invasive Multiple Sensor Hypo-Sense Sensor as a Tool for Detection of Hypoglycemia - Exploratory Study for Prototype Rev 1
Acronym: Hypo-Sense
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00542334
Autoimmune Diseases, Diabetes Mellitus
Ottawa, Ontario, Canada
View Trial DetailsNCT05488119
Autoimmune Diseases, Diabetes Mellitus
Stanford, California, United States
View Trial DetailsNCT07696780
Autoimmune Diseases, Diabetes Mellitus
Bern, Canton of Bern, Switzerland
View Trial DetailsNCT05610111
Autoimmune Diseases, Diabetes Mellitus
Charlottesville, Virginia, United States
View Trial Details