Completed

NCT Number: NCT02225379

Development and Assessment of Feasibility of Non-invasive Multiple Sensor Hypo-Sense as a Tool for Detection of Hypoglycemia

Hypo Sense is a non- invasive method for detection of hypoglycemia. The Hypo Sense combines an array of non-invasive sensors which monitors the patient's physiological parameters (heart & respiration rate, perspiration, skin temperature and arm motion) designed as a wrist watch device.

The Hypo sense is intended for monitoring symptoms of hypoglycemia in diabetic patients in hospital environment among type 1 and type 2 diabetes adults as an adjunctive device to reference methods

The proposed study will be consisting of two main segments:

The primary aim of segment 1 of the study is data collection and calibration of the Hypo Sense sensor prototype compared to standard invasive reference glucometer.

The primary aim of segment 2 of the study is to validate the Hypo Sense prototype performance in detecting hypoglycemic events.

During the first segment of the study we intend to collect in parallel measurements of blood glucose using reference method (capillary glucometer) and continuous data generated by the non- invasive study device during approximately 4 hours, in which a hypoglycemic event will be induced.

The reference and study device data will be analyzed using multivariate regression model to formulate a calibration algorithm model. This model will translate the set of the physiological recorded parameters into detection of hypoglycemic events.

During the second segment of the study we intend to evaluate the validity of the Hypo Sense sensor ability to detect hypoglycemic events compared to standard invasive reference method (capillary glucometer).

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–65 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Schneider Children's Medical Center

Petah Tikva, 49202, Israel

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Signing an inform consent form prior to any trial related procedure
  • Type 1 diabetes diagnosed at least 12 months prior to study inclusion
  • Age > 18 years old

Exclusion criteria

  • Participating in other device or drug study
  • Any significant disease or condition, including psychiatric disorders that in the opinion of the investigator is likely to affect patient's compliance or ability to complete the study
  • Patients with one or more of the following diseases: malignancy, myocardial insufficiency, nephrologic disease or any other chronic disease
  • Chronic skin problem in the lower inner arm
  • Pregnant or breast feeding women

Treatment and study plan

Hypo-Sense (non invasive sensor)

Device

Parallel measurements of capillary blood glucose using reference methods (both capillary glucometer and continuous sensor) and data generated by the non- invasive study device during approximately 4 hours, in which a hypoglycemic event will be induced.

Primary outcomes

  1. Sensitivity

    Time frame: At final visit ( week 4 for participants at segment 1 of the study and week 6 for participants at segment 2 of the study)

    Sensitivity of the Hypo Sense-the proportion of hypoglycemia events detected by the Hypo Sense sensor from total number of hypoglycemic events detected by reference capillary glucometer

  2. Positive Predictive Value of the hypo-Sense

    Time frame: At final visit ( week 4 for participants at segment 1 of the study and week 6 for participants at segment 2 of the study)

    Positive Predictive Value of the hypo sense- the proportion of validated ("true") hypoglycemia events detected by the hypo sense sensor from total alerts detected by the Hypo-Sense sensor

Sponsors and collaborators

Lead sponsor

Rabin Medical Center

Other

Collaborators

  • Night Sense Ltd

Registry information

Official study title

Development and Assessment of Feasibility of Non-invasive Multiple Sensor Hypo-Sense Sensor as a Tool for Detection of Hypoglycemia - Exploratory Study for Prototype Rev 1

Acronym: Hypo-Sense

Important dates

Study start
2014
Primary completion
2016
Study completion
2016
First posted
Aug 26, 2014
Registry last updated
Jan 11, 2017

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.