Guys and St Thomas' NHS Foundation Trust
London, SE1 7EH, United Kingdom
NCT Number: NCT03495505
This feasibility study will use CT scanning to identify the optimal location for placement of the WiSE-CRT system.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
London, SE1 7EH, United Kingdom
Heart failure (HF) is estimated to affect at least 10% of adults aged over 70 years old, with a 12-month all-cause mortality rate of 7% in stable patients. Cardiac resynchronisation therapy (CRT) is an important intervention for patients with severe left ventricular (LV) systolic impairment and helps to improve well-being and reduce morbidity and mortality. CRT is achieved by placing endocardial pacing leads into the right atrium and right ventricle and then placing a third pacing lead through the coronary sinus (CS) to enable epicardial LV pacing, thus achieving ventricular resynchronisation. Although this is a successful therapy, overall 30-40% of patients will fail to respond. Additionally, even in those eligible for CRT, approximately 8-10% of patients cannot have it implanted due to anatomical abnormalities such as venous occlusion, inappropriate CS targets, diffuse scar resulting in inappropriately high pacing thresholds or phrenic nerve stimulation. In these circumstances other avenues to achieve biventricular pacing needs to be sought.
The wireless CRT system (WiSE-CRT, EBR Systems) has been developed to overcome these issues. It uses an endocardial LV electrode to achieve biventricular pacing negating the need to implant a pacing lead through the CS. There are several advantages of endocardial LV pacing such as a greater selection of pacing sites, possibility of lower pacing outputs compared with conventional leads in the CS and it appears to be a more physiological way to pace. The WiSE-CRT system is used in conjunction with a single or dual-chamber pacemaker and is made up of several components. A transmitter is implanted subcutaneously, attached to a battery. This detects right ventricular pacing from the co-implant and then delivers ultrasonic energy which is received by an electrode placed in the LV endocardium to enable biventricular pacing. The transmitter must be placed in a position with an adequate "acoustic window," which requires a line from the transmitter to the LV that is free of significant tissue or bone. Additionally, the endocardial electrode must be placed in close proximity to the transmitter. In a study of 35 patients, the WiSE-CRT system was successful implanted in 97.1% of cases, with 97% achieving biventricular pacing at 1 month and 84.8% showing an improvement in the clinical composite score at 6 months. These results are particularly encouraging given the device was used in patients who had failed conventional CRT, representing a difficult patient group.
The major advantage of the WiSE-CRT system is that the LV electrode can be placed anywhere within the ventricle. Studies have shown that patient outcomes are improved by pacing at a site which avoids ventricular scar and targets an area of latest mechanical activation. These sites will vary according to the aetiology underlying HF. In addition, using conventional fluoroscopy to guide pacing lead implants has been shown to be inaccurate when compared with computed tomography (CT). Our group have previously shown that magnetic resonance imaging (MRI) can be overlaid on fluoroscopic images to help guide epicardial pacing to a specific location and improve outcomes. Many patients undergoing a WiSE-CRT implant will be unable to have a MRI due to their previous pacemaker implant. However, guidance may still be achieved using CT scanning.
The proximity and orientation of the transmitter to the endocardial electrode is important in determining the battery life of the WiSE-CRT system. Yeh et al. showed that a large proportion of patients have at least two suitable acoustic windows for placement of the transmitter. This provides the operator with more opportunities to carefully select an ideal location. However, once implanted the endocardial electrode must be placed within the acoustic window which limits the number of sites the electrode can then be placed.
Identifying the best location for both the transmitter and endocardial electrode is essential to target the most viable myocardium and improve patient outcomes. Ideally this should be achieved before the patient has undergone any procedure since implanting the transmitter limits the potential locations for the electrode. We believe CT guidance can provide satisfactory information to optimise the location for both the transmitter and electrode which will increase the patient response rate and improve outcomes.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Intervention will involve placement of the WISE-CRT system which consists of a transmitter, battery and electrode.
Other names: Cardiac Resynchronisation Therapy, Wireless Endocardial Pacing, WICS device
Time frame: Baseline and during procedure
Time frame: Baseline and 6 months
This measures a change in the patients symptoms from baseline to 6-months post intervention. A higher value represents a better outcome.
Time frame: Baseline and 6 months
Time frame: Baseline and 6 months
Time frame: Baseline and 6 months
This measures a change in the patients symptoms from baseline to 6-months post intervention. A higher value represents a better outcome.
Guy's and St Thomas' NHS Foundation Trust
Other
Image Optimisation and Guidance for Wireless Endocardial Cardiac Resynchronisation Therapy: The CT Guided Trial
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03446651
Cardiovascular Diseases, Heart Diseases
New Haven, Connecticut, United States
View Trial DetailsNCT05806606
Behavior, Cardiovascular Diseases
Hong Kong
View Trial DetailsNCT06847438
Cardiovascular Diseases, Heart Diseases
Atlanta, Georgia, United States
View Trial DetailsNCT07721909
Cardiovascular Diseases, Heart Diseases
Muscat, Oman
View Trial Details