NCT Number: NCT01983293
CRT Implant Strategy Using the Longest Electrical Delay for Non-left Bundle Branch Block Patients
The purpose of this study is to analyze the effect of left ventricular lead pacing location in the non-left bundle branch block (non-LBBB) heart failure patient population. The left ventricular lead pacing location will be guided by either the pacing site with the largest amount of dyssynchrony as measured by the LV electrical delay (QLV) or the physician's standard of care implant approach.
Looking for future studies?
Notify MeKey information
Conditions
Age range
18 year and older
Sex eligibility
All sexes
Study type
Interventional
Phase
Not applicable
Primary location
Glendale Adventist Medical Center, Glendale, California, United States
About this study
This is a prospective, pilot, multi-center, double-blinded, randomized post-market study to assess the effect of left ventricular lead pacing location (guided via QLV measurement vs. standard of care approach) in non-LBBB patients.
In the QLV arm the physician will:
- Assess two branches of the coronary sinus - a non-traditional vessel (inclusive of the anterior region) will be tested first and a traditional free lateral branch will be tested second for LV lead placement.
- Measure QLV for each of the four cathodes of the left ventricular lead in each branch.
- Choose the vein branch and cathode with the longest QLV measurement and program a vector based on that cathode.
In the standard of care group, the left ventricular lead placement will be carried out according to the physician's standard of care implant approach.
The impact of the left ventricular lead position will be evaluated based on the patient's response to CRT utilizing the Clinical Composite Score (cardiovascular death, heart failure hospitalizations, New York Heart Association (NYHA) class, and Patient Global Assessment). Authorized site personnel conducting the NYHA class assessment and Patient Global Assessment will be blinded to the randomization assignment and lead implant technique.
Who can participate
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
- Have non-LBBB morphology (includes complete right bundle branch block and intraventricular conduction delay with a QRS duration ≥ 120ms)
- Have the following indication per the 2013 updated American College of Cardiology Foundation/American Heart Association/Heart Rhythm Society guidelines:
- Left ventricular ejection fraction (LVEF) ≤ 35%, sinus rhythm, ischemic or non-ischemic cardiomyopathy, a non-LBBB pattern with QRS duration ≥ 120 ms, and NYHA) class III/ambulatory class IV on guideline directed medical therapy
- Receiving a new CRT implant or undergoing an upgrade from an existing implantable cardioverter defibrillator or pacemaker implant with no more than 10% right ventricular pacing
- Are 18 years or older, or of legal age to give informed consent specific to state and local law
- Ability to provide informed consent for study participation and is willing and able to comply with the prescribed follow-up tests and schedule of evaluations
Exclusion criteria
- Irreversible occlusion of venous access that will prevent placement of the CRT system either through the right or left upper extremity venous system
- Undergoing left ventricular lead placement via a surgical or epicardial approach
- Cardiomyopathy due solely to valvular disease that is not repaired/replaced
- Enrolled or intend to participate in a clinical drug and/or device study, which could confound the results of this trial as determined by St. Jude Medical, during the course of this clinical study
- LBBB: QRS width ≥ 120 ms, with predominantly negative QRS in lead V1, and upright, monophasic QRS in leads I and V6
- Incomplete right bundle branch block - intraventricular conduction delay with a QRS duration between 110 and 119ms
- Persistent or permanent atrial fibrillation
- Pacemaker dependent
- Patients who are being upgraded primarily due to right ventricular pacing
- Women who are pregnant or who plan to become pregnant during the clinical trial
- Life expectancy < 1 year
Treatment and study plan
Standard of care implant strategy
DevicePrimary outcomes
-
Number of Patients With Improved Clinical Composite Score
Time frame: 12 months
Evaluate the Clinical Composite Score (CCS) at 12 months in NLBBB patients using a standard of care vs. latest electrical delay (QLV) implant strategy. The CCS has 4 components: New York Heart Association (NYHA) functional classification, Patient Global Assessment (PGA), heart failure (HF) events, cardiovascular death. NYHA Class ranges from Class I (least severe) to Class IV (most severe); possible PGA responses are "markedly worse", "moderately worse", "slightly worse", "no change", "slightly better", "moderately better, "markedly better". CCS components were used to classify or score subjects as "IMPROVED" (at least one-class improvement in NYHA Class or improvement by PGA "moderately" or "markedly" better AND no HF events AND no cardiovascular death), or "WORSENED" (worsening in NYHA Class OR worsening by PGA "moderately" or "markedly" worse OR presence of HF events OR Cardiovascular death, or "UNCHANGED" (neither "improved" or "worsened"). Note CCS is not a numeric score.
Sponsors and collaborators
Lead sponsor
Abbott Medical Devices
Industry
Registry information
Official study title
Cardiac Resynchronization Therapy (CRT) Implant Strategy Using the Longest Electrical Delay for Non-left Bundle Branch Block Patients (ENHANCE CRT). A Prospective, Randomized, Postmarket, Pilot Study.
Acronym: ENHANCE CRT
Important dates
- Study start
- 2013
- Primary completion
- 2018
- Study completion
- 2018
- First posted
- Nov 13, 2013
- Registry last updated
- Mar 13, 2019
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Related clinical trials
Published trials that share one or more normalized conditions with this study.
Factors Associated With Response to Cardiac Resynchronization Therapy in Heart Failure Patients With Non-LBBB ECG Pattern
NCT06977217
Cardiac Resynchronization Therapy, Cardiomyopathies
Rochester, New York, United States
View Trial DetailsGoal-directed CPR Using Cerebral Oximetry
NCT04924985
Cardiovascular Diseases, Ischemia
New York, United States
View Trial DetailsRenal Transplantation and Inhaled Anesthetic Sevoflurane (SEVOREIN)
NCT00337051
Acute Kidney Injury, Chronic Disease
Bordeaux, France
View Trial DetailsPassive Limb Movement Study
NCT05935670
Brain Diseases, Cardiovascular Diseases
Milwaukee, Wisconsin, United States
View Trial Details