U-Systems, Inc.
Sunnyvale, California, 94085, United States
NCT Number: NCT01424956
To determine the impact on Interpreting Physician performance in detecting breast cancer, as defined by the area under the Receiver Operating Characteristic Curve (ROC), when Automated Breast Ultrasound (ABUS) and screening mammography (XRM) are combined, compared to screening mammography alone in asymptomatic women with >50% parenchymal density and a screening mammogram assigned a BI-RADS Assessment Category 1 (negative) or 2 (normal with benign findings).
The effect of the improved Reader performance is illustrated by plotting the ROC curves for XRM alone and XRM+ABUS. If Reader performance improves with the addition of ABUS to the XRM, the area under the curve (AUC) for XRM+ABUS (AUCXRM+ABUS) will be greater than the area under the curve for XRM Alone (AUCXRM Alone). This difference is represented as ∆AUCABUS. The null and alternative hypotheses can be formally expressed as follows:
H0: ∆AUCABUS = 0, AUCXRM+ABUS = AUCXRM Alone The null hypothesis is that Reader performance will be unchanged with the addition of ABUS to a screening mammogram assigned a BI-RADS Assessment Category of 1 or 2.
HA: ∆AUCABUS ≠ 0, AUCXRM+ABUS ≠ AUCXRM Alone The alternative hypothesis is that Reader performance will be changed with the addition of ABUS to a screening mammogram assigned a BI-RADS Assessment Category of 1 or 2. A statistically significant change will be considered equivalent to a statistically significant improvement if the estimated value of AUCXRM+ABUS is greater than that of AUCXRM Alone with statistical significance at an alpha level of .05 for a two-sided test.
Looking for future studies?
Notify MeAll sexes
Observational
Sunnyvale, California, 94085, United States
The Clinical Retrospective Reader Study is an observational case-controlled, multi-reader multi-case Receiver Operating Characteristic (ROC) study involving approximately 16 Interpreting Physicians (Readers) and approximately 200 breast screening cases which will be reviewed and of those, 164 breast screening cases will be included in the primary data analysis. All primary data analysis cases will consist of screening mammogram and Automated Breast Ultrasound image readings from the same asymptomatic females with >50% parenchymal density for whom the screening mammogram was assigned a BI-RADS Assessment Category 1 (negative) or 2 (normal with benign findings).
The Readers will review a total of 200 cases: the Non-Cancer Case Set, the Cancer Case Set, and additional Control and Supplemental Case Sets. These cases will be randomized during the reading session so each Reader will review the entire Case Set in a unique order.
The primary Receiver Operating Characteristic analysis will be performed on the Readers' Likelihood of Malignancy (LOM) ratings from their initial review of the screening mammogram alone and their final review of the screening mammogram and Automated Breast Ultrasound for the Non-Cancer Case Set and Cancer Case Set.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Automated Breast Ultrasound (ABUS) as an adjunct to digital screening mammography.
Other names: somo.V ABUS, ABUS, Whole Breast Ultrasound
Time frame: Two months
To determine the impact on Interpreting Physician (Reader) Performance in detecting breast cancer, as defined by the area under the Receiver Operating Characteristic (ROC) Curve (AUC), when ABUS and XRM are combined (XRM+ABUS), compared to XRM Alone in asymptomatic women with >50% parenchymal density (BI-RADS Composition/Density rating of 3 or 4) and a screening mammogram assigned a BI-RADS Assessment Category 1 (negative) or 2 (normal with benign findings).
Time frame: Two months
Reader sensitivity and specificity when XRM+ABUS are combined compared to XRM Alone for two different combinations of cut points.
U-Systems, Inc.
Industry
To Determine the Impact on Reader Performance, as Defined by the Area Under the Receiver Operating Characteristic Curve, When Automated Breast Ultrasound and a Screening Mammogram Are Combined, Compared to a Screening Mammogram Alone in Asymptomatic Women With >50% Parenchymal Density and a Screening Mammogram Assigned a BI-RADS Assessment Category 1 (Negative) or 2 (Normal With Benign Findings).
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT01547741
Breast Cancer, Breast Diseases
Anchorage, Alaska, United States
View Trial DetailsNCT07745946
Anxiety Disorders, Breast Cancer
Ankara, Turkey (Türkiye)
View Trial DetailsNCT00469339
Breast Cancer, Breast Diseases
Houston, Texas, United States
View Trial DetailsNCT06068257
Breast Cancer, Breast Diseases
Orlando, Florida, United States
View Trial Details