Günther Tulip® Vena Cava Filter
DeviceTemporary or permanent filter placement for the prevention of pulmonary embolism
NCT Number: NCT02046096
This prospective, multicenter, double-arm clinical study further evaluated the safety and effectiveness of Cook's commercially available inferior vena cava (IVC) filters (specifically, the Günther Tulip® Vena Cava Filter and Cook Celect® Vena Cava Filters) in patients in need of temporary or permanent IVC filter placement for the prevention of pulmonary embolism (PE).
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
The Alfred Hospital, Melbourne, Victoria, Australia
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Temporary or permanent filter placement for the prevention of pulmonary embolism
Temporary or permanent filter placement for the prevention of pulmonary embolism
Time frame: 12 months
The protocol defined technical placement success as: deployment of filter in a location suitable to provide sufficient mechanical protection against PE with no filter deformation, fracture, premature release, or clinical migration (caudal or cranial movement of a filter resulting in a surgical or endovascular intervention).
Time frame: 12 months
MAEs were defined as clinical perforation, clinical migration, clinical fracture, embolization of the filter or filter fragments to the heart or lungs, IVC thrombotic occlusion, new symptomatic DVT while the filter is indwelling, access site complications with clinical sequelae, and procedure-/device-related death.
Clinical perforation - Protrusion of filter legs through the wall of the IVC causing hemorrhage or hematoma or touching, impressing, or perforating another organ. Documented using CT.
Clinical migration - Caudal or cranial movement of a filter resulting in surgical or endovascular intervention.
Clinical fracture - A loss of structural integrity (breakage or separation) of the filter identified by imaging and associated with clinical sequelae and/or requiring intervention.
Time frame: 12 months
MAEs were defined as clinical perforation, clinical migration, clinical fracture, embolization of the filter or filter fragments to the heart or lungs, IVC thrombotic occlusion, new symptomatic DVT while the filter is indwelling, access site complications with clinical sequelae, and procedure-/device-related death.
A filter retrieval is considered a successful safety result through 12 months, regardless of when the retrieval occurred.
Time frame: 12 months
Rate of technical placement success and 12-month freedom from new symptomatic PE while a filter is indwelling for the combined patient population (i.e., the Celect and Günther Tulip strata).
The protocol defined technical placement success as: deployment of filter in a location suitable to provide sufficient mechanical protection against PE with no filter deformation, fracture, premature release, or clinical migration (caudal or cranial movement of a filter resulting in a surgical or endovascular intervention).
Time frame: 12 months
MAEs were defined as clinical perforation, clinical migration, clinical fracture, embolization of the filter or filter fragments to the heart or lungs, IVC thrombotic occlusion, new symptomatic DVT while the filter is indwelling, access site complications with clinical sequelae, and procedure-/device-related death.
Clinical perforation - Protrusion of filter legs through the wall of the IVC causing hemorrhage or hematoma or touching, impressing, or perforating another organ. Documented using CT.
Clinical migration - Caudal or cranial movement of a filter resulting in surgical or endovascular intervention.
Clinical fracture - A loss of structural integrity (breakage or separation) of the filter identified by imaging and associated with clinical sequelae and/or requiring intervention.
Time frame: 12 months
Filter interaction with IVC is defined as :
Grade 2: Filter strut is entirely outside of the IVC lumen and within the retroperitoneum as evidenced by a "halo" of retroperitoneal fat around axially viewed strut.
Grade 3: Filter strut is touching, impressing, or perforating another organ (e.g., liver, bowel, aorta, psoas muscle, vertebral body, lymph nodes).
Filter migration : Change in filter position compared to its deployed position (cranial or caudal).
Filter fracture : Any loss of structural integrity (breakage or separation) of the filter identified by imaging or autopsy. Documented by imaging or at autopsy.
Filter embolization : Post-placement movement of the filter or its components to a distant anatomic site completely out of the target zone (i.e., heart/lungs). Documented by imaging or autopsy.
Time frame: 2 years
The filters may be retrieved if clinically indicated. The IVC filter and IVC were assessed on imaging before a retrieval attempt.
(The filter retrieval rate= the total number of filter retrieved / the total number of filter retrieval attempted)
Cook Research Incorporated
Industry
Acronym: CIVC
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07689136
Cardiovascular Diseases, Deep Vein Thrombosis
Tunis, Tunisia
View Trial DetailsNCT04737954
Cardiovascular Diseases, Deep Vein Thrombosis
Hamburg, Germany
View Trial DetailsNCT02810704
Cardiovascular Diseases, Embolism
Scottsdale, Arizona, United States
View Trial DetailsNCT04168203
Cardiovascular Diseases, Deep Vein Thrombosis
Boston, Massachusetts, United States
View Trial Details