The objective is to validate the performance of SpO2 measurement of the TipTraQ by comparing it with measurement from an artery blood sample of a healthy volunteer in a controlled desaturation study. The overall framework of the validation process followed Annex EE.2 of ISO 80601-2-61:2019.
Looking for future studies?
Notify Me18 year–50 year
All sexes
Observational
San Francisco, California, 94143, United States
The TipTraQ, is a finger-based wearable for aiding OSA diagnosis at home that consists of a sensor placed on a fingertip. PPG data are produced by a reflective emitter and photodetector elements from one of the fingertips. TipTraQ Companion app can read and record the PPG signals from the TipTraQ Sensor and further transmit the data to TipTraQ Cloud.
The overall framework is a controlled desaturation study follows the Annex EE.2 of ISO 80601-2-61:2019. A radial arterial cannula was placed in each subject's left or right wrist for blood sampling and monitoring blood pressure. Each subject had two control blood samples taken at the beginning of each experiment while breathing room air. Hands with pulse oximeters were maintained motionless on arm boards. Hypoxia was then induced to different and stable levels of oxyhemoglobin saturation (between 70-100%) by having subjects breathe mixtures of nitrogen, air, and carbon dioxide controlled by the study physician by adjusting gas flows with valves according to breath-by-breath estimates of oxygen saturation calculated from end-tidal PO2 and PCO2. Each plateau level of oxyhemoglobin saturation was maintained for at least 30 seconds or until reference pulse oximeter readings were stable. During the procedure, the subject is under stationary status. Two arterial blood samples were then obtained approximately 30 seconds apart. Each stable plateau, therefore, was maintained for at least 60 seconds with SpO2 fluctuating by less than 2-3%. The plateaus were nominally at 100%, room air saturation, 93%, 90%, 87%, 85%, 82%, 80%, 77%, 75% and 70%. At least 200 data points were collected for each type of oximeter and probe combination studied.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: During the controlled desaturation study, within 1 day
oxygen saturation of arterial blood sample measured by blood gas analysis
Time frame: During the controlled desaturation study, within 1 day
Peripheral blood oxygen saturation measured through plethysmography(PPG)
PranaQ Pte. Ltd.
Industry
Accuracy of Pulse Oximeters with Profound Hypoxia
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07720960
Erythropoiesis, Hypoxia
Elche, Alicante, Spain
View Trial DetailsNCT05767385
Cardiovascular Abnormalities, Cardiovascular Diseases
San Francisco, California, United States
View Trial DetailsNCT01542177
Digestive System Diseases, Digestive System Neoplasms
Toronto, Ontario, Canada
View Trial DetailsNCT05354648
Arterial Occlusive Diseases, Arteriosclerosis
View Trial Details