Skip to main content
OpenTrials
Completed

NCT Number: NCT02247765

Controlled Acute Hypoxia Study Comparing Pulse Oximetry to Arterial Blood Samples During Motion

To validate the proposed claims for pulse rate and saturation accuracy in a diverse subject population during motion over a specified saturation range.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–50 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Phase 3

Primary location

Boulder Clinical Laboratory, Boulder, Colorado, United States

Loading trial locations.

About this study

Use an investigational oximetry system to evaluate the triggering of the "sensor off" status with marketed, off-the-shelf sensors.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Male or female of any race
  • 18-50 years old, inclusive
  • Females: negative urine pregnancy test on the day of study participation (prior to exposure to hypoxia)
  • Completed within the last year: physical exam by a licensed physician, physician assistant (PA), or advanced practice nurse; including a 12 lead ECG, a medical history, and blood test (complete blood count and sickle cell trait/disease screening)
  • Meets specific demographic requirements for the monitoring device under study
  • Willing and able to provide written informed consent
  • Able to participate for the duration of the evaluation

Exclusion criteria

  • A room-air baseline % modulation < 1.5% on all four fingers on the test hand
  • Under 18 years or over 50 years of age
  • Pregnant and/or lactating women
  • Hypertension: on three consecutive readings, systolic pressure greater than 145 mm Hg or diastolic pressure greater than 90 mm Hg
  • Ventricular premature complexes (VPC's) that are symptomatic or occur at a rate of more than four per minute
  • History of seizures (except childhood febrile seizures) or epilepsy
  • History of unexplained syncope
  • Daily or more frequent use of anxiolytic drugs (benzodiazepines) for treatment of anxiety disorder
  • Recent history of frequent migraine headaches: average of two or more per month over the last year
  • Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
  • History of acute altitude sickness at or below moderate elevation (up to 5,000-10,000 feet) defined as three or more of the following symptoms: moderate to severe headache, general malaise, dizziness/lightheadedness, nausea/vomiting, fatigue/weakness, shortness of breath, nosebleed, persistent rapid pulse, or peripheral edema
  • History of significant respiratory disease such as severe asthma or emphysema or sleep apnea
  • Sickle cell disease or trait
  • Clinically significant abnormal finding on medical history, physical examination, clinical laboratory test or ECG. Clinical significance will be assessed by the principal investigator or study physician as designated.
  • History of stroke, transient ischemic attack or carotid artery disease
  • History of myocardial ischemia, angina, myocardial infarction, congestive heart failure or cardiomyopathy
  • History of chronic renal impairment
  • Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
  • Unwillingness or inability to remove colored nail polish or artificial nails from test digit(s)
  • Unwillingness or inability to give informed consent
  • Prior or known allergies to: lidocaine (or similar pharmacological agents, e.g., Novocain) or heparin
  • Recent arterial cannulation (i.e., less than 30 days prior to study date)
  • Six or more arterial cannulations of each (right & left) radial artery
  • History of clinically significant complications from previous arterial cannulation
  • Current use of blood thinners: prescription or daily aspirin use
  • History of bleeding disorders or personal history of prolonged bleeding from injury.

Treatment and study plan

Arterial line

Device

Arterial line is inserted to draw a CO-Oximeter value as it is the gold standard and as advised in ISO 80601

Other names: CO-Oximeter SaO2 value

Motion

Device

The subject is asked to move their hand by either rubbing or tapping their fingers at a specified frequency and amplitude during portions of the hypoxic procedure.

Primary outcomes

  1. SpO2 Accuracy During Motion Conditions - MaxA Sensor

    Time frame: up to 6 months

    For each range specified, SpO2 accuracy of the pulse oximeter equipment is stated in terms of the Accuracy Root-Mean-Square (ARMS) difference between measured values (SpO2) and reference blood values (SaO2). The MaxA sensors has a different bandage from the MaxN sensor, and therefore a different form and fit.

  2. SpO2 Accuracy During Motion Conditions - MaxN Sensor

    Time frame: up to 6 months

    For each range specified, SpO2 accuracy of the pulse oximeter equipment is stated in terms of the Accuracy Root-Mean-Square (ARMS) difference between measured values (SpO2) and reference blood values (SaO2). The MaxN sensor has a different bandage from the MaxA sensor, and therefore a different form and fit.

Sponsors and collaborators

Lead sponsor

Medtronic - MITG

Industry

Registry information

Official study title

Controlled Acute Hypoxia Study Comparing Pulse Oximetry to Arterial Blood Samples During Motion in Healthy, Well Perfused Subjects With the USB Pulse Oximetry Monitor Interface Cable

Important dates

Study start
2014
Primary completion
2014
Study completion
2014
First posted
Sep 25, 2014
Registry last updated
May 5, 2017