University of North Carolina at Chapel Hill
Chapel Hill, North Carolina, 27599, United States
NCT Number: NCT04021238
The purpose of this study is to determine if contrast-enhanced ultrasound can detect abnormal features of kidney lesions in patients with suspected kidney cancer with the same accuracy as conventional ultrasound and contrast-enhanced magnetic resonance imaging (MRI)
Looking for future studies?
Notify Me18 year–99 year
All sexes
Interventional
Phase 2
Chapel Hill, North Carolina, 27599, United States
This is a pilot cross-sectional study that compares contrast enhanced ultrasound to conventional ultrasound and contrasted MRI. Eligible patients will include anyone who has suspected kidney cancer and is scheduled for surgery, up to an anticipated total of 40 participants. Subject participation will be only for the day of CEUS study. There will be no follow-up period for this study. However, if results are encouraging, a longitudinal observational study may follow, and these same subjects would be eligible for enrollment. Eligible subjects will undergo a contrast enhanced ultrasound.
Following completion of imaging, all CEUS, MRI (within 4 months) and B-mode (at time of CEUS) US studies will be de-identified. Blinded radiologists will interpret images and provide an overall assessment of risk of malignancy to each kidney using the Bosniak criteria for each kidney lesion present. The Bosniak criteria places cystic lesions into one of 5 categories (I, II, IIF, III and IV) based on lesion characteristics. CEUS based diagnosis will be compared to the diagnoses on routine B-mode US and contrast-enhanced MRI.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Baseline B-mode US will be performed to confirm the lesions being imaged (unless it is a subject with no lesions, then a basic B-mode US will be performed. Perflutren is a diagnostic drug that is intended to be used for contrast enhancement. The vial contains a clear, colorless, sterile, non-pyrogenic, hypertonic solution which is activated by mechanical agitation with Vialmix®.It will be dispensed in inactivated form to a study team member. The contrast agent will be activated just prior to administration (ideally to be used within 5 minutes of activation). Contrast agent Definity (or Lumason) will be activated according to package insert instructions, including use of VialMix®, the device used to activate Definity. Administration will occur IV in coordination with sonography staff trained specifically in contrast ultrasound imaging.The total imaging time is anticipated to be less than 30 minutes.
Other names: Definity, Microbubble contrast agent
Baseline B-mode US will be performed to confirm the lesions being imaged (unless it is a subject with no lesions, then a basic B-mode US will be performed. Lumason will be used as a secondary contrast agent only if Definity(perflutren) is unavailable.This drug will be administered using the dosing range and administration type within the Lumason prescribing information. The contrast agent will be activated just prior to administration (ideally to be used within 5 minutes of activation). Contrast agent Definity (or Lumason) will be activated according to package insert instructions, including use of VialMix®, the device used to activate Definity. Administration will occur IV in coordination with sonography staff trained specifically in contrast ultrasound imaging.The total imaging time is anticipated to be less than 30 minutes.
Other names: Lumason
Time frame: Baseline- up to 3 minutes
A study-specific model was created to predict kidney tumor type. The contrast-enhanced ultrasonogram (CEUS) Wash_In_Slope parameter was generated. This metric indicates the change in contrast with time in the mass using bolus imaging. Bolus imaging injects a small amount of contrast agent and monitors the contrast agent as it flows in (wash-in) and out (wash-out) of the body/mass. The intensity of the signal (the brightness of the image) is plotted over time to generate the time-intensity curve. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White.The Wash_In_Slope increases as the contrast agent is introduced to the body/mass. This value was generated by fitting the data to a linear equation. Increasing brightness is related to the slope of curve.
Time frame: Baseline- up to 3 minutes
A study-specific model was created to predict kidney tumor type. The contrast-enhanced ultrasonogram (CEUS) Wash_In_intercept parameter was generated. This metric indicates the intensity level prior to infusion of contrast. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=white. This value was calculated using an equation for a line which has a slope and the intercept. The values were negative because they were calculated (not measured).
Time frame: Baseline- up to 3 minutes
A study-specific model was created to predict kidney tumor type. The contrast-enhanced ultrasonogram (CEUS) Wash_Out_Slope parameter was generated. This metric indicates the change in contrast with time in the mass using bolus imaging. Bolus imaging injects a small amount of contrast agent and monitors the contrast agent as it flows in (wash-in) and out (wash-out) of the body/mass. The intensity of the signal (the brightness of the image) is plotted over time to generate the time-intensity curve. The intensity of the signal (the brightness of the image) value range was between 0=black and 255=White. The wash Out slope decreases as the contrast agent clears the body/mass. Decreasing brightness is related to the slope of the curve. This results in a negative number. This value was generated by fitting the data to a linear equation.
Time frame: Baseline- up to 3 minutes
A study-specific model was created to predict kidney tumor type. The contrast-enhanced ultrasonogram (CEUS) Wash_Out_Intercept parameter was generated. This metric indicates the intensity level of contrast at a steady state of enhancement in the mass. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White. This value was calculated using an equation for a line that has a slope and an intercept.
Time frame: Baseline- up to 3 minutes
A study-specific model was created to predict kidney tumor type. The contrast-enhanced ultrasonogram (CEUS) FR_a parameter was generated. The intensity of the signal (the brightness of the image) value range is between 0=black, and 255=white. The intensity was plotted by time. FR_a parameters were devised from an exponential fit. The exponential fit of the data provides the height of the curve.
Time frame: Baseline- up to 3 minutes
A study-specific model was created to predict kidney tumor type. The contrast-enhanced ultrasonogram (CEUS) FR_beta parameter was generated. The intensity of the signal (the brightness of the image) value range is between 0=black, 255=White. The intensity was plotted by time. FR_beta parameters were devised from an exponential fit. Exponential fit of the data provides the slope of the curve.
Time frame: Baseline- up to 3 minutes
A study-specific model was created to predict kidney tumor type. The contrast-enhanced ultrasonogram (CEUS) volume parameter was generated. The volumes were generated using the ellipsoid equation on the CEUS image of the kidney mass.
Time frame: Baseline- up to 3 minutes
The association between tumor pathology type (Clear cell renal carcinoma (CCRCC) versus none (CCRCC) pathology) and CEUS-generated Wash_In_Slope parameter was studied. This metric indicates the change in contrast intensity with time in the mass using bolus imaging. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White. The Wash_In_Slope increases as the contrast agent is introduced to the body/mass. This value was generated by fitting the data to a linear equation. Increasing brightness is related to the slope of the curve.
Time frame: Baseline- up to 3 minutes
The association between tumor pathology type (Clear cell renal carcinoma (CCRCC) versus none (CCRCC) pathology) and CEUS-generated Wash_In_intercept parameter was studied. This metric indicates the intensity level prior to infusion of contrast. The intensity of the signal (the brightness of the image) value range is between 0=black, and 255=White. This value was calculated using an equation for a line which has a slope and the intercept. The values were negative because they were calculated (not measured).
Time frame: Baseline- up to 3 minutes
The association between tumor pathology type (Clear cell renal carcinoma (CCRCC) versus none (CCRCC) pathology) and CEUS-generated Wash_Out_Slope parameter was studied. This metric indicates the change in contrast intensity with time in the mass using bolus imaging. The intensity of the signal (the brightness of the image) value range is between 0=black, and 255=White. Decreasing brightness is related to the slope of the curve. The Wash Out slope decreases as the contrast agent clears the body/mass. Decreasing brightness is related to the slope of the curve. This results in a negative number. This value was generated by fitting the data to a linear equation.
Time frame: Baseline- up to 3 minutes
The association between tumor pathology type (Clear cell renal carcinoma (CCRCC) versus none (CCRCC) pathology) and CEUS-generated Wash_Out_Intercept parameter was studied. This metric indicates the intensity level of contrast at steady state of enhancement in the mass. The contrast-enhanced ultrasonogram (CEUS) Wash_Out_Intercept parameter was generated. The intensity of the signal (the brightness of the image) value range is between 0=black, and 255=White. This value was calculated using an equation for a line which has a slope and the intercept.
Time frame: Baseline- up to 3 minutes
The association between tumor pathology type (Clear cell renal carcinoma (CCRCC) versus none (CCRCC) pathology) and CEUS-generated FR_a parameter was studied. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White. The intensity was plotted by time. FR_a parameters were devised from an exponential fit. The exponential fit of the data provides the height of the curve.
Time frame: Baseline- up to 3 minutes
The association between tumor pathology type (Clear cell renal carcinoma (CCRCC) versus none (CCRCC) pathology) and CEUS-generated FR_ beta parameter was studied. The contrast-enhanced ultrasonogram (CEUS) FR_beta parameter was generated. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White. The intensity was plotted by time. FR_beta parameters were devised from an exponential fit. The exponential fit of the data provides the slope of the curve.
Time frame: Baseline- up to 3 minutes
The association between tumor pathology type (Clear cell renal carcinoma (CCRCC) versus none (CCRCC) pathology) and CEUS-generated volume parameter was studied. The volumes were generated using the ellipsoid equation on the CEUS image of the kidney mass.
Time frame: Baseline- up to 3 minutes
The intensity of the signal (the brightness of the image) is plotted over time to generate the time-intensity curve. The intensity of the signal (the brightness of the image) values range between 0=black and 255=White. The Wash_In_Slope increases as the contrast agent is introduced to the body/mass. This value was generated by fitting the data to a linear equation. Increasing brightness is related to the slope of the curve. The association between tumor stage (Tumor stage 0,1,2,3) and CEUS-generated Wash_In_Slope parameter was studied.
Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to American Joint Cancer Committee. All masses in this study were in stage 0, I, II or III (not stage IV). Stage 0 = cancer cell was not found on pathology. Stage I =tumors size ≤7 cm. Stage II = tumor size >7 cm. Stage III includes tumors that are growing into a major vein or tissue around the kidney with no spread to lymph nodes.
Time frame: Baseline- up to 3 minutes
The association between tumor stage and CEUS-generated Wash_In_intercept parameter was studied. These values were generated by fitting the data to a linear equation. This metric indicates the intensity level prior to infusion of contrast. The intensity of the signal (the brightness of the image) value range is between 0=black, and 255=White. This value was calculated using an equation for a line that has a slope and an intercept. The values were negative because they were calculated (not measured).
Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to the American Joint Cancer Committee. All masses in this study were in stage 0, I, II, or III (not stage IV). Stage 0 = cancer cell was not found on pathology. Stage I =tumors size ≤7 cm. Stage II = tumor size >7 cm. Stage III includes tumors that are growing into a major vein or tissue around the kidney with no spread to lymph nodes.
Time frame: Baseline- up to 3 minutes
The association between tumor stage (Tumor stage 0,1,2,3) and CEUS-generated Wash_Out_Slope parameter was studied. The intensity of the signal (the brightness of the image) is plotted over time to generate the time-intensity curve. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White. The Wash_Out_Slope decreases as the contrast agent clears the body/mass. Decreasing brightness is related to the slope of the curve. This results in a negative number. This value was generated by fitting the data to a linear equation. Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to the AJCC. All masses in this study were in stage 0, I, II, or III (not stage IV). Stage 0 = cancer cell was not found on pathology. Stage I =tumors size ≤7 cm. Stage II = tumor size >7 cm, Stage III includes tumors that are growing into a major vein or tissue around the kidney with no spread to lymph nodes.
Time frame: Baseline- up to 3 minutes
The association between tumor stage (Tumor stage 0,1,2,3) and CEUS-generated Wash_Out_Intercept parameter was studied. This metric indicates the intensity level of contrast at a steady state of enhancement in the mass. The contrast-enhanced ultrasonogram (CEUS) Wash_Out_Intercept parameter was generated. The intensity of the signal (the brightness of the image) value range is between 0=black, and 255=White. This value was calculated using an equation for a line that has a slope and the intercept.
Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to the American Joint Cancer Committee. All masses in this study were in stage 0, I, II, or III (not stage IV). Stage 0 = cancer cell was not found on pathology. Stage I =tumors size ≤7 cm. Stage II = tumor size >7 cm. Stage III includes tumors that are growing into a major vein or tissue around the kidney with no spread to lymph nodes.
Time frame: Baseline- up to 3 minutes
The association between tumor stage (Tumor stage 0,1,2,3) and CEUS-generated FR_a parameter was studied. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White. The intensity was plotted by time. FR_a parameters were devised from an exponential fit. The exponential fit of the data provides the height of the curve.
Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to the American Joint Cancer Committee. All masses in this study were in stage 0, I, II or III (not stage IV). Stage 0 = cancer cell was not found on pathology. Stage I =tumors size ≤7 cm. Stage II = tumor size >7 cm. Stage III includes tumors that are growing into a major vein or tissue around the kidney with no spread to lymph nodes.
Time frame: Baseline- up to 3 minutes
The association between tumor stage (Tumor stage 0,1,2,3) and CEUS-generated FR_ beta parameter was studied. The contrast-enhanced ultrasonogram (CEUS) FR_beta parameter was generated. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White. The intensity was plotted by time. FR_beta parameters were devised from an exponential fit. The exponential fit of the data provides the slope of the curve.
Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to the American Joint Cancer Committee. All masses in this study were in stage 0, I, II, or III (not stage IV). Stage 0 = cancer cell was not found on pathology. Stage I =tumors size ≤7 cm. Stage II = tumor size >7 cm. Stage III includes tumors that are growing into a major vein or tissue around the kidney with no spread to lymph nodes.
Time frame: Baseline- up to 3 minutes
The association between tumor stage (Tumor stage 0,1,2,3) and CEUS-generated volume was studied. The contrast-enhanced ultrasonogram (CEUS) volume was generated.
Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to the American Joint Cancer Committee. All masses in this study were in stage 0, I, II or III (not stage IV). Stage 0 = cancer cell was not found on pathology. Stage I =tumors size ≤7 cm. Stage II = tumor size >7 cm. Stage III includes tumors that are growing into a major vein or tissue around the kidney with no spread to lymph nodes.
Time frame: Baseline- up to 3 minutes
The association between pooled tumor stage (tumor stages 0,1,2 versus tumor stage 3) and CEUS-generated Wash_In_Slope parameter was studied. The intensity of the signal (the brightness of the image) is plotted over time to generate the time-intensity curve. The intensity of the signal values range between 0=black and 255=White. The Wash_In_Slope increases as the contrast agent is introduced to the body/mass. This value was generated by fitting the data to a linear equation. Increasing brightness is related to the slope of the curve. Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to the American Joint Cancer Committee. All masses in this study were in stage 0, I, II, or III (not stage IV). Stage 0 = cancer cell was not found on pathology. Stage I =tumors size ≤7 cm. Stage II = tumor size >7 cm. Stage III includes tumors that are growing into a major vein or tissue around the kidney with no spread to lymph nodes.
Time frame: Baseline- up to 3 minutes
The association between pooled tumor stage (tumor stages 0,1,2 versus tumor stage 3) and CEUS-generated Wash_Out_Slope parameter was studied. The intensity of the signal (the brightness of the image) is plotted over time to generate the time-intensity curve. The intensity of the signal (the brightness of the image) values range between 0=black, and 255=White. The Wash_Out_Slope decreases as the contrast agent clears the body/mass. Decreasing brightness is related to the slope of the curve. This results in a negative number. This value was generated by fitting the data to a linear equation. Tumor staging is based on the size of the mass, spread to lymph nodes, and metastasis to distant sites according to the AJCC. All masses in this study were in stage 0, I, II, or III (not stage IV). Stage 0 = cancer cell was not found, Stage I = tumor size ≤7 cm, Stage II = tumor size >7 cm, Stage III includes tumors growing into a major vein or tissue, with no spread to lymph nodes.
UNC Lineberger Comprehensive Cancer Center
Other
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06955910
Behavior, Behavioral Symptoms
Padova, Italy/Padova, Italy
View Trial DetailsNCT03345563
Bladder Cancer, Colon Cancer
New Brunswick, New Jersey, United States
View Trial DetailsNCT00403169
Adenocarcinoma, Carcinoma
Cleveland, Ohio, United States
View Trial DetailsNCT04000880
Adnexal Diseases, Behavior
Birmingham, Alabama, United States
View Trial Details