NCT Number: NCT05889832
The purpose of this study it to evaluate the accuracy of heart rate, heart rate variability and breathing rate measurement with the use of Shen.AI Vitals software developed by MX Labs.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Wroclaw, Lower Silesian Voivodeship, 50-556, Poland
Resting heart rate (HR), heart rate variability (HRV) and breathing rate (BR) will be measured (estimated) with the use of remote photoplethysmography (1-minute face video analysis with Shen.AI Vitals software).
The measurements will be taken with various camera-equipped devices (smartphones, tablets, laptops) using either front camera (self-measurement) or rear camera (measurement made by someone else), in two different lighting conditions (artificial light or natural daylight), with various levels of device stability (fully stable or held in hand/s), and in two different body positions (sitting or supine). In total, each participant will have 8 measurements taken.
Standard deviation of normal heartbeat intervals (SDNN) will be used as the main HRV index.
Apart from the average HR (averaged over the whole 1-minute measurement), instantaneous values of HR (averaged over 10 s) will also be determined and displayed every 1 s by the Shen.AI VItals software.
In all cases, reference measurements will be taken simultaneously (chest ECG for the calculation of average and instantaneous HR as well as HRV; impedance pneumography for the determination of BR).
The reference measurements will be taken with Finapres Nova with the ECG+Resp module (Finapres Medical Systems B.V.).
For each measured parameter, the results from both methods will be compared using the Bland-Altman analysis or non-parametric limits of agreement.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Resting heart rate (HR), heart rate variability (HRV) and breathing rate (BR) will be measured (estimated) with the use of remote photoplethysmography (1-minute face video analysis with Shen.AI Vitals software).
Reference measurements will be taken simultaneously (chest ECG for the calculation of HR and HRV; impedance pneumography for BR).
Time frame: day 1 (single visit, approximately 30 minutes)
In each participant, average HR will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.
The results from both methods will be compared using non-parametric limits of agreement.
Time frame: day 1 (single visit, approximately 30 minutes)
In each participant, the standard deviation of normal heartbeat intervals (SDNN) will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
In each participant, the breathing rate will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded impedance pneumography signal. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
In each participant, multiple instantaneous heart rate values (averaged over 10 s) will be estimated from the software under test and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.
The results from both methods will be compared using non-parametric limits of agreement.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of HR using the rear camera).
For reference, HR will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using non-parametric limits of agreement.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of SDNN using the rear camera).
For reference, SDNN will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of BR using the rear camera).
For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of HR using the rear camera).
For reference, HR will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using non-parametric limits of agreement.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of SDNN using the rear camera).
For reference, SDNN will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of BR using the rear camera).
For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of HR using the front smartphone camera).
For reference, HR will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using non-parametric limits of agreement.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of SDNN using the front smartphone camera).
For reference, SDNN will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of BR using the front smartphone camera).
For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one facial video measurement performed in the supine position (1-minute measurement of HR using the front smartphone camera).
For reference, HR will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using non-parametric limits of agreement.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one facial video measurement performed in the supine position (1-minute measurement of SDNN using the front smartphone camera).
For reference, SDNN will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (single visit, approximately 30 minutes)
Each participant will have one facial video measurement performed in the supine position (1-minute measurement of BR using the front smartphone camera).
For reference BR will be obtained from the simultaneously recorded impedance pneumography signal.
The results from both methods will be compared using the Bland-Altman analysis.
Time frame: day 1 (after a single visit, approximately 30 minutes)
The utility of the software under test for self-measurement of physiological parameters (HR, SDNN, and BR) will be assessed based on of the average score in the participant's questionnaire completed by each participant after the test procedure (i.e. after all measurements in the given participant).
The scale will be 1-5, with higher scores meaning better outcome.
Time frame: day 1 (after a single visit, approximately 30 minutes)
The utility of the software under test for the video-based measurement of someone's physiological parameters will be assessed based on the average score in the participant's questionnaire (the part of the questionnaire related to the measurement with the rear camera of a smartphone) completed by each participant after the test procedure(i.e. after all measurements in the given participant).
The scale will be 1-5, with higher scores meaning better outcome.
SHEN.AI spółka z ograniczoną odpowiedzialnością oddział w Polsce
Industry
Contactless Measurement of Heart Rate, Heart Rate Variability and Breathing Rate Using Face Video Analysis (Remote Photoplethysmography) - Validation of Shen.AI Vitals Software.
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07477834
Air Pollution Exposure, HRV
Abu Dhabi, United Arab Emirates
View Trial DetailsNCT06829615
Blood Pressure, Breathing Rate
Wroclaw, Lower Silesian Voivodeship, Poland
View Trial DetailsNCT05366049
Body Temperature Changes, Fever
Istanbul, Fatih, Turkey (Türkiye)
View Trial DetailsNCT03754231
Heart Rate, Respiratory Rate
Singapore
View Trial Details