NCT Number: NCT05889832

Contactless Measurement of Heart Rate, Heart Rate Variability and Breathing Rate Using Remote Photoplethysmography

The purpose of this study it to evaluate the accuracy of heart rate, heart rate variability and breathing rate measurement with the use of Shen.AI Vitals software developed by MX Labs.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Uniwersyteckie Centrum Wsparcia Badań Klinicznych we Wrocławiu

Wroclaw, Lower Silesian Voivodeship, 50-556, Poland

About this study

Resting heart rate (HR), heart rate variability (HRV) and breathing rate (BR) will be measured (estimated) with the use of remote photoplethysmography (1-minute face video analysis with Shen.AI Vitals software).

The measurements will be taken with various camera-equipped devices (smartphones, tablets, laptops) using either front camera (self-measurement) or rear camera (measurement made by someone else), in two different lighting conditions (artificial light or natural daylight), with various levels of device stability (fully stable or held in hand/s), and in two different body positions (sitting or supine). In total, each participant will have 8 measurements taken.

Standard deviation of normal heartbeat intervals (SDNN) will be used as the main HRV index.

Apart from the average HR (averaged over the whole 1-minute measurement), instantaneous values of HR (averaged over 10 s) will also be determined and displayed every 1 s by the Shen.AI VItals software.

In all cases, reference measurements will be taken simultaneously (chest ECG for the calculation of average and instantaneous HR as well as HRV; impedance pneumography for the determination of BR).

The reference measurements will be taken with Finapres Nova with the ECG+Resp module (Finapres Medical Systems B.V.).

For each measured parameter, the results from both methods will be compared using the Bland-Altman analysis or non-parametric limits of agreement.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • age ≥ 18 years
  • ability to read and write in Polish or English
  • ability to operate a smartphone, tablet and laptop without the need for glasses
  • ability to understand the study procedure
  • ability to communicate and follow the instructions of the study staff
  • written consent to participate in the study and to the processing of personal data by the research team and the sponsor of the study

Exclusion criteria

  • failure to provide the required information in the participant's information survey
  • failure to comply with the study procedure or the instructions of the study stuff
  • extensive facial skin damage (including abrasions, wounds, burns)
  • a disease process involving a large part of the face that makes the video- based measurement impossible
  • extensive face dressing
  • extensive facial skin tattoo or face painting
  • marked deformation of the face (e.g. a tumor)
  • inability to keep the head stable in the required position during the measurement
  • persistent cough making it impossible to remain still during the measurement
  • respiratory dysfunction in the form of shortness of breath or irregular or shallow breathing
  • diagnosed arrhythmia (except sinus bradycardia or tachycardia)
  • moderate or severe anemia
  • diagnosed: heart failure, left ventricular systolic dysfunction, aortic stenosis or other diseases or disorders within the heart vascular system leading to decreased stroke volume, low pressure amplitude or the so-called paradoxical pulse

Treatment and study plan

measurement of physiological parameters wih Shen.AI Vitals software on various camera-equipped devices (smartphones, tablets, laptops)

Diagnostic Test

Resting heart rate (HR), heart rate variability (HRV) and breathing rate (BR) will be measured (estimated) with the use of remote photoplethysmography (1-minute face video analysis with Shen.AI Vitals software).

Reference measurements will be taken simultaneously (chest ECG for the calculation of HR and HRV; impedance pneumography for BR).

Primary outcomes

  1. Accuracy of heart rate (HR) measurement with the software under test compared with the ECG-based measurement

    Time frame: day 1 (single visit, approximately 30 minutes)

    In each participant, average HR will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.

    The results from both methods will be compared using non-parametric limits of agreement.

  2. Accuracy of heart rate variability index (SDNN) measurement with the software under test compared with the ECG-based measurement

    Time frame: day 1 (single visit, approximately 30 minutes)

    In each participant, the standard deviation of normal heartbeat intervals (SDNN) will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.

    The results from both methods will be compared using the Bland-Altman analysis.

  3. Accuracy of breathing rate (BR) measurement with the software under test compared with the impedance pneumography-based measurement

    Time frame: day 1 (single visit, approximately 30 minutes)

    In each participant, the breathing rate will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded impedance pneumography signal. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.

    The results from both methods will be compared using the Bland-Altman analysis.

  4. Accuracy of instantaneous heart rate measurement with the software under test compared with the ECG-based measurement

    Time frame: day 1 (single visit, approximately 30 minutes)

    In each participant, multiple instantaneous heart rate values (averaged over 10 s) will be estimated from the software under test and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.

    The results from both methods will be compared using non-parametric limits of agreement.

Secondary outcomes

  1. Accuracy of measurement of heart rate (HR) with the software under test on a smartphone held by someone else (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of HR using the rear camera).

    For reference, HR will be calculated from the simultaneously recorded ECG signal.

    The results from both methods will be compared using non-parametric limits of agreement.

  2. Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test on a smartphone held by someone else (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of SDNN using the rear camera).

    For reference, SDNN will be calculated from the simultaneously recorded ECG signal.

    The results from both methods will be compared using the Bland-Altman analysis.

  3. Accuracy of measurement of breathing rate (BR) with the software under test on a smartphone held by someone else (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of BR using the rear camera).

    For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.

    The results from both methods will be compared using the Bland-Altman analysis.

  4. Accuracy of measurement of heart rate (HR) with the software under test on a smartphone held in hand/s (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of HR using the rear camera).

    For reference, HR will be calculated from the simultaneously recorded ECG signal.

    The results from both methods will be compared using non-parametric limits of agreement.

  5. Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test on a smartphone held in hand/s (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of SDNN using the rear camera).

    For reference, SDNN will be calculated from the simultaneously recorded ECG signal.

    The results from both methods will be compared using the Bland-Altman analysis.

  6. Accuracy of measurement of breathing rate (BR) with the software under test on a smartphone held in hand/s (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of BR using the rear camera).

    For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.

    The results from both methods will be compared using the Bland-Altman analysis.

  7. Accuracy of measurement of heart rate (HR) with the software under test in daylight conditions (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of HR using the front smartphone camera).

    For reference, HR will be calculated from the simultaneously recorded ECG signal.

    The results from both methods will be compared using non-parametric limits of agreement.

  8. Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test in daylight conditions (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of SDNN using the front smartphone camera).

    For reference, SDNN will be calculated from the simultaneously recorded ECG signal.

    The results from both methods will be compared using the Bland-Altman analysis.

  9. Accuracy of measurement of breathing rate (BR) with the software under test in daylight conditions (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of BR using the front smartphone camera).

    For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.

    The results from both methods will be compared using the Bland-Altman analysis.

  10. Accuracy of measurement of heart rate (HR) with the software under test in the supine position (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one facial video measurement performed in the supine position (1-minute measurement of HR using the front smartphone camera).

    For reference, HR will be calculated from the simultaneously recorded ECG signal.

    The results from both methods will be compared using non-parametric limits of agreement.

  11. Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test in the supine position (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one facial video measurement performed in the supine position (1-minute measurement of SDNN using the front smartphone camera).

    For reference, SDNN will be calculated from the simultaneously recorded ECG signal.

    The results from both methods will be compared using the Bland-Altman analysis.

  12. Accuracy of measurement breathing rate (BR) with the software under test in the supine position (compared with reference measurements).

    Time frame: day 1 (single visit, approximately 30 minutes)

    Each participant will have one facial video measurement performed in the supine position (1-minute measurement of BR using the front smartphone camera).

    For reference BR will be obtained from the simultaneously recorded impedance pneumography signal.

    The results from both methods will be compared using the Bland-Altman analysis.

  13. Utility of the software under test for self-measurement of physiological parameters (assessed with a questionnaire).

    Time frame: day 1 (after a single visit, approximately 30 minutes)

    The utility of the software under test for self-measurement of physiological parameters (HR, SDNN, and BR) will be assessed based on of the average score in the participant's questionnaire completed by each participant after the test procedure (i.e. after all measurements in the given participant).

    The scale will be 1-5, with higher scores meaning better outcome.

  14. Utility of the software under test for the measurement of someone's physiological parameters (assessed with a questionnaire).

    Time frame: day 1 (after a single visit, approximately 30 minutes)

    The utility of the software under test for the video-based measurement of someone's physiological parameters will be assessed based on the average score in the participant's questionnaire (the part of the questionnaire related to the measurement with the rear camera of a smartphone) completed by each participant after the test procedure(i.e. after all measurements in the given participant).

    The scale will be 1-5, with higher scores meaning better outcome.

Sponsors and collaborators

Lead sponsor

SHEN.AI spółka z ograniczoną odpowiedzialnością oddział w Polsce

Industry

Registry information

Official study title

Contactless Measurement of Heart Rate, Heart Rate Variability and Breathing Rate Using Face Video Analysis (Remote Photoplethysmography) - Validation of Shen.AI Vitals Software.

Important dates

Study start
2023
Primary completion
2023
Study completion
2023
First posted
Jun 5, 2023
Registry last updated
Mar 28, 2024

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)