NCT Number: NCT04210830

Comparison Two Point of Care Coagulation Monitoring Systems - the ROTEM Sigma and the Quantra Devic

Human Research Project Risk Category A. In this study no additional intervention or treatment are performed. The whole blood samples are taken from a routinely placed arterial catheter. The blood loss caused by the two additional blood samples is estimated with 2x 2.7ml. The Risk for the patients to participate in this study is minimal.

Clinical observational study of a point of care in vitro diagnostic device, the Quantra®system (Quantra), that received the CE Mark in April 2017 and is approved by the Swiss Medic for clinical use. The Quantra estimates the coagulation strength of whole blood by sonorheometry technology applying a series of ultrasound pulses to a whole blood sample and measuring its stiffness by returning echoes.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Klinik Im Park

Zurich, 8027, Switzerland

About this study

The purpose of this study is to investigate to what extent the results of the Quantra corresponds to those of the ROTEM® Sigma (ROTEM). Additionally, the time required for delivery of final results (turnaround time) of all parameters of the Quantra will be examined. Significant differences in the turnaround time between both devices are of high interest as timing is a crucial factor in the diagnosis-based treatment.

Comparison of the coagulation process measurements in turnaround time and results of the ROTEM INTEM CT, HEPTEM CT, EXTEM A10 FIBTEM A10 with corresponding Quantra parameters such as CT, CTH, CTR, CS, PCS, FCS.

The primary aim of the study is to examine whether the two methods of hemostasis analysis are equivalent and how far they correlate.

Secondary aims are to examine the turnaround time of test results in both devices and to examine whether these results would have led to the same therapeutic decisions in a retrospective review of data.

For both devices following data is documented:

  • Time for cartridge insertion
  • Time first results available
  • Time measurement finished

Parameters measured ROTEM Quantra

  • INTEM CT Clot Time (CT)
  • HEPTEM CT Heparinase Clot time
  • (CTH)
  • INTEM CT/HEPTEM CT Clot Time Ratio (CTR)
  • EXTEM A10 Clot Stiffness (CS)
  • FIBTEM A10 Fibrinogen Contribution
  • to Clot Stiffness (FCS)
  • EXTEM A10 - FIBTEM A10 Platelet Contribution to
  • Clot Stiffness (PCS)

Blood loss estimated (mL/6h; mL/12h; mL/24 h) RCP transfused FFP packages transfused Platelet packages transfused Fibrinogen applied F XIII applied

Quantra clot stiffness values are expressed in hectoPascals (hPa), whereas corresponding ROTEM values are expressed as an amplitude in mm. The relationship between amplitude (A in mm) and shear modulus (G in Pascals) is not linear. For a proper comparison the ROTEM amplitude (mm) needs to be converted to Pascals by following formula:

G (Pascals) = (500 x A) / (100 - A),

as described by Solomon C et al, Anaesth Analg 2015; 121:868-78.

The Quantra data that are collected during the study are only observational and will not be used to guide therapeutic interventions

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patients scheduled for cardiac surgery needing a cardiopulmonary bypass
  • Age older than18 years
  • Written consent of the participation after clarification about the study

Exclusion criteria

  • No consent to participate
  • No German comprehension
  • Known coagulation disorder

Treatment and study plan

Quantra-Device

Diagnostic Test

Point of Care Coagulation Monitoring System

Primary outcomes

  1. Equivalence of the results of both devices

    Time frame: 1 Year

    Comparision of the results of the INTEM CT (R) with the Clot Time (Q)

  2. Equivalence of the results of both devices

    Time frame: 1 Year

    HEPTEM CT (R) was compared with the Heparinase Clot ime (Q)

  3. Equivalence of results of both devices

    Time frame: 1 Year

    INTEM CT/HEPTEM CT (R) was compared with the Clot Time Ratio (Q)

  4. Equivalence of results of both devices

    Time frame: 1 Year

    EXTEM A10 (R) was compared with the Clot Stiffness (Q)

  5. Equivalence of results of both devices

    Time frame: 1 Year

    FIBTEM A10 (R) was compared with the Fibrinogen Contribution to Clot Stiffness (Q)

  6. Equivalence of results of both devices

    Time frame: 1 Year

    Difference of EXTEM A10 and FIBTEM A10 (R) was compared with the Platelet Contribution to Clot Stiffness (Q)

Secondary outcomes

  1. Turnaround time of the measurements

    Time frame: 1 Year

    Time from taking the blood sample until introduction the Cartridge, in seconds

  2. Turnaround time of the measurements

    Time frame: 1 Year

    Time from Insertion the Cartridge until the first plotted result

  3. Turnaround time of the measurements

    Time frame: 1 Year

    Time from insertion the Cartridge until the all results are plotted

Sponsors and collaborators

Lead sponsor

Klinik Hirslanden, Zurich

Other

Registry information

Official study title

Comparison of Methods for Coagulation Analysis Between Rotem Sigma (Instrumentation Laboratory, Bedfort, MA, USA) and Quantra System (Hemosonics, Charlottesville, VA USA) in Cardiac Surgery

Important dates

Study start
2018
Primary completion
2019
Study completion
2019
First posted
Dec 26, 2019
Registry last updated
Dec 26, 2019

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.