This study will compare two supraglottic airway devices currently on the market. Supraglottic airway devices are used during general anesthesia to provide a patent airway. They are inserted blindly into the mouth once the patient has lost consciousness and they cover the laryngeal inlet. The I-Gel and the Supreme laryngeal mask airway (LMA) are the two devices to be compared. The I-Gel has no inflatable cuff, which makes it different from all other supraglottic airway devices currently in use. The Supreme shares some characteristics of the other LMA devices, but it is disposable. Our main goal will be to compare the airway leak pressure and the peak airway pressure of each devices. We will also measure the time needed for insertion, number of attempts needed to secure the airway and side effects related to the airway (cough, dysphagia, trauma, hoarseness of voice, sore throat).
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Montreal, Quebec, H1T 2M4, Canada
First of all, we will record the following demographic variables: age, sex, height, weight, BMI, surgery planned, American Society of Anesthesiology (ASA) status, dental state (good, bad, edentulous), tobacco use (none, ceased, active) and packs year, Mallampati score, thyromental distance and whether or not there is trauma on the tongue, teeth or lips prior to surgery.
Induction of anesthesia will be standardized with Fentanyl 1-3 mcg/kiloGram(kG) and Propofol 1-3,5 mg/kG. The use of myorelaxant will be mandatory, but either succinylcholine or rocuronium will be used as long as complete disappearance of T1 is observed before attempting supraglottic airway insertion.
Prior to insertion, ease of ventilation (yes/no) will be noted.
The two airway devices will be used according to manufacturers instructions. The sizes of the devices will be chosen according to weight:
I-Gel: 50-90 kG: #4 and 90 Kg and up: #5 LMA Supreme: 50-70 Kg: #4 and 70-100 Kg: #5.
Patients will be randomized in either I-Gel or LMA Supreme group. Two insertions with the same device will be allowed before a crossover to the other device. If the third attempt fails, a standard endotracheal tube will be put in the trachea and blood in the oral cavity will be sought. After each failed attempt (no thoracic expansion, no carbon dioxide (CO2) square wave or leak at standardized volume. Blood on the devices will also be recorded.
Since the LMA Supreme has a cuff, we will inflate it to 25 mL and then by intervals of 5 mL of air (max 45 mL) we will inflate it at a volume where no audible leak is heard at a standardized ventilation scheme of volume controlled 8 mL/kg X 10 breaths per minute.
Once fixed with tape, we will measure peak and mean airway pressure for three breaths after 30 seconds of volume controlled breathing. Then the fresh gas flow will be fixed at 3 L/min of 50-50 air-oxygen blend and the Adjustable Pressure Limiting valve will be closed. We will look either for a plateau pressure at which a leak equal to the fresh gas flow or a pressure of 40 cm of water.
We will sought adverse effect like bronchospasm, cough, regurgitation, desaturation, aspiration and deglutition movement.
Ease of insertion will be score on a scale of 1 to 4. 1 being very easy and 4 being very hard.
Endoscopic evaluation will take place afterwards. First in the ventilation tube. Scoring for laryngeal view:
Downfolding of the epiglottis will be sought.
Then we will go through the oesophageal port and see if the oesophageal mucosa il visible.
During surgery, we will check if airway manipulation and/or withdrawal was necessary and why.
If traces of blood on the devices are present at the end of the intervention it will be noted.
Two interviews will take place one at the post anesthesia care unit and another the following day by phone if the patient was having ambulatory surgery. Sore throat, cough, dysphagia and dysphonia will be asked to be graded: absent, mild, moderate or severe.
Pulse and arterial pressure with the operating room's sphygmomanometer and oximeter will be written before induction, after induction before insertion and at 1 and 2 minutes post insertion.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
2 attempts to secure the airway will be tried. If they both fail, there will be a crossover with a Supreme LMA.
In the LMA Supreme group, the same interventions will be done as in the I-Gel group. The cuff of the LMA Supreme will be inflated with 25 mL of air and then by intervals of 5 mL until no leak is heard (max 45 mL). Leak pressure will also be measure at 45 mL. Once the leak test is done, the cuff will be deflated to initial volume necessary to avoid audible leaks.
In case both supraglottic airway devices fail, a standard endotracheal tube will be inserted
Time frame: After introduction of the supraglottic device before the beginning of the surgery.
Time frame: After introduction of the SAD before the beginning of the surgery.
Time frame: From opening of the mouth to thoracic expansion and presence of End Tidal CO2 up to five minutes.
Time frame: At the beginning of anesthesia before the beginning of the surgery.
Time frame: On the day following surgery
Intent to treat assessment of adverse events, per intervention, was intended. The adverse effects were assessed by phone; or in person if patient in the hospital; on the following day.
Time frame: During and after anesthesia when the device is removed.
Maisonneuve-Rosemont Hospital
Other
Prospective Randomized Trial Comparing the Performance of Two Supraglottic Airway Management Devices: I-Gel and Supreme LMA
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07182786
Anesthesia Airway Management, Anesthesia for Endoscopic Procedures
Giza, Egypt
View Trial DetailsNCT07518849
Anesthesia Airway Management, Anesthesia Ergonomics
View Trial DetailsNCT06990308
Anesthesia Airway Management, Respiratory Complications
Marine Corps Base Camp Lejeune, North Carolina, United States
View Trial Details